2nd Year: SRS/SRT Flashcards
For small fields, why do you NOT measure PDD? Only TMR?
It’s very difficult to measure PDD due to chamber movement from CAX as the chamber changes depth. Your alignment is critical and has to be centered perfectly the entire range of depths
For TMR, since the chamber only stays in one spot, it only needs to be aligned once
For cone, why do we always use the same jaw setting for every cone attachment and every cone plan you do?
Changing jaw size affects output factor due to differences in scattering surfaces. You need to stay consistent since your output factors are defined purely for one jaw setting
What does isocentricity measure? What tests do we do that check it?
Size of isocenter
Spoke shot and WL both check it
What does isocenter coincidence check? What test do we do that checks it
Coincidence of radiation and mechanical isocenter
WL checks it
You can argue, if lining WL up using imaging, that imaging isocenter coincidence with the others is also checked, since you define imaging isocenter as the center pixels, so if center sphere aligns to that, then you say imaging isocenter is checked against radiation and mechanical
What test done daily for TB3 checks imaging and radiation isocentricity?
Isocal
(WL also does check mechanical and radiation isocentricity)
For WL, do we collimate with MLC or Jaws?
Always MLC
This is because…
A) We always define the last layer of field collimation using the MLCs
B) MLC positioning is more accurate than Jaw
C) You always want to test the mechanical aspect which is defining a field. For all of IMRT, that is MLCs
What is the isocenter of an image, by definition?
The literal location of the center pixel in your imaging array, rotated about all angles
That is, if something is aligned to imaging isocenter, Ex. a WL cube, then it should be at the center pixel across all projections
What is the benefit of frame delivery vs frameless? What is the drawback
Frame delivery is more accurate and does not require IGRT or any external tracking. It forces the skull to a certain orientation
However, it’s often much more uncomfortable and painful for the patient and requires more setup/time commitment to attach and to do everything same day
How are frames attached? How is patient aligned for treatment?
Screwed to the skull with local anesthesia
There’s a special device that shows torque on skull so you don’t crack it
The frame attaches to the couch top in a known and indexed manner. This makes alignment as close to exact as possible, and does not allow for wiggle
Frame has a known coordinate system for alignments
What are the benefits to using noncoplanar beams over coplanar beams?
Noncoplanar beams can achieve higher conformality and steeper dose gradients, in addition to lower surface dose
In general, what is mechanical tolerances for mecahnical devices?
< 1 mm
What are the main defining characteristics of SRS?
Steep dose gradients
High targeting accuracy (<1 mm)
High dose per fraction ( >= 5 Gy)
High conformity
1 to 5 Fx
Target diameter < 4 cm
How do cones achieve steeper dose fall-off?
Multiple noncoplanar arcs
Typically use lower energies
Smaller collimator-to-tumor distances
How does GammaKnife manage to get steeper dose gradients than Linac based SRS?
Large number of noncoplanar isocentric beams
Small collimator-to-target distances
Lower energy photons, that means lower energy and lower range secondary electrons
How often will a GammaKnife unit require a source exchange?
Every 5 to 6 years
(about 1 half-life)
What is the typical initial dose rate at isocenter of a GammaKnife unit when the sources are first received?
3 Gy/min
What is typical prescription (1 Fx) for the following metastases sizes?
<= 2 cm
2.1 - 3 cm
3.1 - 4 cm
<= 2 cm - 20-24 Gy
2.1 - 3 cm - 18 Gy
3.1 - 4 cm - 15 Gy
What is typical prescription dose for a trigeminal neuralgia?
80 - 90 Gy x 1 Fx to the 100% isodose line
What are the advantages to using GammaKnife over Linac based SRS?
Higher accuracy and pecision
Less moving parts (easier to get everything within QA tolrances)
Quicker QA
Used longer in the field (more research/experience)
Very rare downtimes
What are the disadvantages to using GammaKnife over Linac based SRS?
Can only treat brain
Usually only treats single fractions
Longer treatment times
Not many units come with IGRT capabilities
Requires source exchange after 5 - 6 years
Radiation safety precautions
When would multi-fraction SRT be preferred over single fraction SRS?
Retreatments
Tumors > 4 cm max dimension
Lesion located near critical structures
For a 1.5 T MRI scanner, what the is approximate geometric distortion due to fluctuations of gradient field near the center of the stereotactic space?
What about near the edge (near headframe base for example)
< 0.5 mm near center
about 2 mm near headframe base
How does the strength of magnetic field influece the geometric distortion thatis seen in MRI?
