AEC TEST Flashcards
The primary purpose of AEC is to:
control the amount of radiation reaching the IR
AEC controls the amount of radiation reaching the IR by controlling the:
time of exposure
When using AEC , the radiographer is responsible for determining:
kVp
The predetermined level of radiation exposure needed to terminate the exposure when using AEC is set by:
the people who service the equipment
The two types of detectors used in AEC are:
phototimers and ionization chambers
The original type of AEC detector was the:
phototimer
Today’s AEC system is most likely to use a/an:
ionization chamber
There is typically a configuration of at least ___ AEC detectors in place.
3
An ionization chamber is a hollow device that contains:
air
Once the electric charge from an AEC detector reaches a preset level:
the exposure will stop
The shortest exposure time that a radiographic tube with AEC can have is the:
minimum response time
The maximum length of time that an exposure lasts when using AEC is the:
back-up time
The actual amount of exposure used for an image when AEC is being used is the:
mAs readout
These can be adjusted to increase or decrease the amount of radiation needed to terminate the exposure using AEC.
density controls
When using AEC, the kVp:
is set as appropriate for the study
Using a higher kVp during an AEC examination results in:
a shorter exposure time
The AEC minimum response time may post a problem when:
the patient is unable to hold still for the exposure
If the back-up time is set by the equipment, the exposure time should end when the mAs reaches:
600