Unit 6 - Hemodynamic Monitors & Equipment Flashcards
what do Korotkoff sounds represent
turbulent flow in an artery that was previously occluded by the BP cuff
BP auscultation
- relies on korotkoff sounds
- SBP is measured at the first sound
- DBP is measured when the last sound disappears
what is cuff pressure when korotkoff sounds are produced
between SBP and DBP
how do NIBP machines measure BP
Oscillatory Method
Inflatable cuff occludes arterial blood flow, and as the cuff pressure is released, the monitor measures the pressure fluctuations that occur in response to arterial pulsations
when is SBP measured with oscilattory method
when oscillations 1st appear (the reappearance of flow after cuff occlusion)
when is DBP measured with oscillatory method
measured at the minimum pressure where oscillations can still be generated
BP measurement reading that’s the most susceptible to error with oscillatory method
DBP
why won’t NIBP work in a pt on CPD or with LVAD
requires pulsatile flow
ideal bladder length for BP cuff
80% of extremity circumference
ideal bladder width of NIBP cuff
40% of extremity circumference
NIBP reading with a cuff that’s too small
overestimates SBP
cuff pressure required to occlude artery is higher
NIBP reading when cuff is too large
underestimates SBP
cuff pressure required to occlude artery is lower
what happens to SBP, DBP, and pulse pressure measurements as pulse moves from aortic root toward periphery
SBP increase
DBP decrease
PP widens
SBP, DBP, and PP at aortic root
SBP lowest, DBP highest, PP narrowest
SBP, DBP, and pulse pressure at radial artery compared to aortic root
SBP higher
DBP lower
PP wider
BP reading if BP cuff above heart
BP reading falsely decreased (less hydrostatic pressure)
BP reading if cuff below heart
BP reading falsely increased (more hydrostatic pressure)
every 10 cm change in BP cuff above/below heart = BP changes by _____
7.4 mmHg
Complications of NIBP Measurement
- pain
- neuropathy (radial, ulnar, median)
- measurement errors
- limb ischemia
- compartment syndrome
- bruising
- petechiae
- interference with IV medications
what is measured at the peak of art line waveform
SBP
what is measured at the trough of art line waveform
DBP
what does the upstroke of art line waveform represent
contractility
what part of arterial line waveform represents stroke volume
area under curve
where does art line monitor BP
at level of transducer (not at site of catheter insertion)
art line transducer location that won’t be affected by changes in body or extremity position
level of RA
causes of falsely increased NIBP
- BP cuff too small
- BP cuff too loose
- Bp measured on extremity below level of heart
causes of falsely decreased NIBP
- BP cuff too large
- cuff deflated too rapidly
- measured on extremity above level of heart
measurement that remains constant throughout arterial tree
MAP
where does art line pressure have the greatest pulse pressure
dorsalis pedis
SBP increases along arterial tree as a function of pressure waves reflec
where is arterial DBP measurement lowest
dorsalis pedis
what does optimal waveform morphology balance
amount of damping with amount of distortion from transducer system
High-pressure flush test (square test):
shows how fast the system vibrates in response to a pressure signal
what informs about damping characteristics in art line system
number of oscillations after flush test
when is an art line considered optimally damped
baseline is re-established after 1 oscillation
when is art line waveform considered under damped
baseline re-established after several oscillations
BP measurements with under damped art line
SBP overestimated, DBP underestimated, MAP accurate
causes of underdamped art line
- stiff (non-compliant) tubing
- catheter whip (artifact)
when is an art line system considered over damped
baseline re-established with no oscillations
BP measurements in overdamped art line system
SBP underestimated, DBP overestimated, MAP accurate
causes of overdamped art line system
- air bubble in pressure tubing
- clot in catheter
- low flush bag pressure
- kinks
- loose connection
how does dicrotic notch change with art line monitoring location
moves further away from systolic peak the further the monitoring site is from the heart
where should CVL tip rest
just above the junction of vena cava & right atrium
risks assoc. with CVL catheter in heart chambers
↑ risk dysrhythmias, thrombus formation, cardiac perforation
where should pulmonary artery catheter tip be
- in the pulmonary artery, distal to pulmonic valve
- 25-35 cm from VC junction
3 steps to calculate distance of CVL insertion
- Know the distance from site of entry to VC junction
- Know distance from VC junction to where catheter tip should be (only applies if placing PAC)
- Add these 2 numbers together to determine distance from site of insertion to catheter tip
distance from subclavian vein to junction of vena cava and RA
10 cm
distance from right IJ to junction of vena cava and RA
15 cm
distance from left IJ to junction of vena cava and RA
20 cm
distance from femoral vein to junction of vena cava and RA
40 cm (either side)
distance from median basilic vein to junction of vena cava and RA
right = 40 cm
left = 50 cm
distance from vana cava/RA junction to RA
0-10 cm
distance from vena cava/RA junction to RV
10-15 cm
distance from vena cava/RA junction to PA
15-30 cm
distance from vana cava/RA junction to where PAOP is measured
25-35 cm