The Bain circuit: (select 2)
a. is tested with the Pethick test
b. is a modified Mapleson A
c. delivers fresh gas through the inner tube
d. is best for spontaneous but not controlled ventilation
A. is tested with the Pethick test
C. delivers fresh gas through the inner tube
The Bain system is a modified
Mapleson D design
The Bain circuit is useful for
spontaneous and controlled ventilation
To prevent rebreathing, the FGF should be
2.5x minute ventilation
With the Bain circuit, fresh gas enters the circuit through the
thin inner tubing
With the Bain circuit, exhaled gas exits via the
corrugated tubing
_________- is used to test circuit integrity for the Bain system during the pre-anesthetic checkout procedure
The Pethick test
The outer corrugated tubing of the Bain circuit should be
transparent so that you can assess the integrity of the inner tubing
Describe a concern with using the Bain circuit.
the inner tubing can become kinked or disconnected which converts all of the corrugated tubing to dead space putting the patient at risk for hypercarbia
With the Pethick test, how do you know if the circuit is safe to use?
collapsed reservoir bag- inner tubing is patent and circuit is safe to use
inflated reservoir bag- the inner tubing is occluded and the circuit is not safe to use