CER Chapter 15 Review Quiz Flashcards
Ultrasonic Probes
TEE probes are used for:
a. assessing the heart’s function
b. the insertion of heart catheters
c. visualizing the pulmonary structures
d. all the above
Assessing the heart’s function
Soiled probes should be transported in:
a. closed carts
b. impermeable bags
c. basins with lids
d. all the above
all the above
Small holes in probes:
a. are ok if found on the cord section
b. pose no risk to patients
c. can allow fluid invasion
d. means the probe cannot be cleaned
can allow fluid invasion
Endocavity probes should be HLD:
a. immersing the handle and head
b. no HLD is needed if a sheath was used during the procedure
c. immersed in an AER
d. following manufacturers IFU
Following mfg IFU
All AER’s can be validated to disinfect ultrasound probes.
True or False
False
Endocavity probes can be sored on the console as long as:
a. they do not touch other parts of the console
b. the console is kept in a secured area
c. the probe is covered with a clean storage cover
d. all the above
The probe is covered with a clean storage cover
When manually disinfecting TEE probes:
a. the probe should be completely immersed without coiling the probe in a closed container
b. only the cord and control handle should be disinfected
c. only the transducer should be disinfected
d. immerse the transducer and the handle in the disinfectant
Only the transducer should be disinfected
TEE probes should be:
a. inspected daily
b. fully submerged in the appropriate HLD
c. inspected with a lighted magnifying device
d. all the above
Inspected with a lighted magnifying device
Probes should be stored:
a. in a horizontal position
b. in a closed container
c. with a sterile protective sheath in a vertical position
d. in a vented cabinet
In a vented cabinet
Before cleaning Endocavity probes:
a. check the connector for damage
b. inspect the probe for visible damage
c. check the lens for damage
d. all the above
All the above