Mx908 Flashcards
Mx908 uses ______ type of detection
High pressure mass spectrometry
HPMS
How does the MX908 introduce a sample for analysis?
The sample is drawn in through an inlet, either as a vapor or from a swab.
What happens to the sample inside the MX908?
The sample is ionized using a chemical ionization process to create charged molecules.
How does the MX908 separate different ions?
It uses high-pressure mass spectrometry to filter ions based on their mass-to-charge ratio.
How does the MX908 identify substances?
It compares ion signatures to a built-in chemical database.
What categories does the 908 detect?
Explosives
Cwa
Narcotics
Tics
States of matter and modes
Gas/vapor mode
Trace/aerosols mode
Orientation for trace mode?
Flat
Can vapor and aerosols mode be swapped while on?
No
What mode should be run first when the TRACE MODULE is installed?
Clean Mode
What should be replaced if necessary during system setup?
Sampling Module
What check must be performed before selecting a mission?
System Check
Recommended mode for missions?
Hunter Mode(s)
What error occurs if TRACE MODE Clean is unsuccessful after 3 attempts?
TRACE MODE CLEANING ERROR
How many times can CLEAN MODE be re-run for TRACE MODE cleaning?
Up to 3 times
What should be used to clean the TRACE MODULE?
Alcohol wipes
If a LOW PRESSURE ERROR occurs, what component is likely clogged?
MS Core
What should be done after replacing or cleaning the MS Core?
Re-calibrate MS Core
What module and swabs are required for TRACE Sampling?
TRACE MODULE & TRACE SAMPLING SWABS
What technique should be used for TRACE Sampling?
Swab-the-Swab
What modules are required for VAPOR or AEROSOL Sampling?
VAPOR or AERO MODULE
What function eliminates persistent vapor background?
ZERO function
What tool helps with challenging vapor conditions?
Captive Sampler