Analytic Techniques Flashcards
Which of the following is not necessary for obtaining the spectrum of a
compound from 190 to 500 nm?
a. Tungsten light source
b. Deuterium light source
c. Double-beam spectrophotometer
d. Quartz cuvettes
e. Photomultiplier
a. Tungsten light source
Stray light in a spectrophotometer places limits on
a. Upper range of linearity
b. Sensitivity
c. Photometric accuracy below 0.1 absorbance units
d. Ability to measure in the UV range
e. Use of a grating monochromator
a. Upper range of linearity
which of the following light sources is used in atomic absorption
spectrophotometry?
a. Hollow-cathode lamp
b. Xenon arc lamp
c. Tungsten light
d. Deuterium lamp
e. Laser
a. Hollow-cathode lamp
. Which of the following is true concerning fluorometry?
a. Fluorescence is an inherently more sensitive technique than absorption.
b. Emission wavelengths are always set at lower wavelengths than
excitation.
c. The detector is always placed at right angles to the excitation beam.
d. All compounds undergo fluorescence.
e. Fluorometers require special detectors.
a. Fluorescence is an inherently more sensitive technique than absorption.
Which of the following techniques has the highest potential sensitivity?
a. Chemiluminescence
b. Fluorescence
c. Turbidimetry
d. Nephelometry
e. Phosphorescence
a. Chemiluminescence
Which electrochemical assay measures current at fixed potential?
a. Amperometry
b. Anodic stripping voltammetry
c. Coulometry
d. Analysis with ISEs
e. Electrophoresis
a. Amperometry
Which of the following refers to the movement of buffer ions and solvent
relative to the fixed support?
a. Electroendosmosis
b. Isoelectric focusing
c. Iontophoresis
d. Zone electrophoresis
e. Plasmapheresis
a. Electroendosmosis
Reverse-phase liquid chromatography refers to
a. A polar mobile phase and nonpolar stationary phase
b. A nonpolar mobile phase and polar stationary phase
c. Distribution between two liquid phases
d. Size used to separate solutes instead of charge
e. Charge used to separate solutes instead of size
a. A polar mobile phase and nonpolar stationary phase
Which of the following is not an advantage of CE?
a. Multiple samples can be assayed simultaneously on one injection.
b. Very small sample size.
c. Rapid analysis.
d. Use of traditional detectors.
e. Cations, neutrals, and anions move in the same direction at different
rates.
a. Multiple samples can be assayed simultaneously on one injection.
Tandem mass spectrometers
a. Are two mass spectrometers placed in series with each other
b. Are two mass spectrometers placed in parallel with each other
c. Require use of a gas chromatograph
d. Require use of an electrospray interface
e. Do not require an ionization source
a. Are two mass spectrometers placed in series with each other
Which of the following is false concerning the principles of point-of-care
testing devices?
a. Devices do not require quality control testing.
b. They use principles that are identical to laboratory-based
instrumentation.
c. Biosensors have enabled miniaturization particularly amendable for
point-of-care testing.
d. Onboard microcomputers control instrument functions and data
reduction.
e. Whole blood analysis is the preferred specimen.
d. Onboard microcomputers control instrument functions and data
reduction.
Which is the most sensitive detector for spectrophotometry?
a. Photomultiplier.
b. Phototube.
c. Electron multiplier.
d. Photodiode array.
e. All are equally sensitive.
a. Photomultiplier.
Which of the following is Beer law?
a. A = ε × b × c
b. %T = I/I0 × 100
c. E = hv
d. e = ∆pH × 0.59 V
e. Osmolality = j × n × C
a. A = ε × b × c
Which of the following correctly ranks electromagnetic radiation from low
energy to high energy?
a. Microwaves, infrared, visible, UV, x-rays, gamma, cosmic
b. Cosmic, gamma, x-rays, UV, visible, infrared, microwaves
c. UV, visible, infrared, microwaves, x-rays, cosmic, gamma
d. UV, visible, infrared, cosmic, gamma, microwaves, x-rays
e. Visible, UV, infrared, cosmic, gamma, microwaves, x-rays
b. Cosmic, gamma, x-rays, UV, visible, infrared, microwaves
What is the purpose of the chopper in an atomic absorption
spectrophotometer?
a. Correct for the amount of light emitted by the flame
b. Correct for the fluctuating intensity of the light source
c. Correct for the fluctuating sensitivity of the detector
d. Correct for differences in the aspiration rate of the sample
e. Correct for the presence of stray light
a. Correct for the amount of light emitted by the flame