Lab - BUN & Serum Creatinine Flashcards
Urea + water --> \_\_\_, c/o urease. A. ammonia and carbon monoxide B. ammonia and carbon dioxide C. methane and nitrogen dioxide D. NOTA
B
True about Blood Urea Nitrogen determination
A. In a modified Berthelot reaction, the ammonium ions react w/ hypochlorite and salicylate to give a green dye
B. In a modified Camelot reaction, the ammonium ions react w/ hypochlorite and salicylate to give a blue dye
C. In a modified Berthelot reaction, the ammonium ions react w/ sodium chloride and salicylate to give a blue dye
D. blood serum is used as specimen here
A
In BUN Det, increase of absorbance at \_\_\_ is proportional to urea concentration of sample A. 405 nm B. 350 nm C. 578 nm D. 646 nm
C
Not part of reagent 1 A.sodium salicylate B.sodium hydroxide C.sodium nitroprusside D.sodium hypochlorite
B and D = they are part of reagent 2
True about BUN determination
A.After Reagent 2: Mix and incubate for 5 mins at 20-25 deg C OR 3 mins at 37 deg C
B.After Reagent 1: Mix, incubate for 10 mins at 20-25 deg C OR 5 mins at 37 deg C.
C.Absorbance must be read within 60 minutes
C
baliktad ang A and B
Which is false?
A. All anticoagulants except EDTA can be used
B. The standard contains sodium azide (0.1%). Don’t swallow. Avoid contact w/ skin & mucous membranes
C. Reagent 2 contains sodium hydroxide and hypochlorite which are irritants.
A = ammonium heparin dapat
BUN test is not influenced by A. hemoglobin values up to 200 mg/dl B. bilirubin values up to 10 mg/dL C. BOTA D. NOTA
C
in serum creatinine test: creatinine forms in alkaline solution an orange-colored complex with picric acid.
What reaction is this?
kinetic Jaffe reaction
Which is not used as specimen for serum creatinine? A. serum B. urine C. heparinized plasma D. whole blood
D
Wavelength at which serum creatinine test is conducted: identify
492 nm
Reference range of serum BUN
6-20mg/dL
reference range for serum creatinine, male
Male: 0.6 - 1.1 mg/dL
reference range for serum creatinine, female
female: 0.5 - 0.9 mg/dL
false about serum creatinine det
A. Reaction is not sensitive to temperature.
B. PIC is harmful when inhaled, swallowed, or in contact w/ skin.
C. Assay can be affected by presence of reducing compounds
A - very sensitive dapat
Mix! After 30 sec, read abs (A1)! Read abs (A2) exactly after 2 mins!
In what test is this applicable?
serum creatinine