NBRC SAE - Notable Questions Flashcards
- After several plethysmographic measurements of VTG, a pulmonary function technologist is unable to obtain consistent values. The technologist should instruct the patient to
A. pant deeper and faster.
B. pant gently when the shutter is closed.
C. breathe slowly and deeply.
D. exhale completely before the shutter is closed.
B.√ pant gently when the shutter is closed.
(u) A. Panting deeper may shift lung volumes and generate pressures outside the calibrated range.
(c) B. Panting gently at a rate of 30-60/min when the shutter is closed is the correct maneuver for determining VTG.
(u) C. A panting frequency of approximately 1 per second is adequate to overcome small leaks in the box, such as those used for temperature compensations. See explanation B.
(u) D. Completely exhaling before shutter closure would provide for a measurement of RV.
- An 8-year-old patient with sickle cell anemia is referred for a treadmill exercise (stress) test with pulse oximetry. The following measurements are observed during the procedure:
_________Rest____4 min_____6 min_____8 min_____10 min
BP______98/74____130/70____150/78____152/78____154/78
SpO2_____98______97_______92_______94_______97
HR (min)___88______157______180_______170______190
RR (min)___20______35_______40________45______56
Which of the following should a pulmonary function technologist conclude?
A. The red blood cells are sickling and preventing the uptake of oxygen.
B. Bronchospasm is occurring with exercise, causing desaturation.
C. Motion artifact is preventing an accurate reading of the pulse rate.
D. Increased vasodilatation is caused by increased activity.
C.√ Motion artifact is preventing an accurate reading of the pulse rate.
(u) A. There is no evidence of a change in oxygenation during the test.
(u) B. Bronchospasm would not cause oxygen saturation to drop and then reverse.
(c) C. Motion artifact is causing inaccurate readings in the heart rate and SpO2.
(u) D. There is no indication of vasodilatation.
- A patient’s V̇O2max was determined during a progressive exercise (stress) test on a bicycle ergometer. The test was repeated the next day on a treadmill and the V̇O2 max was 20% less. Which of the following should a pulmonary function technologist conclude?
A. This finding is consistent with published differences between V̇O2max as determined by the ergometer and treadmill.
B. The treadmill test should be repeated and the patient instructed to refrain from supporting his weight on the side rails.
C. The patient pedaled at 30 rpm and needs further instruction.
D. The bicycle ergometer is out of calibration and needs to be returned to the factory.
B.√ The treadmill test should be repeated and the patient instructed to refrain from supporting his weight on the side rails.
(u) A. Published differences show that V̇O2max values on the treadmill are higher than those on a bicycle ergometer.
(c) B. The treadmill test must be incorrect. Treadmill V̇O2max values are usually greater than V̇O2max values achieved on a bicycle ergometer. The observed discrepancy occurs when the patient supports part of his/her weight on the treadmill side rails.
(u) C. A pedal speed of 30 rpm may be too slow on a mechanically braked ergometer. However, this pedal speed may be compensated by an electronically braked ergometer. The observed discrepancy would not be expected on either type of ergometer.
(u) D. V̇O2max is defined as a point where work continues to increase, but V̇O2 does not. Ergometer calibration is not relevant to this measurement.
- A 24-year-old trauma victim received mechanical ventilation for 6 weeks before extubation and early rehabilitation. The following flow-volume loop is obtained and is repeatable:
See “NBRC SAE Exam - Question 11” Image
Which of the following is most likely associated with the shape of the loop?
A. asthma
B. aspiration pneumonitis
C. pulmonary fibrosis
D. upper airway obstruction
D. Upper airway obstruction, as a side effect of long-term endotracheal tube use, would limit inspiratory airflow more severely than expiratory flow because it is outside the chest cavity (extrathoracic).
(u) A. Asthma would affect expiratory flows more than inspiratory flows. See explanation D.
(u) B. Aspiration pneumonitis might cause a decrease in vital capacity, but not inspiratory flow limitation.
(u) C. Pulmonary fibrosis might cause a decrease in lung volume, but no significant inspiratory flow limitation.
(c) D. Upper airway obstruction, as a side effect of long-term endotracheal tube use, would limit inspiratory airflow more severely than expiratory flow because it is outside the chest cavity (extrathoracic).
Note: the image shows a variable extrathoracic obstruction based on the reduced inspiratory flows. An upper airway obstruction is the same as an extrathoracic obstruction.
- An observed PETCO2 in a healthy adult at rest has what relationship with the measured PaCO2 value?
