chapter 6: qc and instrumentation artifacts Flashcards
pet qc procedure in which a source of ge68 as an external or internal rod source is placed into the bore of the gantry. data are acquired and processed to adjust the detector module gain settings
normalization
coincidence timing
energy window calibration (singles calibration)
absolute activity
energy window calibration (singles calibration)
what kind of source is used for coincidence timing?
ge68
this qc procedure evaluates the uniformity of each acquired sinogram and looks for modules that are not performing properly. it can use either ge68 rod sources or a phantom, and is also called the “air scan”.
normalization
blank scan
energy window calibration (singles calibration)
absolute activity
blank scan
this qc procedure often takes many hours to acquire a large number of events from each crystal, then every crystal is compared with the average performance of all detectors, and a calibration sensitivity factor is set for every detector
normalization
blank scan
energy window calibration (singles calibration)
absolute activity
normalization
this qc procedure sets uniform sensitivity of the detector modules and is similar to uniformity calibration on a gamma camera
normalization
blank scan
energy window calibration (singles calibration)
absolute activity
normalization
normalization is performed
daily
weekly
annually
after preventative maintenance and after major service
after preventative maintenance and after major service
this qc procedure matches the activity measured in the dose calibrator to the activity measured in the pet scanner
normalization
blank scan
energy window calibration (singles calibration)
absolute activity
absolute activity
which qc procedure is the last of all the calibrations to be performed?
normalization
blank scan
energy window calibration (singles calibration)
absolute activity
absolute activity
which qc procedure is most important for the calculation of SUVs?
normalization
blank scan
energy window calibration (singles calibration)
absolute activity
absolute activity
what type of pet scanner failure can cause changes in PMT power and drift in gains, energy window settings, and coincidence timing?
daily qc failure
energy window calibration failure
high voltage drift
failed system reboot
high voltage drift
beam hardening, due to large body habitus and low kev/ma settings, on a ct scan appears as
rings
streaks
photopenic area
shading
streaks
select all that apply: a pattern of concentric rings on a ct scan can be caused by
metallic implants
camera operating at lowest kev/ma setting
patient motion
detector out of adjustment
recon artifacts
camera operating at lowest kev/ma setting, detector out of adjustment
patient motion causes what kind of artifact on a ct scan?
ring
streak
shading
photopenic area
streak
after a patient receives a vaccine, fdg uptake will appear in the
contralateral lymph nodes near the site
ipsilateral lymph nodes near the site
focal uptake at the injection site
no fdg uptake is associated with vaccines
ipsilateral lymph nodes near the site
due to the partial volume affect, lesions less than…mm in size will be falsely reduced
10 mm
20 mm
30 mm
40 mm
50 mm
30 mm