MICRA TPS Flashcards
Is the Micra TPS MRI compatible?
yes, for 1.5T and 3T full body MRIs
What is the longevity (battery life) of the Micra TPS?
approximately 12 years
What is the minimum acceptable battery voltage require at implant of a Micra TPS?
at least 3.0 mV
What are the steps to the pre-op pre-program of the Micra TPS?
- interrogate to ensure battery voltage is at least 3.0mV
- switch mode to VVI
- set the date/time
- program pacing values
- switch to Device Off mode before implanting
- click on end session
What needs to be done to the introducer of a Micra TPS prior to use?
wet with NS to activate the hydrophilic coating
How many tines are required to be engaged or in position prior to conducting a pull and hold test of a Micra TPS?
at least 2 tines need to be engaged
What is the recommended R-wave amplitude for a Micra TPS at implant?
5mV ≤
What is the recommended impedance for a Micra TPS at implant?
400 - 1500 ohms
What is the recommended threshold for a Micra TPS at implant?
< 1.0 V
What mode is set when using a Micra TPS?
VVI
What are the 2 methods for conducting a threshold test on a Micra TPS?
- Capture Management (CM) (automated process)
- Amplitude decrement test (manual)
What is the first step in reposition a Micra TPS if unacceptable values result?
turn the device off
What solution is used to flush the delivery system of a Micra TPS prior to removing the tether (strings)?
heparinized saline to remove any formed blood clots
What should be done if resistance if felt when confirming free movement of the tether prior removal following deployment of a Micra TPS?
flush the delivery system with heparinized saline
What technique is used when inserted/placing a Micra TPS?
- Seldinger Technique
- use of a guidewire and dilation cannula to place a catheter (same used for central line placement)
What feature adjusts the patient’s rate response to ensure that it remains appropriate for the full range of patient activities?
Rate Profile Optimization (RPO)
What is does the Rate Profile Optimization (RPO) feature do when programmed on a Micra TPS?
automatically adjusts the patient’s rate response to ensure that it remains appropriate for the full range of patient activities
How does the accelerometer in the Micra TPS tell the difference between the normal resting heart motion and the higher levels of motion that occur during exercise?
- Lower Rate Setpoint feature sets a unique baseline to differentiate normal heart motion at rest versus that occurring with increased activity
What type of accelerometer is contained in the Micra TPS and why?
- a 3 axis accelerometer compared to a single axis in traditional IPG
- to account for the different angle positions the device may be in at implant
When is the Activity Vector Test (AVT) performed after a Micra implant?
- prior to discharge after placement
- when there is concern about adequate rate response
The Micra TPS should be in what mode when performing an Activity Vector Test?
VVIR
What are the main purposes of the Activity Vector Test?
- to set the rate response setpoints
- to determine the activity vector (1 or 3) where rate response performs the best
What should the lower rate setpoint be set to when manually programming setpoints on a Micra TPS following an Activity Vector Test?
set LR to 2 counts above the highest peak counts while at rest
What should the activities of daily living (ADL) setpoint be set to when performing manual programming on a Micra TPS following an Activity Vector Test?
set ADL setpoint to the value of the highest peaks while walking