A. The PETCO2 is lower than the PaCO2.
B. The PETCO2 is higher than the PaCO2.
C. The PETCO2 approximates the PaCO2.
D. The PETCO2 is double the PaCO2.
C.√ The PETCO2 approximates the PaCO2.
The PETCO2 generally approximates the PaCO2; therefore, it is unlikely to be significantly higher or lower than, or double, the PaCO2 in a healthy adult at rest.
- A 30-year-old patient is being tested on a computerized cycle ergometer. The computer display indicates 100 watts, but the patient’s VO2 is at 500 mL/min. What is the most likely cause?
A. There is excessive resistance in the cycle ergometer.
B. There is loss of communication with the cycle ergometer.
C. The patient is pedaling at 30 revolutions per minute.
D. The weighted strap has become disconnected.
B.√ There is loss of communication with the cycle ergometer.
(u) A. A higher than normal resistance would be accompanied by the patient’s VO2 being higher than 500 mL/min.
(c) B. The patient’s VO2 is not consistent with 100 watts and is more consistent with unloaded pedaling. This suggests loss of communication between the cycle ergometer and the computer.
(u) C. Pedaling frequency would not explain the discrepancy between the VO2 and the displayed workload of 100 watts on a computerized cycle ergometer.
(u) D. The computerized cycle ergometer does not use a weighted strap for resistance.
- The following pulmonary function data are obtained from a 60-year-old male who is an ex-smoker and is undergoing a disability evaluation:
FVC 2.2 L FEV1 1.9 L FEF25-75% 4.1 L/sec MVV 75 L/min TLC 2.7 L
Which of the following conclusions should be made concerning the reliability of these data?
A. The FVC is inconsistent with TLC.
B. The FEF25-75% was calculated incorrectly.
C. The patient is malingering.
D. These data can be reported.
D. These data can be reported.
(u) A. The FVC of 2.2 L is consistent with a restrictive disorder, which is also indicated by the TLC and FEV1.
(u) B. The FEF25-75% may be within the normal range in patients with restrictive disorders.
(u) C. The FEV1 and MVV are consistent because one can approximate the MVV by multiplying FEV1 by 35 (= 66.5 L/min). Therefore, malingering is unlikely.
(c) D. The FVC, FEV1, and TLC are consistent with a restrictive process and should be reported.
- Which of the following methods should be used to establish predicted values for a new laboratory?
A. Choose equations from a group similar to the testing population.
B. Use the default settings established by the manufacturer.
C. Average the predicted values from several studies.
D. Select 10 healthy subjects to confirm predicted values.
A. Choose equations from a group similar to the testing population.
(c) A. ATS/ERS recommends selecting predictive equations derived from subjects similar to the population tested.
(u) B. This is commonly done, but it may not be appropriate. See explanation A.
(u) C. Predicted values from several studies should not be averaged under normal circumstances.
(u) D. Global Lungs Initiative recommends 150-300 subjects in each grouping to confirm appropriateness of equations.
- An infrared CO analyzer produces inconsistent readings in response to a calibration gas. Which of the following should a pulmonary function technologist do?
A. Replace the ionization chamber.
B. Refresh the Wheatstone bridge.
C. Adjust the sampling flow.
D. Replace the Nafion (Perma-Pure®) tubing.
D.√ Replace the Nafion (Perma-Pure®) tubing.
(u) A. There is no ionization chamber in this instrument.
(u) B. There is no Wheatstone bridge in this instrument.
(u) C. Adjusting the flow will not cause inconsistent readings since the introduced sample will have a constant concentration.
(c) D. Condensation in the sample chamber will gradually change as the humidity of the introduced sample changes. Nafion (Perma-Pure®) tubing stabilizes the water vapor content of the sample gas to prevent this. This could interfere with the infrared beam and cause the reading to be variable.
- A patient had the following pulmonary function test results at baseline and 6 months after starting bleomycin for lymphoma:
Parameter_______ Baseline_______6-Month Follow-up
FEV1 (L)___________3.90_____________3.80
FVC (L)___________4.59_____________4.50
FEV1/FVC (%)_______85_______________84
TLC (L)____________6.10_____________6.02
DLCO____________30.30____________20.20
V̇ A (L/min)________5.90______________5.81
Which of the following is most consistent with these results?
A. The quality of the test results is questionable.
B. The patient has developed obstruction.
C. The patient has developed medication toxicity.
D. There is no significant change in lung function.
C.√ The patient has developed medication toxicity.
