Chapter 11 Flashcards
Measures must be taken to ensure that radiographic equipment operates safely to protect:
Patients
All personnel
Every diagnostic imaging system must have a:
Protective tube housing
Correctly functioning control panel
What type of tube housing is required to protect the patient
Lead-lined metal
Lead-lined metal housing protects the patient and personnel from what type of radiation
Off-focus or leakage
What is off-focus radiation
Radiation that is not useful - outside of the primary beam
What is the maximum leakage radiation allowed at 1m from the x-ray source
1 mGy/hr (100 mR/hr)
Where is the control panel located?
Behind a suitable protective barrier that has a radiation-absorbent window
Control panel must indicate the conditions of exposure and provide a ______ indication when the x-ray tube is energized
Positive
Control panel must have visible ___ and ___ digital readouts
mA and kVp
Radiographic examination table is frequently made of _______________ material
Carbon fiber
How is SID measured
From anode focal spot to the IR
SID distance and centering indicators must be accurate to within ___% and ___% of the SID, respectively
2% distance
1% centering
How is the x-ray beam limited to the size of the IR
Light-localizing variable-aperture rectangular collimator
Types of x-ray beam limitation devices
Light-localizing variable-aperture rectangular collimator
Aperture diaphragms
Cones
Cylinders
Benefit of restricting x-ray field size to include only the anatomic structures of clinical interest
Significant reduction in patient dose because less scatter radiation is produced
Improves the overall quality of the image
Simplest of all beam limitation devices
Aperture diaphragms
What x-ray beam limitation device is used most in dental radiography
Cones
What types of cones are used for limitation devices
Flared metal tubes and straight cylinders
Purpose of filtration
Removes low-energy photons (long wavelength or “soft” xrays) by absorbing them
Minimum total filtration required for radiographic units operating above 70 kVp
2.5mm Al eq
Types of filtration used for mammographic equipment
Molybdenum and rhodium
The thickness of a designated absorber required to decrease the intensity of the primary beam by 50% of its initial value
Half-value layer
Who measures the half-value layer to verify the x-ray beam is adequately filtered
Radiologic physicist
How often must the HVL be measured
Annually and after an x-ray tube is replaced or repairs have been made on the housing or collimation system
HVL is expressed in:
mm of Al
HVL is a measure of beam:
Quality
Compensating filters are made of
Aluminium
Lead-acrylic
Types of compensating filters
Wedge
Trough
Consistency in output in radiation for identical generator settings from one indvidual exposure to subsequent exposures
Exposure reproducibility
What is the variance percentage for exposure reproducibility
5% or less
How is reproducibility verified
Using the same technical factors for repeated exposures and then observing with a calibrated ion chamber how intensity varies