Distortion magnitudes are approximately proportional to field strength
Ex. For a 1.5 T field, the distortion at center is < 0.5 mm. For a 3T field it is < 1 mm
Ex. For a 1.5 T field, distortion near edge is 2 mm. For a 3T field it is 4 mm
What MRI protocol is the go-to for drawing GTV for SRS/SRT?
T1 with contrast
What is one dose concern for multi-lesion plans that is not present for single lesions? How do you avoid it?
Dose bridging
That is, a certain isodose line bridges between two nearby lesions, giving high dose to the healthy brain between lesions
To minimie, draw control regions around this dose bridging, where you can optimize down the dose
What are some special contours that we draw for SRS/SRT optimization? How are they used?
What are some special tricks we also employ?
Outer brain - 2 cm cropped from PTV. max dose 30% in optimizer
0.5 cm ring - 0.5 cm cropped from PTV. max dose 90% in optimizer
1 cm ring - 1 cm cropped from PTV. max dose 50% in optimizer
2 cm ring - 2 cm cropped back from PTV. max dose 30% in optimizer
Brain - GTV - cropped 0 cm from GTV. You want V12 Gy < 10 cc for SRS
Bridging control - locate bridging dose region after first run. Draw structure around region and minimize dose to this area
NTO = 120 (as opposed to the usual 100)
No entry through eyes
Back-to-back method - that is, give the optimizer the impossible task of having an immediate dose fall-off after giving prescription dose to target, and also conformal dose in target. This will make your coverage worse, but fall-off and conformality better. When you normalize, it will be goood
What couch kicks do we use for our SRS/SRT plans?
0, 45, 315
What is the brainstem max dose constraint in 1 Fx SRS?
15 Gy
What is the chiasm max dose constraint in 1 Fx SRS?
12 Gy
What is the Brain-GTV constraitn in 1 Fx SRS? Where does it come from?
V12 Gy < 10 cc
Quantec and TG-101
What is the cord max dose constraint in 1 Fx SRS?
14 Gy
What is the cochlea max dose constraint in 1 Fx SRS?
9 Gy
What is the optic nerve max dose constraint in 1 Fx SRS?
10 Gy
What global hotspot do we keep our SRS/SRTs to?
125%
What is the coverage goals to GTV and PTV that we normally use in our clinic?
**GTV: ** Dmin = 20 Gy
**PTV: ** V100% > 95% (18*0.95)
What is the chiasm max dose constraint in 5 Fx SRT?
25 Gy
What is the lens max dose constraint in 5 Fx SRT?
7 Gy
What is the chiasm max dose constraint in 5 Fx SRT?
25 Gy
What is the brainstem max dose constraint in 5 Fx SRT?
31 Gy
What is the optic nerve max dose constraint in 5 Fx SRT?
25 Gy
What is the Brain-GTV constraint for 5 Fx SRT?
V28.8 Gy < 7 cc
What is the global max hotspot we use for 5 Fx SRT?
125%
30*(1.25) = 37.5 Gy
What is sphere packing? What is it used for?
The technique of filling non-spherical target volumes with a composition of multiple spherical dose distributions of varying sizes
It is used to create a conformal plan for a non-spherical target when you can only deliver spherical distributions (such as with Cone or GammaKnife)
What is the difference between MLC conformal arc and VMAT?
MLC conformal arc moves the MLCs to conform to the target at all angles. This is forward planning
VMAT is inverse planning and moves the MLCs to fit a required fluence mapping patter
What are the pros and cons to using MLC based conformal arc therapy versus cones?
Pros: can treat multiple targets with a single isocenter. Requires shorter treatment times. Can achieve lower hot spots for irregular targets
**Cons: **Larger penumbra than cone due to final collimation being farther from target
When is it better to use cones or MLCs?
Cones: single, spherical targets or targets where you want fastest possible dose fall-off (Ex. Trigems). Lesions far from inhomogenous material in body and atleast 1.5 cm deep
MLC: larger or irregular shaped targets. Multiple targets
What are some new features added to HyperArc that don’t exist in RapidArc?
- Optimization tools specifically for SRS (automatic lower dose objective and SRS NTO tool)
- SRS specific quality metrics (conformity and dose bridging miniization)
- Virtual dry runs
- 1 Click automated delivery with imaging done between kicks
List off as many clinical uses of SRS as you can
AVMs
Meningiomas
Acoustic Neuromas
Trigeminal Neuralgia
Movement disorders
Malignant brain tumors
Metastatic lesions
What are cones made out? What is their approximate length?
