PROXIMAL AIR IN LINE error?
A
- TURN OFF PUMP and CHECK FOR AIR in the LINE or CASSETTE
- if AIR in CASSETTE, “BACKPRIME” line by attaching a LARGE SYRINGE to the SECONDARY INLET PORT (white cap), then HOLD DOWN BACKPRIME BUTTON until ALL AIR BUBBLES HAVE BEEN FULLY EXPELLED into SYRINGE and IV LINE & CASSETTE are FULLY PRIMED WITH FLUID
1
Q
OCCLUSION error?
A
- TURN OFF PUMP
- CHECK LINE for KINKS or any CLOSED CLAMPS
- check the IV CATHETER FOR PATENCY via FLUSHING IN FLUID
- then RESTART FLUID PUMP
2
Q
DISTAL AIR IN LINE error?
A
- TURN OFF PUMP and CHECK FOR AIR in the LINE
- Then REPRIME IV LINE to CONFIRM FULLY PRIMED WITH FLUID
3
Q
VTBI COMPLETE error?
A
- REENTER VTBI AMOUNT IN SETTINGS
4
Q
MEMORY FAILURE error?
A
- REENTER ALL DATA (both VTBI and FLUID RATE) in SETTINGS
5
Q
CASSETTE error?
A
- TURN OFF PUMP and CHECK CASSETTE
- then REPRIME IF NEEDED and REINSERT CORRECTLY into FLUID PUMP
6
Q
DOOR error?
A
- TURN OFF FLUID PUMP
- OPEN the CASSETTE DOOR
- then CLOSE the CASSETTE DOOR PROPERLY