What is an “equivalent field”
Any field that has the same depth-dose characteristics as an irregular field
(it can be square or circular)
You can even do rectangular, but that in itself would further need another equivalent square field so there’s no reason to really do that…
What field size should you use for Sc for a SSD setup? What about SAD?
You always use field size at 100 cm for Sc regardless of whether you’re SSD or SAD
This should make sense, Sc is dependent only on collimator scatter. The patient setup makes no difference
What field size do you use for Sp for a SSD setup?
Field size at the surface of the phantom/patient
What field size do you use for Sp for a SAD setup?
Field size at the target
When are you allowed to use Scp instead of Sc or Sp?
When the field size you would use for Sc and Sp are both the same
This happens in a 100 SSD setup
or
If you set your target to 100 SAD
What field size do you use for PDD?
Always use field size at the surface
What field size do you use for TMR?
Always use field size at point of measurement
It’s not always 100 SAD. Sometimes they’ll ask for some point off of 100 SAD
Where does TG-71 recommend normalization point for photon outputs be set to?
10 cm
Note: most clinics, including ours, use dmax (which is also acceptable according to the TG report). But TG-71 recommends 10 cm because it’s easier and more clinically relevant
In what scenario would MU calcs still be used?
Emergency setups that don’t require simulation/would be more efficient without simulation
Why does TG-71 actually recommend using 10 cm as normalization depth? (4 reasons)
What normalization depth does TG-71 recommend for electrons?
dmax
If your secondary MU check is failing, what are some steps you can take to explore this issue? (5 suggested steps)
What values are typically taken into account for photon MU calcs (7)?
What values are typically taken into account for electron MU calcs (5)?
What is the equation for the applicator factor of rectangular field (AxB)?
What is the analogous equation for the EDD?
Se(AxB) = [AFAxA*AFBxB]1/2
EDD(AxB) = [EDDAxA*EDDBxB]1/2
For the square root rule for the EDD and output factor of a rectangular field/applicator, what setup table is the data references at?
If SSD = 100, 105, or 110, or any other SSD that you already have a EDD or AF table for, then use the respective SSD
If it’s for an SSD that you don’t have a table for, then assume SSD = 100, and then you need to apply an additional SSDeff factor
When do you use the SSDeff factor?
Only when you use a SSD that you don’t have a table for. Typically this is anything other than SSD = 100, 105 and 110
What is the equation for the SSDeff factor?
[(SSDvirtual + dmax) / (SSDvirtual + dmax + gap)]2
Where the ‘gap’ is the difference between the two SSD setups
So if you have SSD = 103 cm setup, and you only have a SSD = 100 cm table, then gap = 3 cm
Conceptually speaking, what is the “Virtual Source Distance” (VSD)?
It’s an arbitary point in space in which electrons follow a psuedo inverse square relationship relative to
What is an alternative to the extended SSD technique that can be used for electron MU calculations?
The Air-Gap technique
What are the rules for selecting which physical wedge factor to use in a wedge factor (100 SSD) table?
If you’re using the wedged PDD table - use field size at surface and wedge factor at dmax, regardless of depth of target. This is because the table already accounts for the WF at the target, but you need to still normalize to dmax so you have to use the WF for that instead
If you’re using open PDD table - Use field size at surface and wedge factor at depth of target
If you’re using openTMR table - use field size at target and depth if target
DO NOT USE WEDGED TMR TABLE!!!! You haven’t seen any problem solutions that used it, so don’t try it. Just use the open TMR table if need be, then follow the above rule.
What does SPD stand for?
Source to point distance
True or False
For extended SSD or extended SPD setups, you can use TMR for any of them?
True
TMR is independent of distance from source. All you would have to do is adjust for output at the point of interest (inverse square), and field size at point of interest
If you’re comfortable with using TMR instead of SSD+Mayneord factor, then it’s MUCH easier to just use TMR for any extended SSD stuff
True or False
For extended SSD problems, if your isocenter is not in the patient, you cannot use TMR table
False
Nothing is stopping you. Your calculation point does not need to be at isocenter at all. You would just need to do a inverse square correction to account for change in output at isocenter