LINQ ICMs Flashcards

1
Q

What is the longevity (battery life) for a Reveal LINQ ICM?

A

3 years

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What is the longevity (battery life) for a LINQ II ICM?

A

4.5 years

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What are the indications for an Internal Cardiac Monitor (ICM)?

A
  • patients at increased risk of arrythmias

- transient symptoms that may suggest presence of a cardiac arrythmia

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Are the Reveal LINQ and LINQ II ICMs MRI compatible?

A

yes, they are MRI Conditional at 1.5 and 3T with no post-insertion waiting period

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Which programmer is needed to interrogate/set-up the Reveal LINQ ICM?

A
  • 2090 programmer or tablet with the Reveal LINQ manager app
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Which programmer is needed to interrogate/set-up the LINQ II ICM?

A

tablet with the Reveal LINQ manager app

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

What is the number 1 position recommended for ICM placement?

A

4th ICS at a 45 degree angle towards the nipple, left of the sternum

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What is the number 2 position recommended for ICM placement?

A

4th ICS parallel to the left sternal border

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What is the number 3 position recommended for ICM placement?

A

5th ICS perpendicular to the sternum on the left side of the chest

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What is the minimum acceptable measurement for R-waves when using an ICM?

A

0.2mV or greater

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What should be done if the R-wave measures less than 0.2mV when implanting an ICM?

A
  • remove device and reposition at an angle 22 degrees toward the sternum using the same pocket
  • if still not 0.2mV or greater, remove and insert using the 2nd recommend position
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What are the 3 abnormal sensing issues that may occur at implant of an ICM?

A
  • inadequate R-Wave amplitude (< 2.0mV)
  • undersensing
  • oversensing
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What are the steps to correct an inadequate R-Wave amplitude (<0.2mV) at implant of an ICM?

A
  • reposition device is best option

- if unable to reposition, increase R-Wave sensitivity by reducing the sensing value (lowers the fence)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What ICM complications may occur when undersensing occurs?

A
  • inappropriate/false detection of bradycardia or asystole
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What are the steps to correct oversensing of noise at implant of an ICM?

A
  • reposition the device
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What are the steps to correct T-wave oversensing at implant of a Reveal LINQ ICM?

A
  • reprogram sensing threshold decay delay
17
Q

What are the steps to correct T-wave oversensing at implant of a LINQ II ICM?

A
  • reprogram T-wave blanking interval
18
Q

What are the steps to correct P-wave oversensing at implant of an ICM?

A
  • reposition the device

- decrease the R-wave sensitivity (increase sensing value)

19
Q

What are the types of abnormal sensing that can occur at implant of an ICM?

A
  • undersensing
  • inadequate R-wave amplitude (<2.0mV)
  • T-wave oversensing
  • P-wave oversensing