Endoscope Inspection and Preparation
Using poor technique or skipping leak testing can cause harm to the patient? True or False
True
Damage to endoscope lumens increases the risk that debris may be left inside the endoscope after cleaning. True or False
True
The use of a ________during the endoscope inspection process allows the technician to inspect the inner channels of the endoscope.
a. magnifying glass
b. borescope
c. microscope
d. all the above
borescope
Using a venting cap during sterilization will cause the endoscope sheath to rupture. True or False
False
When inspecting an endoscope light cord, broken light fibers can be detected by:
a. using a borescope
b. small nicks in the sheath
c. black spots on the end of the cord
d. scratches in the lumens
Black spots on the end (tip) of the cord
Visual inspection of flexible endoscopes should occur:
a. before the procedure
b. during the procedure
c. after cleaning
d. all the above
All the above
When preparing an endoscope for high-level disinfection it is important to put the valve cover back on the endoscope to ensure the disinfectant can get inside the lumens:
True or false
False
Ethylene oxide sterilization of a flexible endoscope may result in a blown endoscope if:
a. the endoscope is not vented properly
b. the endoscope’s internal channels are moist when sterilized
c. the endoscope has a cracked lens
d. the distal tip is not protected
Not vented properly
Early detection of endoscope damage can protect the patient and:
a. decrease the processing time
b. is required by the FDA
c. Reduce repair costs
d. all the above
Reduce repair costs