Multiple choice question (RT) Flashcards
CHAPTER 1
Which of the following lamps provides a continuous spectrum of radiant energy in the visible, near IR, and near UV regions of the spectrum:
- tungsten-filament
- hydrogen
- deuterium
- mercury vapor
1 - tungsten-filament
Which of the following isolates light within a narrow region of the spectrum:
- photomultiplier tube
- monochromator
- photovoltaic cell
- detector
2 - monochromator
Which of the following is not descriptive of a photomultiplier tube:
- emits electrons proportionally to initial light absorbed
- must be shielded from stray light
- cannot be used with a chopper
- amplifies the initial signal received
3 - cannot be used with a chopper
Which of the following is false about a photomultiplier tube:
- converts radiant energy (light) to electrical energy (current)
- amplifies the current significantly
- has a very rapid response time
- is composed of an iron plate and a layer of selenium
4 - is composed of an iron plate and a layer of selenium
Which type of photodetector employs a linear arrngement that allows it to respond to a specific wavelength resulting in complete UV/visible spectrum analysis:
- photomultiplier tube
- phototube
- barrier layer cell
- photodiode array
4 - photodiode array
When performing spectrophotometer quality assurance checks, what is the holmium oxide glass filter used to assess:
- linearity
- stray light
- absorbance accuracy
- wavelength accuracy
4 - wavelength accuracy
In spectrophotometric analysis, what is the purpose of the reagent blank:
- correct for interfering chromogens
- correct for lipemia
- correct for protein
- correct for color contribution of the reagents
4 - correct for color contribution of the reagents
In regard to bichromatic analysis, which of the following is false:
- absorbance is measured at the spectral absorbance peak for a blank and the sample using the same wavelength
- eliminates background interferences
- sample concentration determined from difference in two measured absorbances
- functions as a reference blank for each sample
1 - absorbance is measured at the spectral absorbance peak for a blank and the sample using the same wavelength
The bandpass of a spectrophotometer is 10nm. If an instrument is set at 540 nm, the wavelength that are permitted to impinge on the sample will be within what wavelength range:
- 530-540 nm
- 530-550 nm
- 535-545 nm
- 540-550 nm
3 - 535-545 nm
Which of the following is not a problem inherent in turbidemetry:
- variation in particle size of samples
- variation in particle size of standards
- rate of aggregation or settling of particles
- need to maintain a constant and specific temperature
4 - need to maintain a constant and specific temperature
Which of the following may be associated with reflectance spectrophotometry as it relates to the dry reagent slide technique:
- light projected to the slide at 180 degree angle
- dye concentration directly proportional to reflectance
- unabsorbed, reflected light detected by photodetector
- reflectance values are linearly proportional to transmission values
3 - unabsorbed, reflected light detected by photodetector
Fluorometers are designed so that the path of the exciting light is at a right angler to the path of the emitted light. What is the purporse of this design:
- prevent loss of emitted light
- prevent loss of the excitation light
- focus emitted and excitation light upon the detector
- prevent excitation light from reaching the detector
4 - prevent excitation light from reaching the detector
Which of the following represents a primary advantage of performing fluorometric over absorption spectroscopic methods of analysis:
- increased specificity and increased sensitivity
- increased specificity and decreased sensitivity
- purity of reagents used not as critical
- ease of performing assays
1 - increased specificity and increased sensitivity
Which of the following may be associated with bioluminescence:
- light emission produced due to enzymatic oxydation of a substrate
- less sensitive than direct fluorescent assays
- electron excitation caused by radiant energy
- employs a radioactve label
1 - light emission produced due to enzymatic oxydation of a substrate
Nephelometry is based on the measurement of light that is:
- absorbed by particles in suspension
- scattered by particles in suspension
- produced by fluorescence
- produced by excitation of ground-state atoms
2 - scattered by particles in suspension
Which of the following instruments is used in the clinical laboratory or in reference laboratories to detect beta and gamma emissions:
- fluorometer
- nephelometer
- scintillation counter
- spectrophotometer
3 - scintillation counter
Which of the following best describes chemiluminescence:
- electron excitation caused by radiant energy
- enzymatic oxidation of a substrate produces light emission
- chemical energy excites electrons that emit light upon return to ground state
- employs a fluorescent label that produces light
3 - chemical energy excites electrons that emit light upon return to ground state
In assaying an analyte with a single-beam atomic absorption spectrophotometer, what is the instrument actually measuring:
- intensity of light emitted by teh analyte on its return to the ground state
- intesnity of light that the analyte absorbs from the hollow-cathode lamp
- intensity of light that the analyte absorbs from the flame
- intensity of the beam from the hollow-cathode lamp after it has passed through the analyte-containing flame
4 - intensity of the beam from the hollow-cathode lamp after it has passed through the analyte-containing flame
What is the function of the flame in atomic absorption spectroscopy:
- absorb the energy emitted from the metal analyte in returning to ground state
- supply the thermal energy needed to excite the metal analyte
- bring the metal analyte to its ground state
- supply the light that is absorbed by the metal analyte
3 - bring the metal analyte to its ground state
Most atomic absorption spectrophotometers incorporate a beam chopper and a tuned amplifier. The purpose of these components is to avoid errors that would be caused by:
- variations in flame temperature
- deterioration of the hollow-cathode lamp
- stray light from the hollow-cathode lamp
- measurement of light emitted by the analyte
4 - measurement of light emitted by the analyte
In potentiometry, which of the following is considered the standard electrode:
- hydrogen electrode
- calcium electrode
- potassium electrode
- copper electrode
1 - hydrogen electrode
In an electrolytic cell, which of the following is the half-cell where reduction takes place:
- anode
- cathode
- combination electrode
- electrode response
2 - cathode
Mercury covered by a layer of mercurous chloride in contact with saturated potassium chloride solution is a description of which of the following types of electrodes:
- sodium
- calomel
- calcium
- silver/silver chloride
2 - calomel