Anesthesia Machines Flashcards
Who Oversees standards of anesthesia machine?
American Society for Testing and Materials
If the master switch is on….. (2)
Pneumatic and electrical functions
Activates alarms and safety features
If the master switch is off the only things that work are….. (3)
Battery charger
Electrical outlets
The oxygen flush valve is active
Power Failure Indicator
Newer machines have 1 backup source for all components (battery)
Keep machine plugged in- to keep batter charging.
If having to use the backup battery what drains the battery the most?
Ventilator -> hand bag to conserve battery
What is the hospital backup
red outlets
Intended to power ANESTHESIA monitor
Electrical outlets (on machine)
Other appliances should be plugged into……
hospital main
If requirements exceed outlet; circuit breaker will activate
3 parts of pneumatic system
High pressure
Intermediate pressure
Low pressure
Receives gases from cylinders at high variable pressure
High pressure system;
cylinders
pressure release valve
check valves
Purpose of the high pressure system
reduces to lower, more constant pressure suitable for use in machine
Hanger yokes ensure______
Ensures a unidirectional flow
Required to have a least 1 yoke for oxygen and 1 for nitrous
Hanger yoke parts
Body- principal framework
Retaining screw- tightens cylinder (clamp)
Nipple- thru which gas enters machine
Index pins- prevents attaching an incorrect cylinder (PISS)
Washer- forms seal between cylinder and yoke
Check valve assembly- ensures unidirectional flow
Prevents gas from exiting machine when there is no cylinder in yoke
Check valve assembly
check valves assesmbly Prevents gas from being transferred from a cylinder with higher pressure to one with lower pressure if…..
if both are in a yoke and ON
Also called reducing valves
Pressure Regulators
Purpose of the pressure regulators (2)
Reduces high, variable pressure in cylinders to lower constant pressure for machine
Regulator required for each gas supplied by cylinder
Without reducing valves……
provider would have to constantly adjust flowmeter to provide constant flow
If master switch is OFF, intermediate pressure system pressure……..
is 0
Receives gases from the pressure regulator or pipeline
Intermediate pressure system
regulator and pipeline pressure
(45 psi cylinder regulator) (50-55 psi pipeline (wall))—–more than that is too high to deliver to patients. stay below these numbers.
Intermediate pressure system includes…..(6)
Pneumatic part of master switch
Pipeline inlet connections and indicators
Piping
Oxygen pressure failure devices
Oxygen flush valve
Flow control valves