Y2 OSCE: Troubleshooting Fluids Flashcards

1
Q

PROXIMAL AIR IN LINE error?

A
  1. TURN OFF PUMP and CHECK FOR AIR in the LINE or CASSETTE
  2. if AIR in CASSETTE, “BACKPRIME” line by attaching a LARGE SYRINGE to the SECONDARY INLET PORT (white cap), then HOLD DOWN BACKPRIME BUTTON until ALL AIR BUBBLES HAVE BEEN FULLY EXPELLED into SYRINGE and IV LINE & CASSETTE are FULLY PRIMED WITH FLUID
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
1
Q

OCCLUSION error?

A
  1. TURN OFF PUMP
  2. CHECK LINE for KINKS or any CLOSED CLAMPS
  3. check the IV CATHETER FOR PATENCY via FLUSHING IN FLUID
  4. then RESTART FLUID PUMP
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

DISTAL AIR IN LINE error?

A
  1. TURN OFF PUMP and CHECK FOR AIR in the LINE
  2. Then REPRIME IV LINE to CONFIRM FULLY PRIMED WITH FLUID
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

VTBI COMPLETE error?

A
  1. REENTER VTBI AMOUNT IN SETTINGS
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

MEMORY FAILURE error?

A
  1. REENTER ALL DATA (both VTBI and FLUID RATE) in SETTINGS
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

CASSETTE error?

A
  1. TURN OFF PUMP and CHECK CASSETTE
  2. then REPRIME IF NEEDED and REINSERT CORRECTLY into FLUID PUMP
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

DOOR error?

A
  1. TURN OFF FLUID PUMP
  2. OPEN the CASSETTE DOOR
  3. then CLOSE the CASSETTE DOOR PROPERLY
  4. RESTART FLUID PUMP
How well did you know this?
1
Not at all
2
3
4
5
Perfectly