Spring semester Test 5 Ch 14-22 Flashcards

1
Q

Exposure artifacts are

A

Improper pt preparation
Double exposure
Blur
Grid cutoff

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

In handling artifacts; light fog is caused by

A

The temp or humidity is too high
The safelight is too bright
Film has been left in X-ray room

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Acceptance testing
Routine performance monitoring
Maintenance

A

An acceptable QC program consists of these 3 steps

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

The Ten Step monitoring system and eval process is promoted by the

A

The joint commission JTC

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

The most important part of patient protection is ?

What is the minimum total required?

A

Filtration -2.5 mm aluminum

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

To measure proper filtration, what is used and how frequently should this be done ?

A

Half value layer

Annually or when there has been a change to tube housing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Misalignment of collimator should not exceed _____

A

2% of the SID

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Most systems use _____ collimators which can not be larger than the receptor size

A

PBL

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Spatial resolution is principally determined by

A

Focal spot size

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

3 tools used for measurement of focal spot size

A

The pinhole camera
The star pattern
Th slit camera

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

How long can a Cr sit before it needs to be erased?

A

24hrs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

The process of assigning a value to a dead pixel based on the recorded values of adjacent pixels

A

Interpolation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

A software correction that is performed to equalize the response of each pixel to a inform X-ray beam

A

Flat-shielding

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

A graph of frequency of occurrence versus digital value intervals
Has discrete plots

A

Histogram

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

IR artifacts for digital

A

Dust, dirt, scratches
Pixel malfunction
Ghost images

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Software artifacts for digital

A

Histograms
Range/scaling
Image compression

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Object artifacts in digital

A

Pt positioning
Collimation/partitioning
Backscatter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

2 artifacts in pre processing are

A

Interpolation

Flatshielding

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Artifacts that occur bcuz of dead pixels or dead rows/columns

A

Pre -processing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

Things that associate with image compression

A

Lossless compression

CAD systems require uncompressed images for processing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Compressed images may cause ____ whole using CAD systems

A

Lesions or lack of data

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Multiple images on a single IP results in a proper separation and collimation of each image

A

Partitioning

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Dust on the CR optical path (mirrors and lenses) can’t be cleaned by _______

A

Rad techs, they must have professional service

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

The backup timer terminates after how long ??

A

6sec or 600ma

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
In flouro, Entrance skin exposure (ESE) should not exceed _____
10R/min
26
Any combo of mA and S should produce same mAs value 10% Make 3 exposures the same
Exposure linearity Exposure reproducibility
27
Kvp calibration uses a _____
Ion chamber and is +-10%
28
A line pair test is used to measure
Resolution
29
Background light is also called
Ambient light
30
ROI
Region of interst -bring sharpness of interest
31
LCD CRT
Flat, active matrix Curved, light emitting, scanned electron
32
Image inversion Window and level
Black to white, white to black Makes all shades of grey visible
33
Materials that emit light Light during stimulation Glows after stimulation
Luminescence Fluorescence Phosphorescence
34
The property of the source of light Basic Photometric unit
Candela aka nit Lumen
35
LED
Light emitting diode
36
Photostimuable phosphor is used in: | TFT and CCD is used in
CR | DR
37
A property of luminescence transmitted by fiber optic bundles
A scintillator
38
Spatial resolution is the ability to
See small objects Soft tissue = low spatial frequency Bone,breast=high spatial frequenc
39
Binary system uses
2 digits (0,1)
40
Bit | Byte
(0,1) | Group of 8 bits
41
CPU | Motherboard
Control unit | Ram, Rom
42
PACS | DICOM
Picture archiving and communications system (Radiologist can access anywhere) Digital imaging and communications in medicine (image storage)
43
RIS | Earliest calculating tools
Radiology information system Abacus
44
1st general purpose computer 1st operational computer
ENIAC Colossus
45
Parallax effect = Image characteristics
Penumbra Spatial res, contrast res, noise, artifacts
46
Test device for Filtration Test device for Collimation
HVL-2.5 mm Aluminum PBL-spatial resolution [not exceed 2%]
47
How to test protective apparel?
Radiograph to check for cracks or tears
48
2 type of luminance meters
Near range Telescopic (0.05 to 1000 cd/m)
49
Test device for Kvp calliboration and criteria needed?
Ion chambers | W/in 10% of selected kvp !!
50
Th frequency for all testing is Annually!!!!! But which is done semi annually??
Collimation
51
Test device used for geometric, reflection, resolution, display noise???
TG-18-
52
Luminance response should be less than 30% _____
Nonuniformity
53
Protective apparel lead aprons for Diagnostic imaging should at least be
.25mm lead | And .50 for Flouro!!!
54
For best digital viewing, we should view __________ | Photometry is the ________
Straight on!! | Science of measuring light
55
4 components of any medical image
Spatial resolution Contrast resolution Artifacts Noise
55
To have quantum mottle is to have Contrast resolution is limited by
Not enough mAs Noise or SNR (signal to noise ratio)
55
The principle source of noise is DR is a ____bit range
Scatter radiation 14Bit
55
Direct DR uses which capture element ?? | Indirect Dr uses with capturing elements??
A-se Amorphous selenium CSI or a-Si Amorphous silicon
55
Spatial resolution is limited by
Pixel size !! And MTF!!(modulated transfer system) | Contrast resolution is more important
55
A measure of X-ray absorption efficiency
DQE | Detective quantum efficiency
55
Photopic Scotopic
Bright light with cones of eye Dim light with rods of eye
56
QC deals with what QA deals with
Instrument and equipment People
57
Exposure timer and exposure reproductiblity is within
5%
58
The SMPTE
The society motion picture and television engineers | Described format for contrast and display resolutions
59
DICOM presented a document to allow medical images to be transferred referred to as
GSDF | Gray scale display function
60
Measurements and observations should be instituted for all display devices
AAPM TG 18
61
Best digital viewing is done
Straight on!
62
The TG 18-QC | Should be viewed
Regularly | Done by technologists!!!