CER Chapter 9 Review Quiz Flashcards
Disinfection Processes for Flexible Endoscope
All endoscopes should be high-level disinfected or sterilized between each patient: True or False
True
Endoscopes are considered _____ devices.
a. critical
b. semi-critical
c. non-critical
d. Class V
Semi-critical
The minimum effective concentration of high-level disinfectants/liquid chemical sterilants should be tested at least daily: true or false
False
Minimum effective concentration and minimum recommended concentration mean the same thing: True or False
True
When manually rinsing an endoscope after high-level disinfection, the endoscope should be rinsed:
a. at least three times using fresh water for each rinse
b. once with sterile water
c. with a diluted disinfectant
d. per the manufacturers’ IFU
IFU
After high-level disinfection, the endoscope should be assembled before storage: True or False
False
High-level disinfection solutions should be discarded after each use. True or False
False
Adding more high-level disinfectant to soak basin or automated equipment reprocessor to bring the solution up to the desired level also extends the length of the time the solution can be used:
True or False
False
Automated endoscope reprocessor device connectors are designed for use with any flexible endoscope. True or False
False
When using an automated equipment reprocessor with an automated cleaning cycle, it is still necessary to thoroughly manually clean the endoscope before using the automated cycle.
True or False
True
Automated equipment reprocessors should be quality tested:
a. at least monthly
b. daily
c. with each use
d. at least weekly
At least weekly
Water bottles can be processed in any automated equipment reprocessor. True or False
False
The following should be documented each time an item is high-level disinfected:
a. name or initials of he person performing the disinfection
b. minimum effective concentration
c. date and time the high level disinfection process took place
d. all the above
All the above
The high level disinfectant solution in an automated equipment reprocessor should be tested for the minimum effective concentration:
a. daily
b. each shift
c. before each use
d. all the above
Before each use