First Year Exam: TG-51 Flashcards
What is the approximate difference in measured output between TG-21 and TG-51?
TG-51 output is about 1% discrepancy relative to TG-21 output
But for electrons, it can be more at dmax than 1%, because it TG-39 gave some good improvements to further better improve the accuracy for plane-parallel chambers
True or False
Pion can never be > 1?
False
Pion can never be < 1 is the proper answer
What does it mean if Pion is < 1?
The voltage itself is ionizing the air and giving you a false signal
What is the limit for Pion?
Pion < 1.05
Why can you not use TG-51 for <= 4 MeV electron beams?
Because you need to measured at dmax for TG-51, and that’s too close to surface to measure for the lower energy electron beams
Why are we allowed to measure dose/MU at 10 cm depth (photons) or dref (electrons) when we do TG-51 in our clinic? Why aren’t we doing Dmax?
Because we can just divide the measured Dose/MU by PDD to get the Dose/MU at dmax
True or False
You can calculate kQ using either a 100 SSD or 100 SAD setup?
False
The tables in TG-51 only give kQ for SSD setups. So it doesn’t matter whether you calibrate for SSD or SAD setup, you much find kQ using an SSD setup regardless
Why is TG-51 titled protocol for clinical reference dosimetry of high energy photon and electron beams instead of calibration?
Because TG-51 allows us to tweak the beam
The actually calibration was performed at the ADCL where the chamber was exposed to a standard Co-60 source
Reference dosimetry means we are tweaking a beam relative to the reference, that being a NIST traceable calibration
How does NIST figure out an exact dose to chamber for the standard Co-60 beam?
The Co-60 beam is exposed to a standard chamber of known calibration
The known calibration was determined by measuring the standard chamber response using a NIST traceable water calorimetry experiment
Thus, NIST first determined dose to the NIST chamber, then correlates the reading to an equivalent dose to your chamber at the same dose, which is why a calibrationf actor is in units of Gy/C
What is the EPOM shift for photons and electrons?
- 6rcav for photons
- 5rcav for electrons
In theory, the effective point of measurement for both electron beams and photon beams is 0.85rcav upstream of the chamber’s central axis? In reality, however, the shift is not as large? Why is this?
0.85rcav is assuming that the entrance of the beam to the chamber only dependents on unscattered, direct beam radiation to the surface of the chamber
The reality is that backscatter and lateral scatter shifts the average a bit closer to the center
Why is the shift from central axis to EPOM smaller for electrons than for photons?
Because electrons tend to scatter more lateral, thus moving the average a little bit closer to the central axis of the chamber, compared to photons that scatter more forward than electrons do
Why for photon beams are we allowed to directly measure a %DD curve, but in electron beams we don’t?
To go from %Ionization curve to %DD curve, you need to multiply by the restrictive stopping power ratio of water to air and depth to dmax
For photons, this ratio stays roughly constant as a function of depth (within 0.1% beyond dmax, so %DD = %I
For electrons, the average electron energies change vs depth, thus the restrictive stopping power also changes vs depth, so %DD = %I * Restrictive stopping power ratio
What is standard temperature, pressure and relative humidity?
T = 22 celsius
P = 101.33 kPa or 760 mmHg
Relative humidity is anywhere from 20% to 80%
What is the equation for PTP?
PTP = (273.2+T) / (273.2 + 22) * (101.33/P)
or
PTP = (273.2+T) / (273.2 + 22) * (760/P)
What does PTP actually account for?
Variation in the amss of gas in the chamber volume as temperature and pressure variations cause expansion and contraction of air
What does Pion measure? What is it a function of and why? (two things)
Accounts for ion recombination within chamber volume before the ions are collected by the electrodes
Function of dose rate (changes density of the charge cloud) and chamber bias (changes speed of ion collection)
What is the equation for Pion for a pulsed beam (as would be used in Linac)? What about continuous.
If you don’t remember, then atleast state what measurements you need to take in order to measure Pion?
For Pulsed: Pion = (1 - VH/VL) / (MH/ML - VH/VL)
For continuous: Just square the ratio of voltages terms
Need to measure voltage and raw readings for low and high voltage
Low and high voltages are usually 150V and 300V respectively
How does Pion change as a function of dose per pulse?
Increases as dose per pulse increases
Does Pion increase or decrease for increasing energies?
Pion increases as energy increases
Is Pion typically higher or lower for electrons compared to photons?
Pion is higher for electrons, than for photons
What does Ppol account for?
Changes in collection efficiency due to beam quality, cable positioning, and chamber bias
What is the tolerance for Ppol?
+- 0.4% from 1.000 and within 0.5% between all energies
How do you measure Ppol?
If you don’t remember the equation, atleast say what values need to be measured.
Ppol = (M+raw - M-raw) / 2*Mstandardraw
Where Mstandardraw is just whatever you normally use, either pos or negative. We usually use positive.
Need to measure raw reading at positive and negative. Same absolute voltage, different signs.
How does Pelec correct reading?
It scales reading of the electrometer to true coulombs
How do you measure Pelec?
