Ultrasound Flashcards
How is a sound wave different to an EM wave?
Unlike electromagnetic waves, sound waves cannot be transmitted through a vacuum; they require a medium in the form of a gas (such as air), a liquid (such as water or blood) or a solid (such as soft tissue, bone or metal).
An electromagnetic wave is transverse but a sound wave is longitudinal i.e. the pressure variation occurs in the direction of travel. Like all other sinusoidal waves, a sound wave is characterised by frequency, wavelength and speed (or velocity).
What is the pulse-echo principle?
The pulse-echo principle states that the distance of a reflecting object can be established by timing the go-and-return of a short pulse if the speed of the pulse is known. The returning pulse is known as the echo.
What is the rough speed of sound in tissue?
around 1540 m/s
What is ultrasound?
The answer is that it is simply sound at a frequency (pitch) which is too high for us to hear. In all other respects it is the same as normal sound.
Why are echoes smaller for deeper tissues?
some of the signal is lost (attenuated) on the way and hence there is a general tendency for echoes to get smaller as the depth increases.
What is TGC?
control which allows the operator to adjust the amount of gain (amplification) which is applied to echoes from different depths. This is called time gain compensation, and is an important control for the operator to identify and master. It normally needs to be adjusted during a scan and certainly between patients.
What is A-mode scanning?
This way of presenting information about the targets along a single direction is called the A-scan. it does not resemble a 2D cross-section image but may be regarded as a 1D image line
What is b-mode scanning?
The assembly of many A-scan lines forming a composite image in this way is called a B-scan. In a modern B-scan, the dots do not have the same brightness. Dot brightness depends on echo amplitude and is displayed on a greyscale with white representing very large amplitude echoes and black being used when no echoes are detected (Fig 2). Furthermore, each B-scan is acquired so rapidly that the image is obtained in ‘real time’.
Concerning the use of ultrasound in a medical setting:
A. Time gain compensation is used to correct for variations in the speed of sound between tissues
B. Medical ultrasound imaging typically uses frequencies in the range 3-150 kHz
C. Each single transmitted pulse may result in the creation of a large number of echoes
D. Soft tissues typically have sound speeds of the order of 1500 m/s
E. The angle at which the beam meets an organ boundary may have a large impact on the echo generated
F
F
T
T
T
Time gain compensation is not used to correct for variations in the speed of sound between tissues, but for the increase in echo attenuation with tissue depth.
Medical ultrasound imaging typically uses frequencies from around 1 MHz to around 10 MHz.
Which of the answers below is/are a correct description of TGC?
A. It is fitted to virtually all ultrasound scanners
B. It alters the way ultrasound is generated
C. It needs adjustment between patients and between views
D. It can create artefacts if incorrectly applied
E. It influences only the B-scan
Answers A, C and D are correct.
Time gain compensation does not alter the way ultrasound is generated.
Time gain compensation influences the A-scan as well as the B-scan.
Although some scanners are now fitted with automatic TGC, it is always present in some guise. Its effect is confined to processing signals from received echoes. Since the TGC is an attempt to correct for tissue attenuation, it follows that it needs to be altered between patients, but also within a scan on an individual patient when moving from one organ or section to another with different acoustic properties.
What is the piezoelectric effect?
If an electric voltage is applied to specific materials, they deform, and if a mechanical pressure is applied to them, they develop a voltage on their surface.
What materials are used for ultrasound transducers?
Some naturally-occurring materials (such as quartz crystals) have piezoelectric properties and these were originally used as ultrasound transducers. However, modern imaging transducers are made of synthetic crystalline ceramic materials such as lead zirconate titanate (PZT).
What is the curie temperature?
Immediately after manufacture, this material is not piezoelectric. However, heating to raise its temperature beyond a certain value (the Curie temperature) while applying an external voltage causes electric dipoles within the crystal to align and give the material piezoelectric properties. If the voltage is maintained as the material cools below the Curie temperature the dipole alignment and hence the piezoelectricity is preserved. The material is then cut into a suitable size and shape for use as an ultrasound transducer. Note that further heating above the Curie temperature destroys the piezoelectric properties.
What does the resonant frequency depend on?
the thickness of the transducer and so the same voltage spike will create different damped responses in different transducers.
At the resonant frequency, the wavelength of an ultrasound wave in the transducer is twice the transducer thickness. Therefore, for a given transducer material, resonant frequency and transducer thickness are inversely proportional to each other; a thick crystal gives a low resonant frequency and vice versa
What is the spatial resolution of ultrasound?
The spatial resolution of an imaging system may be thought of as the extent to which the system blurs the image of an object. Alternatively, it can be regarded as the ability of the system to distinguish two separate small objects lying close together. As the distance between the objects is reduced, there will come a point at which the system merges the two and they blur into a single image
What is axial resolution?
Axial resolution refers to a situation in which two objects lie along a single scan line i.e. along the axis of the ultrasound beam; it is sometimes called the longitudinal resolution.
when the object separation is half the length of the ultrasound pulse i.e. half the spatial pulse length the objects will not be resolvable. This distance is a measure of the axial resolution.
How does axial resolution change with frequency?
speed = frequency x wavelength; if ultrasound frequency increases, wavelength must decrease to compensate since speed is more or less constant (assumed to be 1540 m/s in soft tissue). Therefore, higher frequencies are associated with shorter wavelengths and so 2.5 cycles at a higher frequency will occupy a shorter distance (spatial pulse length).
What is lateral resolution?
resolving objects in the lateral (aka azimuthal plane)
The task of resolving these objects is a different one. Note firstly that it requires more than one beam. In fact we need a minimum of three:
The first beam must hit one object but not the other
The second beam must go through the gap between them
The third must hit the second object only
The beam width is a measure of the lateral or azimuthal resolution.
What is slice thickness resolution?
out-of-plane resolution or elevational resolution. Slice thickness is a measure of elevational resolution and it depends on the dimension of the transducer in this direction
What crtieria does the matching layer of the probe have to meet?
In front of the transducer is a matching layer. It serves as an interface between the PZT transducer element and the patient’s skin. Specifically, it reduces the effect of differences in a property called acoustic impedance, whose value is much greater for PZT than for soft tissue. For it to be effective, the matching layer has a thickness equal to a quarter of the ultrasound wavelength in the material of the layer and acoustic impedance equal to the geometric mean of the values for PZT and tissue. In the absence of a matching layer, much of the ultrasound would be reflected back towards the transducer at its interface with the skin.
What lies behind the PZT crystal in the transducer?
Behind the transducer there is a backing block and an acoustic absorber. The backing block is made of a material that is highly attenuating and has acoustic impedance similar to that of PZT, such as epoxy resin loaded with tungsten. This means that ultrasound energy emitted by the back face of the transducer is transmitted into the block and attenuated by it. This dampens the vibration of the transducer producing pulses of short spatial length and so improving axial resolution. The acoustic absorber attenuates stray ultrasound energy.
What is SPL determined by?
In imaging with pulsed ultrasound, a short spatial pulse length (SPL) improves axial resolution. The SPL is determined by the degree of transducer damping.
What is the Q factor of an ultrasound pulse?
The range of frequencies in the spectrum is known as the bandwidth of the pulse and the Q factor is defined as fo/bandwidth.
What properties does a lightly damped US beam have?
A lightly damped transducer is associated with long SPL, narrow bandwidth and high Q