AGAPPE Flashcards

1
Q

This reagent is intended for in vitro quantitative determination of glucose in serum, plasma and CSF

A

Intended use: Glucose

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

GOD-PAP abbrev

A

Glucose oxidase- Phenol amino phenazone

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

GOD-PAP methology kay linear siya up to _________

A

600 mg/dl

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

_______ is a major carbohydrates present in blood and serves as primary source of energy

A

Glucose

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

ingested from starch and sugar

A

Glucose

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Quinonimine produce color ________

A

Red

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

MSDS will be provided on request

A

Risk and Safety

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Glucose reagent and standard are ready to use

A

reagent preparation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

The sealed reagents _________ are stable up to the expiry date on the label, when stored at ______ from the light

A

Reagent storage and stability - 2-8 degree celcius

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Once opened, the reagent is stable up to _______ at __________, if contamination is avoided.

A

Open Vial Stability - 2-8 degrees celcius

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

On-board calibration stability is _______

A

20 days

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

turbidity or precipitation in any kit component indicates detoriarition and the component must be discarded

A

reagent detoriation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

to avoid contamination use clean laboratory wares.

Avoid direct exposure if reagent to light. do not blow into the reagetnt

A

Precaution

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Reagents must be disposed off in accordance with local regulations

A

waste managements

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Serum/plasma free of hemolysis/CSF

A

Sample

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

mode of reaction

A

End point

17
Q

Slope of reaction

A

increasing

18
Q

wavelength

A

505 (440-550) nm

19
Q

Temperature

A

37 degrees celcius

20
Q

Standard concentration

A

100 mg/dl

21
Q

Linearity

A

600 mg/dl

22
Q

incubation time

A

10 minutes

23
Q

blank

A

reagent

24
Q

sample volume

A

10 uL

25
Q

Reagent volume

A

1000 uL

26
Q

Cuvette

A

1 cm light path

27
Q

It is recommended that each laboratory establish its own reference values

A

reference range

28
Q

It is recommended to use aggape qualicheck norm and path

A

quality control

29
Q

this reagent is linear upto ________

A

linearity