Lab practical 1 Flashcards
Also called steroscopic microscopes
Dissecting microscopes
includes both scanning and transmission
electron microscopes
also called light microscopes
compound microscopes
are used to view the surface of relatively larger specimens at lower power typically no more than 10x,
dissecting microscopes
functions like elaborate magnifying glasses and would be used to observe the gross anatomy of a living, preserved, or dissected specimen at greater detail
dissecting microscopes
scientific instruments that use a beam of highly energetic electrons to examine objects on a very fine scale (up to several 100,000x
electron microscopes
What type of microscope do we use in lab
compound microscopes
Describe how the compound microscope works
compound microscopes are limited by the physics to 500x or 1000x magnification, it consists of two lenses and associated hardware that make viewing of specimen easier
the lower lens of the compound microscope is called
objective lens
the uppermost lens of the compound microscope is called
the ocular lens
the part through which the person looks through is
the uppermost lens which is the ocular lens
are mounted on a turret, allowing rapid changing
the lower lens which are the objective lens
name the power listed on the objective lenses scope
4x scanning power, 10x low power, 40x high power
hols the ocular and objective lenses in place
the body tube
most microbiological specimens are mounted on glass
slides
glass slides with cover slips are placed on
the stage
holds the slide firmly on the stage
usually clips or clamps
what two objects are located beneath the stage
a light source and a condenser lens
why is the compound microscope called light microscope
because it has a light source
why must the specimen be sliced thin
the specimen must be sliced then enough so that light can pass through the specimen to be viewed
focuses on light through a hole in the stage
condenser
what does the condenser include
an iris
varies the amount of light passing through a specimen
the iris
what happens to the light as it passes through the specimen
the goes through the objective and ocular lens and through the eye of the observer
this image is calculated by multiplying the magnification of the ocular by the magnification of the objective
the total magnification
what happens to light as it passes through any object (glass, air, specimens, etc.),
it bends
the bending of light is called
refraction
is a measurement of the extent that the substance bends light
the refractive index of a substance
distortion of the image
excessive refraction
at magnifications of less than 500x
the distortion is minimal
we will not view specimens at any total magnification greater than
400x
when magnifications are high what becomes the problem
the distortion becomes so great the image details are lost
helps to remedy the distortion problem by eliminating the air gaps between the specimen and the objective lens
oil immersion lens
the 4th objective lens is
the oil immersion lens
topmost series of lenses through which an object is viewed
ocular lens or eyepiece
what is the magnifying power of the ocular lens on your microscope
10
holds the nosepiece at one end and eyepiece at the other end; conducts light rays
body tube
supports upper parts and provides carrying handle
arm
objectives
objective lens, scanning, low, and high power
hold 4x lens used to view the whole slide
scanning power objective lens
Holds 10x lens used to view objects in greater detail
low-power objective
holds 40x lens used to view the objective in even greater detail
high- power objective
holds 100x lens and is used in conduction with oil to view objects with the greatest magnification
oil immersion objective lens
what lens are you never suppose to use
the oil immersion objective lens if your compound/light microscope has one
knob used to bring object into approximate focus; used only with low-power objective
coarse-adjustment knob
knob used to bring object into final focus
fine-adjustment knob
controls amount of illumination used to view the object
diaphragm or diaphragm control lever
an attached lamp that directs a beam of light up through the object
light source
the flat surface of the microscope that rests on the table
base
holds and supports microscope slides
stage
aids in the accurate position of the slide
mechanical stage
what structures are present to move the stage
the mechanical stage control knobs
should you ever use something other than the mechanical stage control knobs
NO, you should never use anything to move the stage it will cause damage to the stage
two knobs that are usually located below the stage. one knob controls forward/reverse and the other controls right/left
mechanical stage control knob
what is the first rule for microscope use
the lowest power objective should be in position booth at the beginning and end of microscope use
what is the second rule for microscope use
use only lens paper for cleaning the lens
what is the third rule for microscope use
do not tilt the microscope when viewing a wet mount
what is the fourth rule for microscope use
keep the stage clean and dry to prevent rust and corrosion
what is the fifth rule for microscope use
do not remove parts of the microscope
what is the sixth rule for microscope use
keep the microscope dust-free by covering it after use
what is the 7th rule for microscope use
report any malfunctioning when you notice it. This will benefit you and those who use the microscope after you
what is the should you always do before the 1st rule for focusing the microscope
always start with lowest power objective lens on you microscope or the scanning lens with the compound microscope
what is the first thing you do after making sure you at the lowest power objective lens when focusing the microscope
turn the nosepiece so that the lowest power lens is in straight alinement over the stage
what is the second rule of focusing the microscope
alway begin with focusing with lowest power objective
what is the 3rd rule of focusing the microscope
with the coarse-adjustment knob, lower the stage until it stops
what is the 4th rule of focusing the microscope
place slide on the stage and stabilize it with the clips
that is the 5th rule of focusing the microscope
again be sure the lowest power is in place
look grin the side
decrease the distance between the stage and the objective lens until the lens comes to an automatic stop or is no closer than 3mm above the slide
Why would it be a bad idea for the slide and objective lens to come into contact
you could break the slide or damage the lens
what is the 6th rule of focusing the microscope
while looking in to the eyepiece, rotate the diaphragm lever to give the maximum amount of light