Point-of-care Testing, Semi Automated & Automated Blood Cell Analysis Flashcards
The LeukoChek Blood Diluting System
for Manual White Blood Cell and
Platelet Counts Consists of a 20 µL capillary pipette and plastic reservoir containing__________ that makes a 1:100 dilution of whole blood.
1.98 mL of 1% buffered ammonium oxalate
In the HemoCue Hb 201 1 System for
Measuring Hemoglobin the hemoglobin is converted to ______ and is read photometrically at two wavelengths _______
570 nm and 880 nm
This method avoids the necessity of specimen dilution and interference from turbidity
The HemoCue Hb 201 1 System for
Measuring Hemoglobin
Eliminates the use of sealants. They have a factory-inserted plug that seals automatically
when the blood touches the plug.
SUREPREP Capillary Tubes
(Becton, Dickinson)
POCT microhematocrit centrifuge- based device
Hematastat II Separation Technology
Use the conductivity method to determine the hematocrit. Plasma conducts electrical
current, whereas WBCs act as insulators.
I-STAT Instrument for Measuring Hematocrit
Measures total hemoglobin by a spectrophotometric method.
AVOXimeter 1000E
Performs a CBC along with platelet aggregation.
Ichor Hematology Analyzer
Involves centrifugation in specialized capillary tubes designed to expand the buffy coat layer. The components (platelets, mononuclear cells, and granulocytes) can be measured with the assistance of fluorescent dyes and a measuring device.
Quantitative buffy coat analysis (QBC STAR,
manufactured by QBC Diagnostics, Inc., Philipsburg, PA)
TWO BASIC PRINCIPLES OF HEMATOLOGY
ANALYZERS:
● Electronic impedance (resistance)
● Optical scatter
Based on the detection and measurement
of changes in electrical resistance produced
by cells as they traverse a small aperture.
EELECTRONIC IMPEDANCE (RESISTANCE)
The data are plotted on a frequency
distribution graph, or volume distribution
histogram, with relative number on the y-
axis and volume (channel number equivalent
to a specific volume) on the x-axis.
ELECTRONIC IMPEDANCE (RESISTANCE)
Low-voltage DC impedance, as described
previously, may be used in conjunction with
RF resistance, or resistance to a high
voltage electromagnetic current flowing
between both electrodes simultaneously.
Radiofrequency (RF)
A hydrodynamically focused sample
stream is directed through a quartz flow
cell past a focused light source.
OPTICAL SCATTER
The light source of optical scatter
tungsten-halogen lamp or a helium-neon laser.