EAC Suction Equipment Flashcards
Indications for use of suction
Regurgitation in the unconscious pt
Excessive sputum
Blood from maxilla-facial or head injuries, epistaxis or abdominal haemorrhage
Excessive salivation in stroke patients
Injuries to pharynx or obstruction of oesophagus
Clearance of oro-pharyngeal airways
incorrect use of suction could lead to:
Soft tissue damage
Vagal stimulation
Removing oxygen from the airways (too long at a time)
how to run serviceability test on LSU
press and hold the test button while setting the operating knob to 500+mmHg
Keep test button pressed for two seconds
The test will start immediately
As soon as LED two lights up occlude the patient suction tubing
Keep occluded until LED one relights
capabilities of running times LSU
500mmHg is 25L/min and from full charge runs for 45min
1L capacity
cleaning LSU
Replace disposable items
Ensure contaminated parts are placed in yellow clinical waste bin
Do not dispose of canister holder or vacuum connector
Clean all external surfaces using LAS infection control cleaning systems
Do not stand or soak and parts of LSU in water
capabilities of Suction-Easy
fluid capacity 1L
vacuum force 100mmHg
suction catheters
Yankuer - hard/angled catheter
CH8 soft, flexible
CH18 soft flexible (larger)
cautions suctioning laryngectomy pt
Use catheter to suction the stoma for not more then 3-5secs inside
And no longer then 15secs at a time around the outside
Only remove the tube from a stoma if completely blocked and only as a last resort
correct suction technique
Only suction on the way out of the mouth
Suction in a figure of 8 motion
Suction for no longer than 10 seconds at a time
unconscious patients suctioned on their sides where possible
how to use:
Suction-Easy
Remove packaging
Pull out tube until stop/lock is felt.
Remove obvious solids too big for tube from pt’s airway
grasp bulb between thumb and forefingers… squeeze bulb
General Points
Always use aseptic technique
Reduce suction pressure when aspirating children
Don’t allow reservoir to overfill as aspirant will contaminate pump unit
Clean thoroughly after use
Perform function test at start of shift and check battery is in
when using on a pt who has taken poison try and retain a sample