Typically you don’t. It’s measured at NIST/ADCL
If it’s calibrated with the ion chamber, then whatever correction factor the ion chamber gets will already account for Pelec, so you assume Pelec = 1. If they’re calibrated separately, typically the electrometer measured charge of a known circuit, and the Pelec is a correction based off the ratio of known charge to read charge
What does PQgr account for?
Difference in the effective point of measurement versus the point of measurement during output measurements?
How do you measure PQgr for photons and electrons for cylindrical chambers?
For photons you don’t have to measure it, because it’s already accounted for in kQ
For electrons you have to measure it as follows…
PQgr = Mraw(dref+0.5rcav) / Mraw(dref)
You need to know dref, rcav and Mraw upstream and at dref
Why do you have to shift a PDD curve, but yet when you take your reference dosimetry, you’re allowed to center the chamber either at dref or 10 cm?
Because the kQ (for photons) or the PQgr (for electrons) already effectively take into account the shift
In general, is Pion larger for pulsed or continuous radiation sources?
Pulsed
If Ppol exceeds tolerance threshold, how do you go about accounting for it?
You divide the ND,WCo-60 value by Ppol that was found from time of calibration, and you use that new value in place of ND,WCo-60
How do you use a chamber that differs from those specified in TG-51?
Simply find the closest matching chamber to the list of chambers that they provide
What 4 values determine if your chamber is similar enough to one of the TG-51 chambers?
- Chamber wall material
- Radius of cavity
- Central electrode material
- Wall thickness
What should a waterproofing sleeve be made out of?
PMMA
What setup is required for %DD or %I curves?
100 cm SSD, 10x10 cm2 field (or 10x10 cm2 cone)
Shift using EPOM
What are the differences between the following?
%DD(10)
%DD(10)X
%DD(10)Pb
Which of these three values do you want as your end result?
%DD(10) - PDD at 10 cm including electrom contamination and all
%DD(10)X - PDD at 10 cm not including electron contamination
%DD(10)Pb - PDD at 10 cm with 1 mm of lead in the beam placed 50 cm from the phantom surface (or 30 cm if LINAC doesn’t permit 50 cm)
In the end, you want %DD(10)X for calculating kQ
Why for photon beams of 10 MV or greater do contaminant electrons need to be removed according to TG-51? Why not for energies < 10 MV?
Contaminant electrons cause an increase in the dose at dmax and a subsequent decrease of %DD(10)
For enegries < 10 MV, the contaminant electrons have a short range so don’t have as much of an effect
What is the purpose of the lead foil?
Remove contaminant electrons and introduce a known quantitiy of electrons into the beam
How do you find %DD(10)X?
For E < 10 MV, %DD(10) = %DD(10)X
For E >= 10 MV, or for FFF beams, you use a lead foil to find %DD(10)Pb then use analytical equations to convert from that value to %DD(10)X
True or False
For photon output measurements you are allowed to measure in either 100 cm SSD or 100 cm SAD?
True
Just make sure if you have a SSD setup, you use PDD to scale final reading to desired depth, and for SAD setup, you use TMR to scale final reading to desired depth
In general, what is greater pas dmax, PDD or TMR? Why?
TMR
Because in TMR, the depth in the phantom is the only thing that changes, distance from source remains the same. In PDD, botht he depth AND the distance from source change as you scan. Meaning, the difference between Dose at max vs dose at a deeper depth is larger for a PDD setup (changes more because of the two factors) than for a TMR setup.
How is dref determined?
dref = 0.6*R50 - 0.1
How is R50 calculated?
By using analytical expressions from TG-51 to convert from I50 to R50. These expressions take the form of linear equations
Why for electron measurements is a plane parallel chamber always preferred?
In order to limit variations in Pion, Ppol and PQgr as a function of depth
For what energies is a plane parallel chamber preferred for electrons? For what energies is it required.
Preferred for <= 10 MeV
Required for <= 6 MeV
But most people usually use cylindrical chambers for 6 MeV anyway
For electrons, what three correction factors replace kQ?
PQgr
k’R50
kecal
What is kecal?
A function that takes your chambers calibration factor (which is derived for photons) and applies it to electrons
This factor is chamber specific and constant value
It takes the ADCL calibration factor, and applies it to an arbitary electron beam energy with arbitary beam quality, and is also chamber specific. It’s a singular value because it onlyc ares about the chamber, not the quality of any beam because it just assumes something arbitrary.
What is k’R50?
A factor that accounts for variations in readings due to electron beam quality
It converts from an arbitary electron beam, to your electron beam quality and is a function of R50
How do you meausre k’R50 and kecal?
kecal is found using a look up table in TG-51
k’R50 is plottes vs R50. So first find R50, then use the lookup table or an analytically derived fit function
Where is your k’R50 defined?
It’s only defined/valid for measurements taken at dref
So it’s derived using R50, but it’s only allowed to be applied to readings at dref
What are the 5 goals of the TG-51 addendum?