Cerrobend lead encased in stainless steel
Approximately 15 cm long
Give a general outline for an “ideal” end to end test
An ideal end to end test is one that tests every step of the SRS workflow and involves everyone who would be responsible for individual steps
This would mean going very deep into detail with an anthropromorphic head phantom, with a mask placed on it during sim by sim therapist, planned by the dosimetrist, approved by physician and physicist, QA performed, and treated by a therapist
Where do we get the majority of our SRS/SRT dose constraints?
TG-101, Quantec and Timmerman
Why in some clinics is it preferred to get a very high hotspot at the center of a lesion?
Center of tumors are typically hypoxic
So a high hotspot at the GTV center can actually partially compensate for the loss in effectiveness due to OER
For what dose per fraction is the LQM generally supported?
1 - 5 Gy/Fx
Beyond this range, the validity of the model is debated
What three non-LQM mechanisms form the basis of SRS radiobiology?
Abscopal Effect
Vascular Damage
Immune-Mediated Response
Describe briefly the abscopal effect
Phenomenon in which local radiotherapy is associated with regression of metastatic cancer in unirradiated lesions
For what dose per fractions does the abscopal affect contribute more? Low or high?
High
Briefly describe the idea of vascular damage in relation to SRS radiobiology
At large dose per fractions (> 10 Gy), death of vascularture leads to lack of blood supply to the tumor
Cell killing continues up to 3 days after irradiation due to tumor starvation
Briefly describe the mechanism of “Immune-Mediated responses” as it pertains to SRS
For high dose/fraction it has been shown that induced local release of tumor specific antigens combined with pro-inflammatory and pro-axidant cytokines improves priming of tumor T cells, contributing to increased immune response to combat cancer
**Note: **For standard fraction the opposite actually occurs. Lower fractionation induces immunosuppression
What does the “Geometric Window of Opportunity” refer to?
Ability of highly conformal dose distributions to avoid critical structures all together
How does one improve the “Geometric Window of Opportunity” (two ways)
Improved localization techniques
Lower intrafraction time
How is TCP affected by inter-fraction time?
You need to give some time between fractions for re-oxygenation and re-distribution to occur, increasing TCP
But at the same time it allows for proliferation and repopulation of tumor, reducing TCP
It’s a balancing act
Describe the procedure of a WL test
A tungsten ball is centered at isocenter as defined by either lasers or imaging
At a combination of different couch, gantry, and collimator angles, a small radiation field will be shot at the ball
During analysis, the center of the radiation field, as defined by the FWHM, is compared to the center of the tungsten ball shadow on the field profile
What is the purpose of the WL test?
It’s both an isocentricity test and a coincidence of radiation and mechanical isocenter test
It measures the size of your isocenter but also how the radiation isocenter agrees with mechanical isocenter
What does a spoke shot test measure?
The size of your isocenter (isocentricity)
What is “Stereotaxis”
Use of a 3D coordinate system to localize a target
In general, how does the size of the tumor affect the prescribed dose? Why?
As size of tumor increases, prescribed dose decreases
This is due to the higher amount of normal tissue dose due to larger beams
What are the positional and dosimetric accuracy requirements for a SRS delivery?
Dose: 3%
Positional: 1 mm
What imaging modality is used for AVMs?
Digital subtraction angiogram (DSA)
What is “General Supervision”
Procedure and staff are under control of rad onc and QMP, but their presence is not required during procedure
What is “direct supervision”
Procedure and staff are under control of rad onc and QMP, and they must be immediately available in the department during procedure
What is “personal supvervision”
Procedure and staff are under control of rad onc and QMP and they must be present during procedure
What level of supervision do we have at our site?
Hybrid,
QMP is personal supervision (they are at the treatment the entire time)
Rad Onc is direct supervision and is only present during setup and time out
Which two reports recommend that QMP should provide personal supervision for atleast the first fraction of SRS/SRT?
TG-100 and MPPG 9
Per MPPG 9, what are some QMP responsibilities in an SRS program? List as many as you can, don’t have to know all
- Be available for consult at CT
- Be available for personal supervision during treatment
- Oversee acceptance testing and commissioning of IGRT/localization, hetereogeneity calcs, TPS, SFD
- Develop E2E tests and perform yearly
- Develop an ongoing QA program
- Develop SOPs and SWIs for everyone involved in the SRS workflow
- Develop safety checklist
- Incorporate an incident learning system
- Supervise or perform treatment planning
- Initial and final plan review
- Perform appropriate plan-specific QA
What is the definition of the RTOG and ICRU defined conformity index?
