NBRC Practice Exam - Notable Questions Flashcards
- Instrumentation/Equipment
Set Up, Maintain, Calibrate
An elevation in which of the following may distort COHb and SaO2 (meas) values obtained from a hemoximeter (CO-oximeter)?
A. fetal hemoglobin
B. methemoglobin
C. deoxyhemoglobin
D. total hemoglobin
A. fetal hemoglobin
- Instrumentation/Equipment
Set Up, Maintain, Calibrate
An autocalibration system indicates 8 cm H2O while an external manometer indicates 10 cm H2O during a manual calibration of a mouth pressure transducer in a body plethysmograph. Which of the following should a pulmonary function technologist do to correct the problem?
A. check for a leak in the u-tube manometer.
B. add water to the u-tube manometer.
C. service the autocalibration device
D. check for a leak in the mouth shutter.
C. service the autocalibration device.
2 devices are measuring different values upon calibration. One device is erroneous.
- Instrumentation/Equipment
Set Up, Maintain, Calibrate
Which of the following is required to verify speed accuracy of a new treadmill?
A. stop watch
B. rotameter
C. odometer
D. biologic control
A. stop watch
You need belt lenth x rpm.
to find rpm (revolutions per minute) you need a way to keep time.
- Instrumentation/Equipment
Set Up, Maintain, Calibrate
According to the ATS/ERS 2005 recommendations on diffusing capacity, a carbon monoxide analyzer should be within what percentage of linearity?
A. 0.5
B. 1.0
C. 5.0
D. 10.0
A. 0.5
Carbon monoxide linearity should be within 0.5%
- Instrumentation/Equipment
Troubleshooting
An automated blood gas analyzer indicates that it is unable to calibrate the pH electrode. Which of the following would cause this problem?
A. the wrong calibration gas is installed.
B. the internal barometer is malfunctioning
C. the system is unable to maintain electrode temperature.
D. the hemolyzer failed.
C. the system is unable to maintain electrode temperature
- Instrumentation/Equipment
Troubleshooting
A consistently slow response by a blood gas O2 electrode is most likely an indication of
A. moisture contamination
B. electrode not at thermal equilibration
C. membrane contamination with protein.
D. air bubbles in the electrolyte solution.
C. membrane contamination with protein
- Instrumentation/Equipment
Troubleshooting
A hemoximeter (CO-oximeter has produced the following results on a series of patient samples during a 1-hour period
MetHb: COHb:
- 9_______-0.10
- 2_______0.01
- 8_______0.02
- 4_______-0.04
Which of the following corrective actions should a pulmonary function technologist recommend?
A. Clean the cuvette
B. Change the flush solutions
C. Run a hemoglobin calibration
D. Change the calibration gases
A. Clean the cuvette
- Instrumentation/Equipment
Troubleshooting
The measured FVC of a biologic control has decreased by 0.50L for the past 2 days. Which of the following should a pulmonary function technologist do FIRST?
A. replace the flow sensor and recalibrate
B. replace the batteries in the portable spirometer.
C. place a bacterial filter inline and recheck the biologic quality control.
D. check the FVC of the biologic quality control on a different spirometer.
A. replace the flow sensor and recalibrate
FVC decreased by half a liter for 2 days in a row (on a known biologic control.) First thing to do is replace the sensor and recalibrate.
- Procedures
Select Test Protocols and Equipment
Which of the following tests should be used to evaluate a 5 year old child with asthma like symptoms?
A. maximal voluntary ventilation
B. methacholine challenge
C. mannitol challenge
D. lung volumes by gas dilution
B. methacholine challenge
- Procedures
Select Test Protocols and Equipment
Pulse oximetry is requested during a maximum exercise stress test performed on a patient with severe COPD who smokes. Which of the following should a pulmonary function technologist recommend?
A. measure exhaled CO2
B. obtain an arterial blood gas sample.
C. perform baseline spirometry
D. proceed with the study as requested
B. obtain an arterial blood gas sample
- Procedures
Select Test Protocols and Equipment
A high school athlete is referred for an exercise induced bronchospasm evaluation. During a history evaluation, she states it is hard to breathe when running. A pulmonary function technologist should
A. initiate 25% supplemental O2 for the exercise test.
B. modify the workload for the exercise test.
C. obtain tidal breathing loops during the exercise test.
D. administer 2 puffs of albuterol via MDI/spacer prior to testing.
C. obtain tidal breathing loops during the exercise test.
“hard to breathe when running.”
Flow volume loops should be obtained during exercise.
We are looking for an obstruction (fixed or variable) that would cause difficulty during exercise. We will need flow volume loops to assess the problem.
- Procedures
Perform the Procedure
According to ATS/ERS recommendations for children less than 10 years of age, what is the minimal exhalation time (sec) for acceptable spirometry?
A. 3
B. 4
C. 5
D. 6
A. 3 seconds
Minimum exhalation time for kids under 10 years of age is 3 seconds.
- Procedures
Perform the Procedure
When measuring the FRC by N2 washout, which of the following decreases washout time most effectively?
A. initiating the test at FRC level
B. performing an IC maneuver periodically
C. initiating a 10 second breath hold every 30 seconds
D. performing an FVC maneuver at the end of the test
B. performing an IC maneuver periodically
- Procedures
Perform the Procedure
A pulmonary function technologist notices that FRC is 200% of predicted at the conclusion of a helium dilution test. Which of the following should the technologist do next?
A. report the results
B. check the supplemental O2 supply
C. check for a leak in the system
D. reduce the BTPS correction factor.
C. check for a leak in the system
Too much helium has been diluted, which has caused the FRC to appear higher than it is.
If too much helium has been diluted, it will result in an unusually high FRC. The helium was diluted by leaking out of the system.
- Procedures
Evaluate Validity of Result
A patient without lung disease has an elevated RV with a normal FRC measured by N2 washout. Which of the following should a pulmonary function technologist do?
A. repeat the FRC measurement
B. repeat the vital capacity maneuver
C. check for a leak in the system
D. perform an inspiratory capacity maneuver
B. repeat the vital capacity maneuver.
There is no leak (FRC is normal).
No need for an IC.
No indication that the N2 system is malfunctioning, so no need for an additional FRC.