Chapter 11 Flashcards

1
Q

Measures must be taken to ensure that radiographic equipment operates safely to protect:

A

Patients
All personnel

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Every diagnostic imaging system must have a:

A

Protective tube housing
Correctly functioning control panel

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What type of tube housing is required to protect the patient

A

Lead-lined metal

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Lead-lined metal housing protects the patient and personnel from what type of radiation

A

Off-focus or leakage

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What is off-focus radiation

A

Radiation that is not useful - outside of the primary beam

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What is the maximum leakage radiation allowed at 1m from the x-ray source

A

1 mGy/hr (100 mR/hr)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Where is the control panel located?

A

Behind a suitable protective barrier that has a radiation-absorbent window

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Control panel must indicate the conditions of exposure and provide a ______ indication when the x-ray tube is energized

A

Positive

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Control panel must have visible ___ and ___ digital readouts

A

mA and kVp

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Radiographic examination table is frequently made of _______________ material

A

Carbon fiber

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

How is SID measured

A

From anode focal spot to the IR

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

SID distance and centering indicators must be accurate to within ___% and ___% of the SID, respectively

A

2% distance
1% centering

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

How is the x-ray beam limited to the size of the IR

A

Light-localizing variable-aperture rectangular collimator

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Types of x-ray beam limitation devices

A

Light-localizing variable-aperture rectangular collimator

Aperture diaphragms

Cones

Cylinders

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Benefit of restricting x-ray field size to include only the anatomic structures of clinical interest

A

Significant reduction in patient dose because less scatter radiation is produced

Improves the overall quality of the image

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Simplest of all beam limitation devices

A

Aperture diaphragms

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

What x-ray beam limitation device is used most in dental radiography

A

Cones

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

What types of cones are used for limitation devices

A

Flared metal tubes and straight cylinders

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Purpose of filtration

A

Removes low-energy photons (long wavelength or “soft” xrays) by absorbing them

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

Minimum total filtration required for radiographic units operating above 70 kVp

A

2.5mm Al eq

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Types of filtration used for mammographic equipment

A

Molybdenum and rhodium

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

The thickness of a designated absorber required to decrease the intensity of the primary beam by 50% of its initial value

A

Half-value layer

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Who measures the half-value layer to verify the x-ray beam is adequately filtered

A

Radiologic physicist

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

How often must the HVL be measured

A

Annually and after an x-ray tube is replaced or repairs have been made on the housing or collimation system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
HVL is expressed in:
mm of Al
26
HVL is a measure of beam:
Quality
27
Compensating filters are made of
Aluminium Lead-acrylic
28
Types of compensating filters
Wedge Trough
29
Consistency in output in radiation for identical generator settings from one indvidual exposure to subsequent exposures
Exposure reproducibility
30
What is the variance percentage for exposure reproducibility
5% or less
31
How is reproducibility verified
Using the same technical factors for repeated exposures and then observing with a calibrated ion chamber how intensity varies
32
Consistency in output radiation intensity at selected kVp settings when generator settings are changed from one mA and time combination to another
Exposure linearity
33
Ratio of the different in mR/mAs values between two successive generator settings to the sum of those mR/mAs values
Linearity
34
What must the value be of linearity
Less than .1
35
When settings are changed from one mA to a neighboring mA station, the most that linearity can vary is:
10%
36
Intensifying screens convert xray energy into ______ to produce radiographic density on the film
Visible light
37
Appox ___% of the radiographic density of the recorded image results from the visible light photons that are emitted from the intensifying screens
95%
38
A single x-ray photon can produce how many light photons with an intensifying screen
80 to 95
39
Intensifying screens enhances the film exposure process and __________ exposure time
Reduces
40
Rare-earth screen phosphors
Gadolinium Lanthanum Yttrium
41
What are the speeds for rare-earth screens and matching film combinations
200-1200
42
What screen-film speeds are currently standard for general radiography
400 speed
43
When the speed of the screen-film system doubles, radiation exposure is reduced by
50%
44
As kV increases, effective screen speed __________ for rare-earth screens, which _________ patient dose
Increases Reduces
45
What is used as a front material in screen-film
Carbon
46
Carbon fiber absorbs approximately ____ as much radiation as previously used materials
Half
47
What is the result of using asymmetric film emulsion and intensifying screen combinations
Image with greater uniformity and a decrease in patient dose
48
A device made of parallel radiopaque strips alternately separated with low-attenuation material
Grid
49
What are the strips made of in a grid
Aluminum, plastic or wood
50
Minimum skin-to-source distance for mobile radiography
30cm (12 inches)
51
Distance generally used for mobile radiography is
100cm - 120cm (40-48 inches)
52
With _________ SSD there is a more uniform distribution of exposure throughout the patient
Increased
53
CR produces what type of image
Analog
54
DR produces what type of image
Digital
55
How are x-rays turned into electrical signals in indirect conversions using thin-film transistor arrays
Scintillator Photodiode TFT array Electrical signals
56
How are x-rays turned into electrical signals in indirect conversions using CCD
Scintillator CCD Electrical signal
57
How are x-rays turned into electrical signals in direct converions
Photoconductor TFT array Electrical signal
58
How are repeat rates in DR effected
Eliminates the need for retakes due to improper technical selection Does not effect reports for positoning
59
CR contains what type of phosphor
Photostimulable
60
What type of laser beam is used to read the PSP plate
Helium-neon laser
61
Sensitivity of the phosphor used in CR is approximately equal to a ______ - speed screen-film combination
200
62
CR has a ______ kV flexibility than does conventional screen-film radiography
Greater
63
CR is _________ sensitive to scatter so a grid should be used more frequently
More
64
What type of imaging has the greatest patient radiation exposure rate in diagnostic radiology
Fluoroscopy
65
Benefits of image intensification in Fluoroscopy
Increased image brightness Saving of time for the radiologist Patient dose reduction
66
What is the kVp range for fluoroscopic procedures on adults
75-110
67
What is the SSD for stationary fluoroscopes
Not less than 38cm (15 inches)
68
In fluoroscopic procedures, how is kVp changes for children
Decrease by as much as 25%
69
How to minimize exposure to children
Decrease technical factors Maintain SSD Minimize the height of the image intensifier entrance surface above the patient
70
What is the HVL for fluoroscopy units when the kVp is between 80-100
3-4.5mm aluminum
71
What is the maxium exposure rate limitation for general purpose fluorscopy
100 mGy per minute (10 R/min)
72
What is the maximum exposure rate for high-level control fluorscopic units
200 mGy per minute (20 R/min)
73
What is the primary protective barrier required for a fluoroscopic unit
2-mm lead
74
Fluoroscopic exposure control switch must be what type
Dead-man
75
For C-arms, source-to-end of collimator assembly distance is required to a minimum of
30cm (12 inches)
76
For C-arms, patient-image intensifier distance should be as ______ as possible to reduce entrance does
Short
77
Dose reduction methods when using fluoro
Pulsed imaging Last-image-hold
78
FDA has recommended that a notation be placed in the patient's record if a skin dose is in the rank of:
1 to 2 Gy
79
Procedures involving extended fluoroscopic time
Percutaneous transluminal angioplasty Radiofrequency cardiac catheter ablation Vascular embolization Stent and filter placement Thrombolytic and fibrinolytic procedures Percutaneous thanshepatic cholangiopancreatography Transjugular intrahepatic portosystemic shunt Percutaneous nephrostomy Biliary drainage Urinary or biliary stone removal