Test 1: 8 anethesia machine Flashcards
most anesthetic machines will contain:
— + —= delivery to patient
carrier gas source (O2)
inhalant volatization method
compressed oxygen can be stored as
gas cylinders
liquid oxygen
oxygen concentrator
safty connection of E cylinders
PISS (pin index safety system): prevents connecting the wrong type of gas to the machine
Fusible plug- prevents over pressurization due to increased temp
a full E tank has
660 L= 2200 psi
a full H tank has
6900 L= 2200 psi
Nitrous oxide tank will read — psi
745
will remain pressurized within the tank until the entire liquid phase is gone
measure how much is left by weighing the tank
NO is stored as liquid and gas (95% liquid when full)
liquid oxygen is stored at — °C
-160 (is a liquid at this temp)
is then pumped through hospital in pipes at 58 psi
oxygen concentrators work by
concentrating O2 in room air into 90-95% O2 at 25 psi
1-4 L/min: not strong enough to ventilate off of
pressure regulators will drop psi from — to —
2200 psi (from O2 tank) (high pressure)
45-50 psi (guarantees constant pressure in the flowmeter) (intermediate pressure)
what kind of pipelines can connect to O2 lines
DISS (diameter index safety system)
Manufacturer-specific connectors
prevents attaching to wrong gas
O2 flush valve
connects the 50 psi pipeline coming from regulator or pipeline inlet to the fresh gas outlet, bypassing the flowmeter
O2 coming out has 50 psi, 35-75 L/min → risk of barotrauma
O2 flush valves bypass — and produce O2 at — psi
flowmeter
50 psi, 35-75 L/min
go from regulator (midpressure) to fresh gas outlet
risk of barotrauma
flowmeter allows precise delivery of gas to — and —
vaporizer and common gas outlet
flow meter drops pressure to —
low (14.7 psi)
goes from tank(2200-high)
regulator (50- mid)
flowmeter (14.7-low)