TG-179 Values and Information Flashcards

1
Q

What are the recommended daily tests for CT based IGRT systems? And their respective tolerances

A

Safety (collision interlocks, warning lights)

Laser/image/treatment isocenter coincidence (+- 2mm)

Phantom localization and repositioning with couch shift (+- 2 mm)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

How often do you need to perform geometric calibration maps?

A

Monthly or upon upgrade

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

How often do you need to measure couch shift accuracy?

A

Monthly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Tolerance for kV/MV/laser alignment

A

+- 1 mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Tolerance for couch shifts

A

+- 1 mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What image quality tests are required monthly?

A
Scaling
Uniformity
Noise
High contrast spatial resolution
Low contrast detectability
  • CT number accuracy (but only if IGRT is used for dose calculation)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

How often should you measure imaging dose?

A

Annually

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Typical dose for Varian OBI (CBCT)

A

0.2 - 2.0 cGy

How well did you know this?
1
Not at all
2
3
4
5
Perfectly