HEMATOLOGY INSTRUMENTATION Flashcards
Beckman Coulter Unicel DxH 800 WBC count method
Impedance volume and conductivity and five-angle light scatter measurement
Sysmex XN Series WBC count method
Fluorescent staining; forward light scatter and side fluorescent light detection
Abbott CELL-DYN Sapphire WBC count method
Light scatter (primary count)
impedance (secondary count)
Siemens ADVIA 2120 WBC count method
Light scatter and absorption
Beckman Coulter Unicel DxH 800 RBC count method
Impedance
Sysmex XN Series RBC count method
Impedance
Abbott CELL-DYN Sapphire RBC count method
Light scatter (primary count), impedance (secondary count)
Siemens ADVIA 2120 RBC count method
Low-angle and high-angle laser light scatter
Beckman Coulter Unicel DxH 800 HGB method
Modified cyanmethemoglobin (525 nm)
Sysmex XN Series HGB method
Sodium lauryl sulfate-HGB (555 nm)
Abbott CELL-DYN Sapphire HGB method
Modified cyanmethemoglobin (540 nm)
Siemens ADVIA 2120 HGB method
Modified cyanmethemoglobin (546 nm)
Sysmex XN Series HCT method
Cumulative RBC pulse height detection
Beckman Coulter Unicel DxH 800 HCT method
(RBC x MCV) / 10
Abbott CELL-DYN Sapphire HCT method
(RBC x MCV) / 10
Siemens ADVIA 2120 HCT method
(RBC x MCV) / 10
Beckman Coulter Unicel DxH 800 MCV method
Mean of RBC volume distribution histogram
Sysmex XN Series MCV method
(HCT/RBC) x 10
Abbott CELL-DYN Sapphire MCV method
Mean of RBC volume distribution histogram
Siemens ADVIA 2120i MCV method
Mean of RBC volume distribution histogram
Beckman Coulter Unicel DxH 800 MCH method
(HGB/RBC) x 10
Sysmex XN Series MCH method
(HGB/RBC) x 10
Abbott CELL-DYN Sapphire MCH method
(HGB/RBC) x 10
Siemens ADVIA 2120i MCH method
(HGB/RBC) x 10
Beckman Coulter Unicel DxH 800 MCHC method
(HGB/HCT) x 100
Sysmex XN Series MCHC method
(HGB/HCT) x 100
Abbott CELL-DYN Sapphire MCHC method
(HGB/HCT) x 100
Siemens ADVIA 2120i MCHC method
(HGB/HCT) x 100
Beckman Coulter Unicel DxH 800 RDW method
RDW as CV (%) of RBC histogram or RDW-SD (fL)
Sysmex XN Series RDW method
RDW-SD (fL) or RDW-CV (%)
Abbott CELL-DYN Sapphire RDW method
Relative value, equivalent to CV
Siemens ADVIA 2120i RDW method
CV (%) of RBC histogram
Beckman Coulter Unicel DxH 800 Reticulocyte count method
Supravital staining; impedance volume and conductivity and light scatter measurement
Sysmex XN Series Reticulocyte count method
Fluorescent staining; forward light scatter and side fluorescent light detection
Abbott CELL-DYN Sapphire Reticulocyte count method
Fluorescent staining; low-angle scatter, and fluorescent light detection
Siemens ADVIA 2120i Reticulocyte count method
Supravital staining (oxazine 750); low-angle and high-angle light scatter and absorbance
Beckman Coulter Unicel DxH 800 NRBC method
Impedance volume and conductivity and five-angle light scatter measurement
Sysmex XN Series NRBC method
Fluorescent staining; forward light scatter and side fluorescent light detection
Abbott CELL-DYN Sapphire NRBC method
Red fluorescent dye staining; forward light scatter and fluorescent light detection
Siemens ADVIA 2120i NRBC method
Multi-angle light scatter measurements in the two WBC differential channels
Beckman Coulter Unicel DxH 800 WBC Differential method
Impedance volume and conductivity and five-angle light scatter measurement (‘VCS’ – volume, conductivity, scatter)
Sysmex XN Series WBC Differential method
Fluorescent staining; forward and side light scatter, and side fluorescent light detection
Abbott CELL-DYN Sapphire WBC Differential method
Multi-angle polarized scatter separation (MAPSS) and three-color fluorescence detection
Siemens ADVIA 2120i WBC Differential method
Peroxidase staining, light scatter and absorption; for basos, differential lysis, low-angle and high-angle laser light scatter