The technologist is performing calibration of a blood gas analyzer that has the following calibration gases:
CO2 5%
O2 12%
N2 83%
Which of the following would be the correct value for the Severinghaus electrode with a PB of 760 mm Hg?
A. 50.6 mm Hg
B. 45.6 mm Hg
C. 35.6 mm Hg
D. 25.6 mm Hg
C. 35.6 mm Hg
When preparing for a routine exercise stress test, the technologist should set the paper speed on the ECG machine to record at
A. 5mm/sec
B. 10mm/sec
C .15mm/sec
D. 25mm/sec
D. 25mm/sec
When checking the operation of a mass spectrometer to analyze gas concentration during pulmonary function testing, the technologist observes that the total sum of all the gases analyzed is less than 100%. The technologist should
A. increase the concentration of each gas until it reaches 100%
B. increase the overall gain until the sum reaches 100%
C. obtain another analyzer
D. decrease the sample line.
B. increase the overall gain until the sum reaches 100%.
Which of the following is/are used to properly calibrate a body plethysmograph?
- rotameter
- barometer
- dosimeter
- sine-wave rotary pump
A. 1 and 2 only
B. 1, 2, and 4 only
C. 2 and 4 only
D. 1 and 3 only
B. 1, 2, and 4 only
A technologist is peforming DLCO testing. Which of the following analyzers will provide the quickest response time for this diagnostic test?
A. infrared absorption
B. gas chromatograph
C. geissler tube ionizer
D. wheatstone bridge
A. infrared absorption
The most appropriate application of inductive plethysmography is while performing
A. exercise stress testing
B. spirometry
C. bronchial challenge testing
D. sleep studies
D. sleep studies
inductive = “motion”
inductive pleythysmography is used in sleep studies to monitor chest movement and tidal volume. The sensors are wrapped around the chest and abdomen.
All of the following are components of the Sanz electrode EXCEPT
A. KCl buffer solution
B. mylar membrane
C. ceramic plug
D. silver-silver chloride wire.
B. mylar membrane
Mylar membranes are associated with the clark electrode.
While performing routine spirometry testing using a uni-directional valve, the patient informs the technologist that he can inhale, but it is difficult to exhale. In this case, the valve
A. is malfunctioning
B. has very high resistance
C. placement is reversed
D. is stuck in the closed position
C. placement is reversed
While performing a helium dilution procedure for FRC measurement, the technologist notices that the patient is tachypneic two minutes into the testing period. The most common cause is
A. the scrubber is exhausted
B. the patient has had an allergic reaction to the helium
C. the desiccant is missing
D. this is normal response to testing
A. the scrubber is exhausted
The two most common indications of CO2 rebreathing are tachypnea and feeling hot. If the scrubber is exhausted, the patient will rebreathe CO2.
The technologist is asked to measure PEF at the patient’s bedside. The technologist has the patient perform the maneuver a total of six times. The last four readings were exceptionally low. Which of the following is the most likely cause for the difference in the readings?
A. patient fatigue
B. use of a low range peak flow meter
C. heating element is not working properly
D. bent vane
A. patient fatigue
A technologist is working at a small rural hospital which has a fully functioning water-seal spirometer. While performing calibration, the technologist notices that the volume calibration produces a value of 3.2L consistently. All of the following could be the cause of the readin EXCEPT
A. a bent bell.
B. low water level
C. air is leaking into the system
D. the potentiometer is malfunctioning.
B. low water level
Which of the following items of equipment can produce a consistent flowrate to properly calibrate a spirometer for flow?
A. dry rolling seal
B. wedge bellows
C. rotameter
D. potentiometer
C. rotameter
The pulmonologist has issued an order for pre & post bronchodilator spirometry. The technologist has confirmed the order and noted in the chart that the patient is unable to correctly use an MDI. Which of the following devices should the technologist recommend to administer the bronchodilator?
A. dry powder inhaler
B. small volume nebulizer
C. dosimeter
D. breath-actuated nebulizer
B. small volume nebulizer
The main purpose of the vent in the body plethysmograph is to
A. allow the patient to pant easier.
B. provide accurate measurement of FRC
C. maintain the temperature
D. account for a small leak
C. maintain the temperature
All of the following are advantages of using a leg ergometer for exercise stress testing EXCEPT
A. cycling is a natural motion for most patients
B. less danger of patient falls or injury
C. results are independent of body weight
D. patient can stop the exercise at any point during the test
A. cycling is a natural motion for most patients