Transcutaneous Oximetry (TcPO2) Flashcards
trancutaneous oximetry helps determine ___
wound healing and amputation level
transcutaneous oximetry reflects ___
tissue partial pressure of oxygen (PO2)
transcutaneous oximetry electrode houses a heating element that heats the skin to ___ to increase blood flow
45°C
transcutaneous oximetry
a self-adhesive molded plastic fixation ring is applied to the intact skin and a few drops of ___ are put inside the plastic ring
electrolyte solution
PO2 readings are noted within ___
15 to 20 minutes
where is the transcutaneous oximetry reference site?
upper left chest
transcutaneous oximetry
the electrode is applied (2)
near a wound
at the anticipated level of amputation
if the PO2 levels is less than normal or poor, TcPO2 with ___ can be performed
oxygen challenge
TcPO2 interpretation
normal
poor
critical ischemia
normal ( > 50mmHg )
poor ( < 40mmHg )
critical ischemia ( < 30mmHg )
factors that affect TcPO2 values (4)
arterial PO2
skin blood flow
skin composition
capillary temperature under the sensor