JVD 2015 #4 - Managing Endodontic Instrument Separation Flashcards
“Managing endodontic instrument separation” by McCoy,
what are the two most common causes of instrument separation?
instrument fatigue from overuse, applying excessive apical pressure (rotary motion);
separation from manufacturing defect is rare
“Managing endodontic instrument separation” by McCoy,
where are NiTi files designed to break?
at the shaft-thank jnctn under extreme force/torque, but it can occur anywhere along the file
“Managing endodontic instrument separation” by McCoy,
what should you do if an instrument breaks?
(cry)
take a radiograph, use loupes or surgical microscope for visualization, strong attempts to remove (file braiding, magnetized instruments, ultrasonic endo tips), enlarge access, owner informed, then consider switching to different option
“Managing endodontic instrument separation” by McCoy,
what if you can’t get the file out?
can obturate around it (?!) as long as cleaning, shaping is adequate with regular radiographs monitored, can convert to an apico, can extract
“Managing endodontic instrument separation” by McCoy,
how often should NiTi files be changed? what about path finders?
3-6 uses;
pathfinder only use once!