MRI Flashcards
What are the hazards to the device related to the Static magnetic fields?
- force and torque (pt discomfort and dislodgement)
- vibration (pt discomfort and device damage)
- device interactions (may inhibit/cause therapy)
What are the hazards to the device related to the Gradient magnetic fields?
- unintended stimulation (arrhythmia induction)
- vibration (pt discomfort and device damage)
- case heating (pt discomfort, necrosis)
- device interactions (may inhibit/cause therapy)
What are the hazards to the device related to the RF magnetic fields?
- unintended stimulation (arrhythmia induction)
- case heating (pt discomfort, necrosis)
- device interactions (may inhibit/cause therapy)
- lead electrode heating
Which type(s) of magnetic fields can cause unintended stimulation (arrhythmia induction)?
gradient and RF
Which type(s) of magnetic fields can cause vibrations leading to patient discomfort and possible device damage?
static and gradient
Which type(s) of magnetic fields can cause the case to heat resulting in possible patient discomfort and tissue necrosis?
gradient and RF
Which type(s) of magnetic fields can cause lead electrode heating that may negatively impact the delivery of therapy and proper sensing/threshold capture?
RF
Which label given to items that pose no known hazards resulting from exposure to any MRI environment?
MR Safe
Which label is given to items which pose unacceptable risks to the patient, medical staff, or other persons within the MRI environment?
MR Unsafe
Which label is given to items with demonstrated safety in the MRI environment within defined conditions?
MR Conditional
What device settings/values will prevent MRI SureScan mode from being turned on?
- if pacing impedances are < 200 or 1500 <
- if defib lead impedances are < 20 or 200 <
- if RRT/ERI/EOS has been reached
Which pin plug is approved as MR Conditional with the CRT MRI system?
the 6725 atrial pin plug
What is the MRI SureScan timeout value for all ICD’s and CRT systems?
6hrs
What is the MRI SureScan timeout value for all pacemaker systems?
24hrs
What is the recommended RV pacing capture threshold value for PM-dependent patients going for an MRI?
< 2.0V at 0.4ms