QA/ QC Flashcards
what is the principal goal of a quality assurance program
to provide accurate and timely diagnosis 
what is the secondary goal of a quality assurance program
to minimize radiation exposure to the patient while the principal goal is achieved 
quality assurance
deals with people
quality control
delta switch instrumentation and equipment
* together they are designed to ensure the best possible image quality with the lowest possible exposure and minimum costs
QC tests are performed by
physicists
filtration beam quality test 
tested using a digital dosimeter half value layer measurement is required. performed annually.
- A filter is usually a piece of aluminum placed in the primary beam just as it exits the x-ray to before it reaches the collimator which helps lower patient skin dose.
- total filtration must be at least 2.5 mm Aluminum equivalent 
Collimation/light field alignment
must be accurate to within +/- 2% of the source to image receptor distance - -performed semi annually
positive beam limitation
within +/- 2%
central ray must be within
1%
focal spot testing
- slit camera ( most reliable)
- star test pattern
- pin hole camera
focal spot size can be determined by observing the regions of blurring 
spinning top test
spinning test is used to test timer accuracy a rectifier function in three phase equipment( older equipment)
exposure linearity test
they take MA stations next to each other and test them in a row.
they have to be within 10% with one another 
exposure reproducibility
The same exposure repeated. same mAs and Kvp repeated over and over
-5%
how do you test AEC?

use phantoms to simulate variations in patient thickness 
Quality control of digital equipment

- initial acceptance testing when you equipment is purchased it is tested to make sure it matches specifications also after a major repair!
- CR plate maintenance inspect and clean imaging plate regularly watch for dust! Plate needs to be erased every 24 to 48 hours.
- CR reader processor- calibrate annually,
- DR plate maintenance inspect and clean regularly detector calibration 
monitors
-they need to be dust free.
-Most monitors are flat panel, LCD screens that are designed to reduce the effect of ambient lighting monitors control the brightness of the image. 
lead checks
- performed annually to check for brakes and lead lining using image intensified fluoroscopy equipment as preferred.
- lead aprons should never be folded it will lead to cracks.
in what format must images being so that they can be sent through out the image viewing system?
DICOM
Makes it possible for equipment from different manufacturers to communicate with each other