First Year Exam: MU Calcs Flashcards
What is an “equivalent field”
Any field that has the same depth-dose characteristics as an irregular field
(it can be square or circular)
You can even do rectangular, but that in itself would further need another equivalent square field so there’s no reason to really do that…
What field size should you use for Sc for a SSD setup? What about SAD?
You always use field size at 100 cm for Sc regardless of whether you’re SSD or SAD
This should make sense, Sc is dependent only on collimator scatter. The patient setup makes no difference
What field size do you use for Sp for a SSD setup?
Field size at the surface of the phantom/patient
What field size do you use for Sp for a SAD setup?
Field size at the target
When are you allowed to use Scp instead of Sc or Sp?
When the field size you would use for Sc and Sp are both the same
This happens in a 100 SSD setup
or
If you set your target to 100 SAD
What field size do you use for PDD?
Always use field size at the surface
What field size do you use for TMR?
Always use field size at point of measurement
It’s not always 100 SAD. Sometimes they’ll ask for some point off of 100 SAD
Where does TG-71 recommend normalization point for photon outputs be set to?
10 cm
Note: most clinics, including ours, use dmax (which is also acceptable according to the TG report). But TG-71 recommends 10 cm because it’s easier and more clinically relevant
In what scenario would MU calcs still be used?
Emergency setups that don’t require simulation/would be more efficient without simulation
Why does TG-71 actually recommend using 10 cm as normalization depth? (4 reasons)
- Different machines at same energy will be matched to a more clinically relevant depth
- May decrease differences in programmed MU when moving patients from one machine to another
- Some TPS systems specifically require measured output factors at d = 10 cm, thus you’d have to take this measurement anyway. May as well kill two birds with one stone
- Some field size dosimetric quantities vary less at depths > dmax
What normalization depth does TG-71 recommend for electrons?
dmax
If your secondary MU check is failing, what are some steps you can take to explore this issue? (5 suggested steps)
- Verify a calculation error hasn’t been made by reviewing the parameters used in the MUV (this includes beam modifiers and beam parameters that are used for the TPS)
- Confirm an appropriate comparison point was used (in field and away from heterogeneity)
- Confirm geometric information between MUV and TPS are consistent (should be since the body contour and radiological densities are typically assigned in the MUV)
- Investigate if the algorithm for calcing itself has some magnitude and direction of error that is expected to be constant and predictable that wasn’t taken into account
- If you can’t figure out whats causing the discrepency, alert the Physician and see how they would like too proceed given some known uncertainty
What values are typically taken into account for photon MU calcs (7)?
- PDD or TMR
- OAR
- Collimator scatter
- Phantom scatter
- Beam modifier transmission factors
- Mayneord factor (only for PDD)
- Inverse square corrections
What values are typically taken into account for electron MU calcs (5)?
- Applicator factor
- Cutout factor
- The treat to isodose line
- Inverse square correction (SSDeff factor)
- Electron depth dose (but note, if you normalize to an isodose like, then the EDD just becomes the treat to isodose line, as is the case most of the time). So EDD will often = 0.9 or 0.8
What is the equation for the applicator factor of rectangular field (AxB)?
What is the analogous equation for the EDD?
Se(AxB) = [AFAxA*AFBxB]1/2
EDD(AxB) = [EDDAxA*EDDBxB]1/2