Why is QC of digital display devices important?
Malfunctioning displays can degrade images, simulating or obscuring disease.
Who should supervise a comprehensive QC program for digital radiography?
qualified medical physicist
What described the format, dimensions, and contrast characteristics of a pattern used to make measurements of the resolution of display systems?
SMPTE
SMPTE
Social of Motion Picture and Television Engineers
NEMA
National Electrical Manufacturers Association (NEMA)
The intent of this standard was to allow medical images to be transferred according to the DICOM standard to be displayed on any DICOM-compatible display device with a consistent gray scale appearance
NEMA-DICOM
DIN
Deutsches Institut für Normung
equal changes in digital values associated with an image translate into equal changes in perceived brightness at the display
principle of perceptual linearization
DIN published a document in 2001 called?
“Image Quality Assurance in X-ray Diagnostics; Acceptance Testing for Image Display Devices”
This was developed as an acceptance testing standard to address the requirements for digital display devices
DIN 2001
In 1998, the Flat Panel Display Measurement (FPDM) standard, version 1.0, was released by???
VESA
VESA
Video Electronics Standard Association
This standard provides a set of instructions that can be used to help in the evaluation of system performance according to a compliance standard
VESA
This was developed to ensure acceptable clinical performance and to developed set of test patterns.
AAPM TG 18
AAPM
American Association of Physicists in Medicine
TG 18
Task Group Report 18
What is used to measure luminance response and uniformity of monitors?
luminance meter
2 types of luminance meter
*near-range photometer
*telescopic photometer
This is used in close proximity to monitors
near-range photometer
This is used to test from a distance of 1m.
telescopic photometer
The luminance meter should use a calibration method that is traceable to the National Institute of Standards and Technology (NIST) and should be able to measure luminance in the range of 0.05 to 1000 cd/m2 with better than 5% accuracy and a precision of at least 0.01.
basahin mo lang d ko alam how iquestion to
What device used to measure CIE-specified color coordinates of a digital display device?
colorimeter
The photometer also should comply with the Commission Internationale de l’Éclairage (CIE) standard photopic spectral response within a range of 3%.
isa pa to
What device is used to evaluate monitor display reflection and to assess ambient light conditions?
illuminance meter