Kettering Practice Exam - Notable Questions Flashcards
Equipment Operation:
Pneumotachometers can provide an accurate means for documenting flowrate if which of the following conditions are met?
1. The same pressure and gas composition used for calibration of the device are used for measurements. 2. The device is calibrated according to the manufacturer’s specifications. 3. The device is properly cleaned and maintained.
A.1 and 3 only
B.2 and 3 only
C.2 only
D.1, 2, and 3
D.1, 2, and 3
Equipment Operation:
The primary operational difference between a flow-type and pressure-type plethysmograph is measurement of
A.box pressure.
B.mouth pressure.
C.box volume.
D.Raw and Gaw.
A. box pressure
Lung Testing:
Pulmonary function studies on a 29-year-old male patient with no history of lung disease reveal the following results:
ACTUAL PREDICTED FVC (L) 4.90 5.0 FEV1(L/sec) 3.40 4.0 FEF25-75% (L/sec) 4.05 4.50 FRC (L) 2.64 2.20 TLC (L) 6.67 5.80 DLCO (mL CO/min/mm Hg) 25 25
The pulmonary function technologist should interpret these results as
A.normal pulmonary function tests.
B.the increased FRC indicates that the patient has significant air-trapping.
C.the increased FRC indicates that “switch-in” occurred above end-tidal expiration.
D.the FVC is invalid and is not consistent with the FRC values.
C.the increased FRC indicates that “switch-in” occurred above end-tidal expiration.
Equipment Operation:
Which of the following types of breathing valve should be used when performing a single-breath CO diffusion study?
A.One-way directional
B.Two-way nonrebreathing
C.Electronic solenoids
D.Free-breathing
C.Electronic solenoids
Exercise Testing:
The following pattern is observed on the ECG monitor while a patient performs an incremental stress test:
See “Practice Exam - Question 21” image
The pulmonary function technologist should
A.terminate the stress test.
B.administer a beta blocker.
C.initiate oxygen therapy via nasal cannula.
D.continue the test and monitor the patient closely.
D.continue the test and monitor the patient closely.
Lung Testing:
The following input/output signal response curve was recorded with a Fleisch pneumotach prior to measurement of Raw in a body plethysmograph:
See “Practice Exam - Question 25” image
This curve is best described as an example of
A.excessive effort.
B.underdamping.
C.overdamping.
D.alinearity.
A. excessive effort.
Exercise Testing:
The following ECG pattern is observed during a cardiopulmonary exercise stress test using a cycle ergometer:
See: “Practice Exam - Question 28” image
The technologist should interpret this rhythm as
A.myocardial infarction.
B.premature ventricular contraction.
C.ischemia/injury.
D.normal sinus.
B. premature ventricular contraction
Exercise Testing:
While performing a multi-stage exercise stress test, the patient’s heart rate does not increase normally and the test is stopped after 3 minutes because of marked dyspnea and fatigue. This response would be associated with all of the following conditions EXCEPT
A.sympatholytic drug use.
B.complete heart block.
C.myocardial ischemia.
D.poor conditioning.
D. poor conditioning
Spirometry:
While performing an MVV maneuver, the patient begins to complain of light-headedness and tingling in her fingers. The pulmonary function technologist should associate these findings with
A.bronchodilator toxicity.
B.myocardial infarction.
C.respiratory alkalosis.
D.impedance of venous return.
C. respiratory alkalosis
Equipment Operation:
Which of the following will result if a mylar membrane is used on a Clark electrode?
A.Proper diffusion of carbon dioxide
B.Insufficient diffusion of carbon dioxide
C.Insufficient pH change of electrolyte
D.Insufficient current flow
D. Insufficient current flow
Lung Testing:
If a patient’s airways resistance is reported to be 0.8 cm H20/L/sec, what is the patient’s conductance?
A.0.4 L/sec/cm H20
B.0.8 L/sec/cm H20
C.1.25 L/sec/cm H20
D.2.50 L/sec/cm H20
C.1.25 L/sec/cm H20
Equipment Operation:
Which of the following analyzers requires a carrier gas of helium or neon?
A.Gas chromatograph
B.Mass spectrometer
C.Infrared absorption
D.Thermal conductivity
A. Gas chromatograph
Equipment Operation:
The partial pressure of oxygen in a gas mixture is measured by all of the following EXCEPT
A.gas chromatograph.
B.clark electrode.
C.polarographic.
D.fuel cell.
A. gas chromatograph
Lung Testing:
The normal value for the DLCO is
A. 25 mL CO/min/mm Hg BTPS
B. 25 mL CO/min/cm H2O BTPS
C. 25 mL CO/min/cm H2O STPD
D. 25 mL CO/min/mm Hg STPD
D. 25 mL CO/min/mm Hg STPD
25 “mm Hg”
DLCO is always STPD
Equipment Operation
The best method for disinfection of pulmonary function breathing hoses and mouth pieces is
A. ethylene oxide.
B. steam autoclave.
C. glutaraldehyde.
D. ethyl alcohol.
C. gluteraldehyde