ICD Inductions and Programming Flashcards

1
Q

Do Medtronic ICD’s have programmable tilt for ventricular therapies?

A
  • No

- have fixed tilt 50/50

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What acronym is used when performing an ICD device functional integrity test prior to implantation?

A

VTOP:

  • Voltage
  • date and Time
  • verify detections of Off
  • Print
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What is the minimum battery voltage allowed when implanting an ICD?

A

voltage must be > 2.97V

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What are the 2 methods for inducing VF during a DFT?

A

T-shock and 50Hz burst

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What do the following T-Shock Parameters represent?

1) #S1
2) S1S1
3) delay

A

1) number of pacing pulses delivered in the pacing train
2) cycle length (rate) pacing train is delivered
3) time between the last pacing pulse and the shock

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What are the sequence of events seen on an EGM when performing a DFT?

A

1) pacing train
2) shock on T-wave following delay interval
3) FS sensing
4) FD detected
5) charging of capacitor
6) shock delivered
7) NS or RNID

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

What are the programmable T-Shock values used for induction during a DFT?

A
  • number of pulses in the pacing train (#S1)
  • cycle length or rate of the pacing train
  • length of delay between the last pulse and shock
  • energy level
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What are the programmable 50Hz Burst values used for induction during a DFT?

A
  • chamber
  • amplitude
  • pulse width
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What is the maximum amount of time the device will allow delivery of a 50Hz Burst used for induction during a DFT?

A

burst times out after 10 seconds regardless of how long the button is pressed

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What does the pneumonic “STD” stand for in relation to a DFT?

A
  • Sensitivity (set RV to least sensitive - 1.20mV)
  • Therapies (start low, 15-25J)
  • Detections (set to lowest NID and RNID)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What does the pneumonic REPD stand for in relation to a DFT?

A
  • Resume at delivery (want checked)
  • Enable (want checked)
  • Print
  • Deliver shock
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What should the RV sensitivity be set to prior to performing a DFT?

A

1.2mV or the least sensitive setting

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What should the VF therapy energy be set to after performing a DFT?

A
  • add 10 Joules to the DFT rescue energy
How well did you know this?
1
Not at all
2
3
4
5
Perfectly