CCS CMR 2021 Update Flashcards
1
Q
What should PPM be reprogrammed to during MRI for nor non-compatible device in a patient that is not dependent
A
Sense only mode (OSO / OVO / ODO)
2
Q
What magnet should be used in MRI nonconditional CIEDs?
A
1.5 T
3
Q
What are 2 absolute CIs to MRI?
A
- Fractures leads
- Presence of lead extenders or adapters
4
Q
Name 4 relative CIs
A
Leads implanted < 6 weeks before the planned CMR
Epicardial leads
Abandoned TV leads
Scan overlaps with the region of the CIED
5
Q
How to set PPM if dependent?
A
Asynchronous pacing
6
Q
ICD management?
A
Turn off Tachy therapies
7
Q
ILR?
A
Download data before scan