FAULTS IN TISSUE PROCESSING AND SECTIONING Flashcards
reason of brittle and hard tissue
prolonged fixation
prolonged dehydration
prolonged clearing
prolonged paraffin infiltration in overheated paraffin oven
drying out of tissue before actual fixation
failure to give special handling of tissues when treated with routine processing methods
remedy in brittle and hard tissue
Tissues may be softened by soaking in a small dish or bowl containing water with detergent, phenol or Molliflex.
causes of clearing agent turns milky as soon as tissue is placed on it
water not completely removed due to incomplete dehydration
remedy for the clearing agent that turns milky as soon as tissue is placed in it
Repeat dehydration with absolute alcohol, then clear again
give the reason
On trimming, tissue smells of clearing agent
Clearing agent not completely removed due to insufficient impregnation
give the remedy
tissue smells of clearing agent
Block is trimmed down nearest to the tissue, remaining wax is melted on embedding oven and paraffin impregnation is repeated, changing the paraffin at least once before blocking
give the reason
Tissue is opaque, section–cutting is difficult due to presence of alcohol
Insufficient clearing
give the remedy
Tissue is opaque, section–cutting is difficult due to presence of alcohol
Repeat clearing; if object has already been embedded, prolong clearing up to 12 hours then re–embed
reason for
Tissue shrinks away from wax when trimmed
Insufficient dehydration, therefore incomplete clearing and impregnation
remedy
Tissue shrinks away from wax when trimmed
repeat the whole procedure
reason
Tissue is soft when block is trimmed
incomplete fixation
remedy for
Tissue is soft when block is trimmed \
repeat fixation
reason for
Airholes found on tissue during trimming
incomplete impregnation
remedy for
Airholes found on tissue during trimming
repeat impregnation
reason for
On trimming, wax appears crystalline
contaminated wax
remedy for
On trimming, wax appears crystalline
re-embed in freshly filtered wax
reason for
Paraffin block after cooling is moist and crumbles
insufficient impregnation
remedy for
Paraffin block after cooling is moist and crumbles
repeat paraffin impregnation then re-embed
what are the faults in tissue processing
brittle or hard tissue
clearing agent turns milky as soon as tissue is placed in it
on trimming, tissue smells clearing agent
tissue is opaque, section-cutting is difficult due to presence of alcohol
tissue shrinks away from wax when trimmed
tissue is soft when block is trimmed
airholes found on tissue during trimming
on trimming, wax appears crystalline
paraffin block after cooling is moist and crumbles
reason
sections are too thick
wrong micrometer setting
remedy if the sections are too thick
microtome needs recalibration
give the reason
sections fall to form ribbons
surfaces and edges of the block are not parallel
horizontal surface of the block is not parallel to the knife
paraffin wax is too hard
knife is tilted too much
sections are too thick
knife is dull
give the reason if the sections roll up on cutting so that they adhere and get broken against the knife edge
knife is blunt
tilt of knife is too great
knife edge is dirty
remedy if the sections roll up on cutting so that they adhere and get broken against the knife edge
sharpen the knife
reduce the tilt
clean the knife edge
sections fall to form ribbons
reason: Surfaces and edges of the block are not parallel
remedy:
re trim the block
sections fall to form ribbons
reason: horizontal surface of the block is not parallel to the knife
remedy:
re adjust and re orient the block
sections fall to form ribbons
reason: paraffin wax is too hard
remedy:
coat horizontal edges of the block with wax of lower melting point
sections fall to form ribbons
reason: knife is tilted too much
remedy:
reduce the tilt