AEC TEST Flashcards
The primary purpose of AEC is to:
control the amount of radiation reaching the IR
AEC controls the amount of radiation reaching the IR by controlling the:
time of exposure
When using AEC , the radiographer is responsible for determining:
kVp
The predetermined level of radiation exposure needed to terminate the exposure when using AEC is set by:
the people who service the equipment
The two types of detectors used in AEC are:
phototimers and ionization chambers
The original type of AEC detector was the:
phototimer
Today’s AEC system is most likely to use a/an:
ionization chamber
There is typically a configuration of at least ___ AEC detectors in place.
3
An ionization chamber is a hollow device that contains:
air
Once the electric charge from an AEC detector reaches a preset level:
the exposure will stop
The shortest exposure time that a radiographic tube with AEC can have is the:
minimum response time
The maximum length of time that an exposure lasts when using AEC is the:
back-up time
The actual amount of exposure used for an image when AEC is being used is the:
mAs readout
These can be adjusted to increase or decrease the amount of radiation needed to terminate the exposure using AEC.
density controls
When using AEC, the kVp:
is set as appropriate for the study
Using a higher kVp during an AEC examination results in:
a shorter exposure time
The AEC minimum response time may post a problem when:
the patient is unable to hold still for the exposure
If the back-up time is set by the equipment, the exposure time should end when the mAs reaches:
600
When the radiographer has the opportunity to set the back-up time, it should be approximately _____ of the expected exposure time.
150%
If the back-up time is shorter than the actual exposure needed for a properly exposed image:
the image receptor will be underexposed
The detector(s) selected for an image should be:
the ones that are superimposed by the anatomic structure of interest
Which of the following is the most critical aspect of successfully performing an examination using AEC?
centering the anatomy of interest over the selected detector(s)
If the patient for an AP thoracic spine is centered so that the spine is not over the top of the detector:
the film image will be too light
If the AEC detectors are not completely covered by the part being imaged and are exposed to primary radiation:
the exposure will terminate very quickly