Work Area Design
Effective work area design should incorporate:
a. an irrigating ultrasonic cleaner
b. multiple automated equipment reprocessors
c. storage for common endoscope replacement parts
d. one-way flow
One way flow
Inadequate workspace in the endoscope work area can be a cause of:
a. poor workflow
b. aerosols entering the clean work area
c. endoscope damage
d. all the above
endoscope damage
Endoscopes that have been high-level disinfected should be stored on open wire shelving: True or False
False
The recommended temperature for the endoscope clean processing area should be:
a. 75 or less
b. 60 to 65
c. at least 75
d. can be found in ASHRAE/ASHE 170
Can be found in ASHRAE/ASHE 170
When airflow is negative it means air flows into that area.
True or Falst
True
Doors and pass-through windows between the decontamination and disinfection area should remain closed; however, they may remain open between the disinfection and sterilization area.
True or False
False
The relative humidity in all endoscope work areas should not exceed…..
a. 20%
b. 30%
c. 50%
d. 60%
60%
The automated endoscope reprocessor should be placed in the decontamination area to ensure the endoscopes are clean before sending them to the preparation area. True or False
False
Physical barriers between soiled and clean areas can prevent aerosols from entering the clean area. True or False
True
When sending endoscopes from the user area to the decontamination area, endoscopes and their accessories, such as forceps and wires, should be transported together in the same closed container. True or False
False