TG-179 Values and Information Flashcards
1
Q
What are the recommended daily tests for CT based IGRT systems? And their respective tolerances
A
Safety (collision interlocks, warning lights)
Laser/image/treatment isocenter coincidence (+- 2mm)
Phantom localization and repositioning with couch shift (+- 2 mm)
2
Q
How often do you need to perform geometric calibration maps?
A
Monthly or upon upgrade
3
Q
How often do you need to measure couch shift accuracy?
A
Monthly
4
Q
Tolerance for kV/MV/laser alignment
A
+- 1 mm
5
Q
Tolerance for couch shifts
A
+- 1 mm
6
Q
What image quality tests are required monthly?
A
Scaling Uniformity Noise High contrast spatial resolution Low contrast detectability
- CT number accuracy (but only if IGRT is used for dose calculation)
7
Q
How often should you measure imaging dose?
A
Annually
8
Q
Typical dose for Varian OBI (CBCT)
A
0.2 - 2.0 cGy