Ratio of volume of prescription isodose surface and PTV volume
What is the ideal value for the conformity index?
1.0
What does a conformity index > 1 represent? What about < 1?
> 1 means over-coverage
<1 means under-coverage
What is the major limitation of the conformity index?
It is only a ratio of volumes. It gives no indicate whatsoever about the overlapping of the volumes
For all you know the volumes of the target and the prescription line can be exactly equal, so you get a CI = 1. But in reality the isodose line may be completely offset from the target volume
What is the Paddick Conformity Index? What is the equation for it?
It’s a modified version of the conformity index used to account for the overlap of the prescription isodose line and the target
CI = (TVPIV)^2 / (TV*PIV)
- TVPIV is the target volume covered by the prescription isodose volume
- PIV is the prescription isodose volume
- TV is the target volume
What is the ideal value of the paddick conformity index? What is an expected value?
Ideal is 1.0
Expected is > 0.8
True or False
The conformity index is not applied to OARs
True
What is the one limitation of the Paddick Conformity Index?
It does not distinguish between overcovering and undercovering
If a value is not = 1, you don’t know if you’ve over or undercovered the target
Pairing it with the Conformity index may help
What is the definition of the gradient index? What is the ideal value for SRS? What is the expected value for SRS?
Ratio of 50% isodose volume and prescription isodose volume. It’s an indicator of dose fall-off
Ideally 1.0, but this is physically impossible
For SRS, expectation is 3 - 4
Will a smaller target yield a larger or smaller gradient index?
Smaller
What is the homogeneity index?
Max dose in target divided by prescription dose
It’s an indication of the uniformity across the target volume
What is the expected homogeneity index for a Linac based SRS? What about GammaKnife?
Linac SRS: around 1.2 - 1.3
GammaKnife: Around 2.0 or more
Eclipse defines the gradient index a bit differently than most reports. How does eclipse calculate it?
Difference of equivalanet sphere radii of prescription and 50% isodose lines
(unit is cm)
Describe the general function of a CyberKnife unit. How does it work?
Compact x-band LINAC mounted a robotic arm weighing 160 kg
6 MV x-rays are generated froma tungsten alloy target with primary and removable secondary collimators
The arm can move in a hemisphere around the patient
Radiation dose delivery is non-isocentric
Couch is also automated and robotic for repositioning
Arm and couch movements are determined by the computer system
Synchrony respiratory motion management available for lung tumors
How does the CyberKnife Synchrony respiratory motion management tool work?
Can correlate external infrared monitoring system and fiducial movements to internal tumor movement and adjust the targeting of the beam to focus on the tumor in real time while beam is still on
What kind of secondary collimation is available for a CyberKnife?
Circular apertures with dimensions 5 to 60 mm
Why do you need to warmup the chambers of a CyberKnife daily?
They are open chambers and require about 6000 MU to warm them up to the average temperature that they’ll be operating at during a typical treatment day
**Note: ** Closed chambers wouldn’t require this warmup
Newer models have closed chambers. Older models have open chambers
What is the term used to define a combination of location of couch and arm that is beam is delivered from for CK?
Node
What two types of dose calc algorithms are employed for CK?
ray-tracing or monte carlo
How is geometric isocenter defined for CyberKnife?
There is a isocrystal mounted on an isopost, and using the orthogonal x-ray system it serves as the origin for the CyberKnife coordinate system
For CyberKnife, what is the recommended frequency for E2E testing?
Monthly for both intracranial and extracranial
For small field dosimetry of a CyberKnife system, which secondary collimation attachment do you use as your msr field?
60 mm
For all msr fields you want to use the largest possible field size if your machine is incapable of producing a 10 x 10 cm2
Which TG report covers CyberKnife QA?
TG-135
Which protocol is used for CyberKnife reference dosimetry and output calibration? How?
TG-51
But since you can’t find kQ in the traditional way, there are publications that compare PDD10 of a 60 mm collimation to a 10 x 10 cm2 field
You still measure PDD for SSD = 100 cm
What additional safety tests should you consider for Cyberknife?
Since both the couch and arm move often for CyberKnife treatments, there is a lot of collisional safety checks needed for CyberKnife
Additionally, since the arm as it moves can cover camera, you need to make sure you have multiple cameras so not all can be blocked at the same time
There’s also a patient safety zone programmed in which, unless overriden by difficult treatment plans, will not allow the robotic arm to move into
Which TG report covers QA for GammaKnife?
TG 178