Radiation Fields Flashcards
T/F: radiation fields can differ depending on location of tumor and depth of tumor
TRUE
-every plan and patient can be unique however there are some standards
a single field, can also be called a single ____
port
single fields are used to treat what type of treatments?
superficial electron fields or palliative trts
why aren’t single fields super common?
because the dose distribution is not ideal for MV beams
the following are examples of which radiation field?
-AP IM chain
-PA spine
-supraclavicular nodes
single field
the following is an example of what radiation field?
*hint - notice there is no high isodose line on other side of patient
single field
T/F: you can use “Field” and “Port” interchangeably
TRUE
Parallel Opposed Fields (POPs) have ___ treatment fields/ports aimed at/in opposing directions
2
POP stands for -
Parallel Opposed Ports
POPs have a hinge angle of _____ !!!
180 degrees
POPs have a ______ dose distribution !!!
homogenous
POPs are used for more ____curative/palliative____ treatments?
curative
for POPs 2 fields are aimed along the same ____ from opposite sides
same axis
AP/PA and RT lat/LT lat are examples of what radiation field?
Parallel Opposed Fields//Ports
POPs lead to higher ____ doses, therefore it is even more common to use more than two fields to reduce these reactions
skin
_____ are the simplest form of multiple fields, and are very reproducible
POPs
_____ have a homogenous dose to tumor
POPs (Parallel Opposed Ports/Fields)
the following is an example of what radiation field?
POPs (Parallel Opposed Ports/Fields)
For which radiation field, is there less sparing of tissues around tumor - when it is compared to the other techniques
POPs
good note - in general, multiple fields (so anything over a single field) are used more for ___curative/palliative___ treatments
curative
___less/more___ fields allow for higher doses to the tumor and lower doses to surrounding OARs
more!!!! with more fields you can conform to tumor better and find more angles to avoid OARs
a “hot spot” would indicate overdose, or underdose?
overdose
a “cold spot” would indicate overdose, or underdose?
underdose
Concept Check! - For parallel opposed fields, the fields must be parallel… so if they are NOT fully opposing… what would you check/fix?
check to see if between the 2 parallel opposed fields the collimator was flipped OR if the jaws were flipped/reversed
–one or the other, not both collimator AND jaws
if both the collimator and jaws were flipped for your POPs, is this good or bad? and Why
bad - only one can flip/be reversed… because if BOTH the collimator and jaw were flipped between fields, it will be like nothing changed on the opposite field
if using blocks for POPs what must you do between fields?
FLIP THEM
–tbt to sim lab!
the following is a POP field, can we treat if it looks like this?
NO - the POPs are not fully opposed
for POPs, the gantry is the exact _____ between the 2 fields, and field size is a ____ image
gantry is exact opposite (ie. parallel) and field size is a mirror image
why does the POP fields look like this? is it wrong or right?
because neither the collimator or field size were adjusted
in the following POP example - is the collimator or jaws adjusted?
collimator
in the following POP example - is the collimator or jaws adjusted?
jaws