M&M NonCardiac Monitoring Flashcards
Mandatory monitor for any anesthetic, including moderate sedation, no contraindications
pulse oximetry
Where is the pulse ox sensor applied?
A sensor with a light source and detector is placed across the tissue, that can be transilluminated.
Oxygenated hgb absorbs more
Infrared light (940nm)
Deoxygenated hgb absorbs more
RED light (660 nm)
What is the basis of oximetric determinations?
Change in light absorption
How does the pulse oximetry?
Microprocessor analyzes the ration of red to infrared absorptions. to provide O2 saturation.
Pulse ox , arterial pulsations, identified by
Plethysmography, allows for correction for absorption by nonpulsative venous blood and tissue.
Clinically detectable cyanosis requires
5 g of desaturated hgb and corresponds to an SPO2 of less than 80%
Usually goes undetected by pulse ox in the absence of lung disease or low FiO2
Bronchial intubation
Methemoglobin has the
Same absorption coefficient at both red and infrared wavelengths.
Red and infrared with methemoglobin absorps in what ratio?
1:1
1:1 absorption ration seen with methemoglobin corresponds to a saturation reading of
85%
Things causes pulse ox artifact.
Methylene blue dye
low perfusion
malpositioned sensor
leakage of light from the light emiting diode to the sensor.
What is required to obtain mixed venous blood oxygen saturation
Pulmonary artery catheter with fiberoptic sensor to determine SvO2
Noninvasive brain oximetry
monitors regional O2 saturation of Hg in the brain.
In contrast to pulse ox, brain oximetry measures
Venous
Capillary
Arterial saturation
thereby providing oxygen saturation of all regional hemoglboin
What can dramatic decrease regional oxygen saturation (rSO2)
Cardiac arrest
Cerebral embolization
Deep hyothermia
Severe hypoxia
Capnographs rely on the absorpition of
Infrared light by CO2
Capnography: Nondiverting (mainstream)
Measure Co2 passing through an adpator placed in the breathing circuit.
Capnography: Diverting (aspiration)
Continously suction gas from the circuit into a sample cell in the monitor.
Capnography: Diverting (aspiration): HIGH ASPIRATION rates
defined as up to 250ml/min and low dead-space sampling tubing usualy increase SENSITIVITY and decrease lag time.
Capnography: Diverting (aspiration): Low ASPIRATION rates
50ml/min can underestimated ETCO2 during rapid ventilation.
Capnography : diverting (sidestream) units are prone to
Water precipitation that can cause obstruction of the sampling line and erroneous readings.
How is expiratory valve malfunction determined
Detected by the presence of CO2 in inspired gas.
Capnographs reliably indicate ______Intubation but do not reliably detect ______intubation
Esophageal intubation ; bronchial intubation
May indicated a circuit disconnection is
Sudden cessation of CO2 during expiratory phase of the capnograph
A marked rise in ETCo2 may be caused by
the increased metabolic rate associated with MH
The PaCo2-ETCO2 gradient is usually
2-5 mmHg reflects alveolar dead space (alveoli that ventilated but not perfused)
What are the 3 phases of capnography and what do they mean?
Phase I : Dead space
Phase II: Dead space and alveolar gas
Phase III: Alveolar gas plateau
Characteristic of capnograph with no plateau before inspiration
COPD , shark fin
COPD on the ETCO2 and arterial CO2 gradient
Increases the gradient
Any depression during phase III (plateau) indicates
SPONTANEOUS respiratory effort.
Failure of the inspired CO2 to return to zero may indicate?
Incompetent expiratory valve or exhausted Co2 absorbent
Persistence of exhaled gas during the inspiratory phase signals the presence of
an incompetent inspiratory valve
2 ways anesthetic gas are analyzed
Infrared absorption
Piezoelectric Analysis
Infrared absorption is based on this law
Beer-Lambert law
Do not absorbed infrared light must be measure by other means
N2 and O2
What does the BEER-LAMBERT law states:
The absorption of infrared light passing through a solvent (inspired or expired gas) is proportional to the amount of unknown gas.
Piezoelectric analysisi uses
oscillating quartz crystals one of which is covered in lipid.
In piezoelectric analysis, Volatile anesthetic dissolve in the
Lipid layer, and their concentration is determined by the change in oscillation frequency. This method cannot distinguish different anesthetic agents.
3 ways OXYGEN is analyzed?
Galvanic cell
PARAMAGNETIC ANALYSIS
Polarographic electrode
Galvanic cell for the analysis of oxygen explain?
Galvanic cell hydroxyl ions are formed at the gold cathode and react with lead anode.
Galvanic cell for the analysis of oxygen An
electrical current is produced that is proportional to the amount of oxygen being measured
Oxygen is a nonpolar gas that is paramagnetic and expands when placed in a magnetic field. By switching the field on and off, the volume change can be used to measure O2 content. which concept of O2 analysis?
PARAMAGNETIC ANALYSIS concept
A semipermeable membrane separates gold cathode and silver anode . A voltage is applied and OH- are formed from O2; resulting current is proportional to the amount of O2. which concept of O2 analysis?
Polarographic electrode concept
SPIROMETRY: Low and HIGH peak inspiratory pressure:
Indicates circuit disconnect or airway obstruction, respectively
SPIROMETRY: Minute ventilation
Obtained by measuring Vt and breathing frequency.