Most multi-part flexible endoscopes are their accessories must be disassembled for sterilization. True or False
True
When the color of chemical indicator tape used to seal flat wrap changes, it indicates that the item is sterile. True or False
False
Gas plasma sterilizers must be in an area physically separated from all other work areas. True or False
False
The OSHA admin requires a method to alert employees to emergency situations for this type of low-temperature sterilizer:
a. ethylene oxide
b. hydrogen peroxide gas plasma
c. vaporized hydrogen peroxide
d. all the above
Ethylene oxide
Items being prepared for ethylene oxide sterilization should be:
a. clean
b. dry
c. compatible with ethylene oxide
d. all the above
All the above
All flexible endoscopes should have a ventilation adaptor or water resistant cap in place when undergoing ethylene oxide sterilization; True or False
False
An effective way to dry endoscopes before ethylene oxide sterilization is the use heated forced air. True or False
False
Peel pouches are an acceptable packaging material for ethylene oxide sterilization. True or False
True
When running an ethylene oxide sterilizer, the load should be full - unless there is a medical necessity to run a partially filled load. True or False
True
A biological indicator should be run in each ethylene oxide cycle. True or False
True