What filter do GE lunar systems use
Cerium
The cerium filter spectrum at 80KV produces 2 photo electric peaks at about what value
40keV and 70keV
Norland Swissray uses what type of filter
Samarium
The Samarium filter spectrum at 100kV a low energy peak and high energy peak at what range
47keV low, Less than 100keV high
Hologic scanners use what type of filter system
Energy switching xray beam between 100 and 140kVp
Two dimensional scan area is also known as
Projectional or areal
Three dimensional scans are also known as
Volumetric
A decrease in the BMC also results in a decrease in
BMD
A decrease in area causes an increase in
BMD
This is the sum of the data values, divided by the number of values
Mean
Measure of variability, it shows how spread out the data is around the mean
Standard deviation
This is A statistic that helps compare a variability between different datasets whether or not they have the same mean
Percent coefficient of variation
This refers to an inconvenient anomaly in which T scores from different atomic sites within the same patient fall into different diagnostic classifications
Discordance
This describes how closely a measurement approximates with things that is measuring
Accuracy
This measures how closely a test`s results approximate the correct value
Trueness
This is the measure of how likely a test is to achieve the same result when repeated on the same subject in the same conditions
Precision
This is used to describe the consistency of measurement
Precision
This is used to describe how closely it measures the expected values
Trueness
What are the rules for in vivo scans
Either 15 patients 3 times each or 30 patients 2 times each
What is the maximum acceptable precision error for technologies performing a lumbar spine scan
1.9% with an associated LSC of 5.3%
What is the maximum precision error for a scan of the total hip
1.8% with an LSC of 5%
What is the maximum precision error for a scan of the femoral neck
2.5% with an LSC of 6.9%
How do you cross calibrate machines
Scan 30 patients representative of the facility’s patient population on the original system and then twice on the new system for this method all scan should be completed within 60 days
This is a measure of bone micro texture and is a marker for the risk of osteoprosis
Trabecular bone score