Rad Eye Flashcards
Radeye g measures
Gamma
Radeye g tech
Geiger mueller tube
Utilizes ____ to avoid false positives
NBR
After background established
Press ____ button ____ to enter ____ mode
Black
2x
Finder
_____ indicates functioning meter
Heartbeat icon
2 alarms
___ for low and ___ for high
1 arrow
2 arrow
Radeye prd measures
Gamma and x ray
Radeye prd measures up to
25mR/hr
Radeye prd tech
Scintillating tube
RIID EYE GIN+
Tech
Scintillating tube with QCC tech
QCC tech
Quadratic conpression conversion
Can RIIDEYE identity isotopes
Yes
Riideye identifies what types
Gamma and neutron
What is the detector type of the RIIDEye GiN+?
2” x 2” NaI(Tl) scintillator with optional neutron detection channel.
What is the energy range of the RIIDEye GiN+?
20 keV to 3 MeV.
Does the RIIDEye GiN+ have neutron detection capabilities?
Yes, with optional neutron detection channel.
What type of detector does the RadEye G use?
Energy-compensated Geiger-Müller (GM) tube.
What is the measurement range of the RadEye G?
50 µR/h to 10 R/h.
Does the RadEye G have neutron detection capabilities?
No.
What type of detector does the RadEye PRD use?
NaI(Tl) scintillation detector with a miniature photomultiplier tube.
What is the measurement range of the RadEye PRD?
1 µR/h to 25 mR/h.
Does the RadEye PRD have neutron detection capabilities?
No.
Is there natural background rejection (NBR) for the RadEye PRD?
Yes.