Chapter 15: Energy Sources: Electrosurgery and Lasers Flashcards

1
Q

Define electrical voltage, current and resistance and how they relate in regards to Ohm’s law:

A
  • Volatage (V) = the difference in potential between two points. (The force that drives particles across a potential difference). Volts
  • Current (I) = Represents the flow of electons. Amperes
  • Resistance (R) = The impedence to the flow of the electrons. Ohms

Ohm’s Law: V = IR or 1 volt = 1 amp x 1 ohm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What is power?

A

Power is the work performed expressed in watts

W = VI =RI^2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What is the difference between electrosurgery and electrocautery?

A
  • Electrosurgery = Passing an electrical current from an instrument through tissues in order to achieve a desired result
  • Electrocautery - Using electrical current to heat a metal instrument first before applying it to a targeted area
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What are differences between radiowave surgery and electrosurgery?

A
  • Radiowave uses current in the higher frequency range (>4MHz as compared to 500kHz)
  • Less tissue penetration
  • Less lateral tissue damage
  • No ground plate required
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

How does pure cut mode on monopolar devices work in terms of waveform?

A
  • Current is delivered uninterrupted 100% of the time (and therefore V is low as W=VI)
  • Best cutting achieved with electrode held slightly away from tissue
  • Generator power settings 50-80W
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

How does coagulation mode work in terms of waveform?

A
  • Series of interrupted waves at a higher voltage than cut mode
  • Generator power setting 30-50W (spikes of V as high as 9000V can be produced at 50W)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

What percentage of the time is current available due interrupted coagulation waveforms?

A
  • Current is available 6% of the time to heat the tissue.
  • Remaining 94% of the time the tissue cools to produce a coagulation effect
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What are fulguration, desiccation, and coaptive coagulation?

A
  • Fulguration coagulation - holding the electrode slightly away from the tissue with a resulting arc and spark. Good if a discrete bleeder cannot be identified.
  • Desiccation Coagulation is achieved by direct contact of the electrode and tissue. Heats tissue more efficiently with deeper necrosis and thermal spread.
  • Coaptive coagulation is a type of desiccation coagulation which occurs when the lumen of a vessel is occluded by a metal instrument (Collagen weld)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What are some advantages / disadvantages of monopolar electrosurgery?

A

Advantages:
Commonly available
Hemostasis for vessels <2mm

Disadvantages:
- Does not function well in liquid
- Potential for burns at grounding pad
- Increase in histopath scores of wound inflammation
- Delays in wound healing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What are some advantages and dsadvantages of bipolar electrosurgery?

A

Advantages
- Much more precise
- Can be used in a wet environment
- Lower voltage and power setting
- Cosidered to be safer

Disadvantages
- Increased time for coagulation
- Charring and adherence can lead to tearing
- Limited use for dissection and ineffectual for cutting

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What are the main mechanisms of injury when using electrosurgery devices in minimally invasive surgery?

A
  • Insulation failure (reuse issue)
  • Direct coupling (touches another metal instrument)
  • Capacitive coupling - When 2 conductors or instruments are seperated by an insulator (air) and form stored energy
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What size vessels is the LigaSure approved for?

A
  • Blood vessels 7mm or smaller and tissue as big as fits in jaw
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

List benefits of the Harmonic Ultrasonic device over electrosurgical units:

A
  • Simultaneously cuts and coagulates tissue
  • Lower temperatures (50-100C vs 150-400C)
  • Minimal lateral thermal spread
  • Virtually no charring, desiccation or sticking
  • Reduces smoke and vapour

However, does not reliably seal vessels >3mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What does laser stand for?

A

Light Amplification by Stimulated Emission of Radiation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What are the characteristics of a laser beam?

A
  • Monochromic (photons of the same waveform)
  • Spatial coherence (parallel photons)
  • Temporal coherence (photons in phase with each other)
  • Collimated (travel in a linear direction)

These photons will have consistent energy released at the same wavelength, amplitude, frequency and time allowing it to be focused 100x better than ordinary light

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What is power density in regard to lasers?

A

Power density is diretly proportional to the power (W) that a laser can deliver and inversely proportional to the surface area the the beam strikes

PD = W/cm^2

17
Q

What is the focal length of most lasers

A

1-3mm

18
Q

What are the 4 basic laser-tissue interactions?

A
  • Absorption
  • Transmission
  • Scatter
  • Reflection
19
Q

What are the five forms of tissue injury formed by lasers?

A
  • Photothermal
  • Photochemical
  • Photomechanical
  • Photodynamic
  • Photoablation
20
Q

List some potential uses of a CO2 laser. What size vessels will it seal?

A
  • Oral surgery
  • Upper airway surgery
  • Staphylectomy
  • Perianal/rectal surgery

Seals vessels <0.6mm

21
Q

What form of laser is good for cutting bone? Why?

A

Er:YAG
It is absorbed by hydroxyapatite

22
Q

What are the safety considerations when using a laser?

A
  • Toxic smoke production - must have a suction evacuation device
  • Fire Hazard
  • Ocular hazard - wave-length specific safety goggles
  • Potential mutagenesis
23
Q

What is the only vessel sealing device that controls energy deposition at the electrode-tissue interface to spread <2mm?

A

Enseal
Can seal up to 7mm