POC machines for Blood Test 2 Flashcards
Hemochron Response ?
POC Test: Sample Size: Mech of detection: QC: Data Storage: Activator:
ACT,PT,PTT,HRT,PRT,TT,HiTT,Fibrinogen 2 mL in ICT Tube Magnet moves until it clots EQC & Plasma Internal Memory 600 pts / 64QC Silica
Hemochron Jr. Signature & GEM PCL ?
POC Test: Sample Size: Mech of detection: QC: Data Storage: Activator:
ACT+, ACT-LR, PT/PTT 1 drop on card Sample well >Channel >Activator EQC & Plasma Internal Memory 234pts / 85QC Silica & Kaolin
Helena Actalyke ?
POC Test: Sample Size: Mech of detection: QC: Data Storage: Activator:
Max-ACT, HDR, HPT
0.5 mL in ICT Tube. can use homochron response tubes
Magnet moves until it clots / 2 point detection
EQC & Plasma
No internal memory
Silica+Kaolin+Glass = maxACT
Hemochron Jr. Signature
&
Gem PCL
POC Test: Sample Size: Mech of detection: QC: Data Storage: Activator:
ACT + , ACT-LR, PT/PTT
1 Drop
Blood placed on sample well, > sample channel > Constriction > Mixed with Activator
EQC & Plasma
- Internal memory (234 pts & 85 QC)
Silica & Kaolin
Hepcon HMS (Heparin Management System)
POC Test: Sample Size: Mech of detection: QC: Data Storage: Activator:
HR-ACT, LR-ACT,Heparin assay HDR/HPT, PFT
1.5 mL
Sample cartriges come in 2, 4, or 6 wells loaded with different levels of heparin. Syringe of blood is placed into the machine > blood is distributed into all the wells. The machine lifts the black portion of the well every 5 seconds until a clot is formed
EQC & Plasma QC
- No internal memory * No Software analysis
Kaolin
What is a good thing about the upgraded version of Hepcon HMS Plus ?
• Internal memory • Operator lock out
ACT II & ACT Plus
POC Test: Sample Size: Mech of detection: QC: Data Storage: Activator:
HR-ACT, LR-ACT, RACT, PT/PTT, HTC,
0.3 mL / 1 Drop
Infrared light beam through the cartridge detects how long it takes for the black plunger / daisy star to fall through the reaction chamber into the reagent chamber.
EQC & Plasma QC
500 point Internal memory
Celite and CaCl
Bayer/Chiron RapidPoint Coag AKA: TAS
(Throbolytic Assessment System)
POC Test: Sample Size: Mech of detection: QC: Data Storage: Activator:
ACT / PT / PTT
0.3 mL / 1 Drop
Credit card style sample cartrige with capillary action that draws the blood sample in through the sample well into the reaction chamber . Paramagnetic Iron Oxide Particles (PIOP) suspended in
liquid sample is influenced by the changing magnetic
forces below the test card
Coagulation prevents the PIOP from moving and the photodetector detects a change in the quantity of light
reflected off the back of the card
EQC & Plasma QC
Internal memory (1000 pts)
Celite
Sienco Sonoclot II
POC Test: Sample Size: Mech of detection: QC: Data Storage: Activator:
ACT & Clot rate
360 ul
The mechanism is a tubular probe that moves up and down within a blood sample As the sample progresses through various stages of clotting, the electronic circuitry detects increasing resistance.
Plasma & Oil QC
No internal memory
Silica
Heamoscope TEG
* Sensative to motion even from the building*
POC Test: Sample Size: Mech of detection: QC: Data Storage: Activator:
(Clotting Time) do not call it an ACT. Heparinase Test Platelet Function Clot Strength Fibrinolysis
A thin wire probe is used to measure, which the clot forms around. The speed and strength of clot formation is measured in various ways, typically by computer.
No internal memory
What is the most significant difference between the TEG and the ROTEM ?
The Sensor shaft rotates on the ROTEM rather than the rotational CUP with the TEG.