Threshold Testing Flashcards
1
Q
What is the minimum acceptable acute value for atrial sensing?
A
≥ 1.5 mV
2
Q
What is the minimum acceptable acute value for ventricular sensing?
A
≥ 5.0 mV
3
Q
What is the minimum acceptable acute value for atrial capture threshold?
A
≤ 1.5V @ 0.5ms
4
Q
What is the minimum acceptable acute value for ventricular capture threshold?
A
≤ 1.0V @ 0.5ms