(u) A. The results of each test are internally valid.
(u) B. There is no evidence of obstruction.
(c) C. A decrease in DLCO is a known and common adverse reaction to bleomycin. It is the reason why serial pulmonary function tests are performed on these patients.
(u) D. This is untrue, as there is a significant decrease in the DLCO.
- A pulmonary function technologist is developing an asthma management plan for patient education for use in a pediatric population. Which of the following should be considered in the plan?
- population diversity
- literacy level
- socioeconomic status
- zone system settings
A. 1, 2, and 3 only
B. 1, 2, and 4 only
C. 1, 3, and 4 only
D. 2, 3, and 4 only
B.√ 1, 2, and 4 only
- True. Population diversity will influence the language for delivery of the education and asthma management plan.
- True. Literacy level impacts the selection of appropriate educational materials.
- False. Socioeconomic status may influence access to healthcare, but does not impact the development of the educational materials or asthma management plan.
- True. The zone system related to peak flow or symptoms is an essential component of asthma management plans.
(u) A. incorrect and incomplete
(c) B. correct
(u) C. incorrect and incomplete
(u) D. incorrect and incomplete
- An arterial blood gas sample is collected into a plastic syringe. To ensure the analysis of the sample is accurate, it is critical that the sample
- contain no bubbles.
- not be placed in ice slush.
- be analyzed within 5 minutes.
- be mixed before analysis.
A. 1, 2, and 3 only
B. 1, 2, and 4 only
C. 1, 3, and 4 only
D. 2, 3, and 4 only
B.√ 1, 2, and 4 only
- True. The sample must not contain bubbles.
- True. The sample should not be placed in a container with ice slush because the pO2 will be affected.
- False. The sample does not need to be analyzed within 5 minutes, but should be analyzed within 30 minutes.
- True. The sample must be mixed before analysis.
(u) A. incorrect and incomplete
(c) B. correct
(u) C. incorrect and incomplete
(u) D. incorrect and incomplete
- A 70-year-old patient’s VC is 1.4 L and FEV1 is 1.0 L. A pulmonary stress test indicates that maximum exercise V̇ F is 40 L/min. Which of the following should a pulmonary function technologist conclude?
A. The data are consistent with severe obstructive disease.
B. The results should be reported.
C. The VC is inconsistent with stress test results.
D. The FEV1 is inconsistent with stress test results.
B.√ The results should be reported.
(u) A. The FEV1/FVC = 0.71, which is NOT consistent with severe obstructive disease in a 70-year-old patient.
(c) B. Based on an FEV1 of 1.0 L, the expected V̇ Fmax will be approximately 35 to 40 L/min. The measured exercise ventilation is consistent with the FEV1 and should be reported as ventilatory limitation.
(u) C. The VC is not related to the expected V̇ Fmax.
(u) D. See explanation B
- A DLCO system that uses a gas chromatograph shows inadequate separation between the Ne and CO signals. Which of the following will correct this problem?
A. Replace the column.
B. Increase the carrier gas flow.
C. Replace the DLCO gas tank.
D. Decrease the breath-hold time.
A.√ Replace the column.
(c) A. Inadequate separation of the gas signals occurs when the gas chromatograph column is exhausted, and reducing the carrier gas does not solve the problem.
(u) B. Decreasing, not increasing, carrier gas flow may help to extend the useful life of the column.
(u) C. The DLCO gas does not affect the analysis of the exhaled gas.
(u) D. Decreasing breath-hold time will not change the analysis problem described.
- An ECG simulator is used to assess the accuracy of a 12-lead system used for exercise (stress) testing. The following ECG is recorded, including a 1-millivolt calibration signal:
See “NBRC SAE Exam - Question 64” Image
Which of the following correctly describes this simulated tracing?
A. The voltage is set to the standard ECG sensitivity.
B. The voltage is set to half-standard ECG sensitivity.
C. The simulated voltage in leads aVL and V3 is incorrect.
D. The simulated voltage in leads II and aVR is incorrect.
A.√ The voltage is set to the standard ECG sensitivity.
(c) A. The standard ECG sensitivity setting for a 1-millivolt calibration signal is 10 mm and is correctly displayed.
(u) B. See explanation A.
(u) C. The leads are correct as displayed. See explanation A.
(u) D. See explanation C.
Look at the calibration box on the left of the strip. It is one large box wide and 2 boxes high. This is 10mm of deflection, and the correct calibration for a 1 millivolt calibration at 25mm/sec speed.