- Update kQ tables to include more chambers and models so we can do TG-51 with not just farmer chambers
- Provide guidance on what makes a “suitable” chamber for reference dosimetry
- Provide guidance for using TG-51 for new tech (Ex. FFF beams)
- Discuss more details about the determination of ND,W
- Disucss uncertainties present throughout TG-51 and how each part contributes to a final combined uncertainty (establishing an uncertainty budget)
What major feature is left out of the TG-51 addendum?
The addendum only talks about photons. It gives no additional guidance for electrons
According to the addendum, what 6 features must a chamber have in order to be an “appropriate” reference chamber?
- Needs to have chamber stabilization capabilities (time required to get equilibrium reading is not too long and that reading from non-equilibrium to equilibrium states are not too drastically off)
- Should have reasonable leakage current (<0.1% for no correction factor, allowed up to 0.5% if you do have a correction factor to apply. So Pleak < 0.5%, and if < 0.1% then Pleak = 1)
- Minimal polarity correction (Ppol < 0.4%). and correction shouldn’t change across energies by more than 0.5% (relatively stable vs energy)
- High degree of chamber stability (Change in calibration coefficient over 2 years should be < 0.3%)
- Well understood initial and general Pion
- Is not small volume (having volume < 0.05 cc) or high-Z
What are the following recommendations that the addendum makes for…
Chamber equilibration
Pleak
Ppol
Pion
Stability
Chamber equilibration: < 0.5% from first irradiation to equilibrium readings
Pleak < 0.1% of chamber readings
Ppol: 1+-0.4% and < 0.5% variation over all photon energies (meaning, if 6X you get 0.997 but 10X you get 1.003, technically they’re both within 1+-0.4%, BUT, the difference between them is 0.6% so it fails
Pion: Should vary linearly with dose rate, Cinit < 0.002, polarity dependence < 0.001
Stability: Calibration cerificates < 0.3% variation year-to-year
What does the addendum say about the 0.6rcav correction?
For now, it’s fine to continue using it, BUT, it has been shown since TG-51 that the correction is not actually true, and that ti dependents on the length of the cavities and central-electrode diameter. So expect in a future addendum that the recommendation for shift becomes chamber specific.
What changes to kQ values did the addendum make?
TG-51 originally used a semi-analytical approach to calculate kQ values
The addendum used monte-carlo, which is more precise, to calculate kQ
The addendum values agree within 0.5% of the original values
Experimental kQ values were collected and showed excellent agreement with MC
Approximately how much of an error in kQ occurs when a change in %dd(10)X of 1% is measured?
Change in %dd(10)X of 1% yields a change of ~ 0.15% in kQ
When was TG-51 published?
When was the TG-51 addendum published?
TG-51: 1999
Addendum: 2014
True or False
The addendum provides a graph of kQ for chamber and beam quality using analytical data fit
False
The addendum provides a table. That table has a bunch of difference detectors, and gives 5 kQ values for each detector depending on the %dd(10)X (63, 67, 73, 77, 81%). The data is found using monte carlo for the addendum.
What is the major flaw in the conversion from %dd(1) to %dd(10)X equation from TG-51?
The quation is only defined for flat fields, not FFF
What uncertainty does using the %dd conversion equation introduce for flattened fields?
0.2% error, so the addendum recommends you can still continue to use the equation for flattened fields, but you should account for the error in your “budget”
What recommendation does the addendum make regarding lead foil and FFF beams?
The lead foil should still be used for FFF beams. The equation is not suitable
Summarize the recommendations to calculate %dd(10)X for the following energies…
6X
6FFF
10X
10FFF
15X
6X - you may measure %dd(10). No conversion needed. This is approximate = to %dd(10)X
6FFF - you should use a lead foil, measure %dd(10)Pb, then use conversion to get to %dd(10)X
10X - You can use the equation if you want. It’s a 0.2% error, so just keep that in mind. Or you can use the lead foil and convert to %dd(10)X.
10FFF - you should use a lead foil, measure %dd(10)Pb, then use conversion to get to %dd(10)X
15X - You can use the equation if you want. It’s a 0.2% error, so just keep that in mind. Or you can use the lead foil and convert to %dd(10)X.
Summarize the lead foil recommendations from TG-51 and the addendum
If energy < 10 MV, and beam is not FFF, you don’t have to use foil or any extra conversion
If beam is FFF, you have to use foil and convert to %dd(10)X using equation
If beam is >= 10 MV and is flat, you can either use a lead foil or not use one. But if you decide not to use one, account for the 0.2% error since %dd(10) is not exactly equal to %dd(10)X. If you do use the foil, measure %dd(10)Pb and conver to %dd(10)X using equation
What two correctionf actors does the addendum introduce?
Prp and Pleak
what is Prp and how do you measure it?
It’s a factor that takes into account radial asymmetry of a beam resulting in differential fluence being delivered to the reference chamber
For flattened beams - you can do a 1D integration along the length of the chamber and ratio the central value quantity to average value
For unflattened beams - you do the same but using a 2D integration (likely with film). Take the average and do central value quantity / average value
How do you determine Pleak?
If < 0.1%, you can just ignore it and call Pleak = 1
If < 0.5% and > 0.1%, it’s tolerable, but you need to apply a non-unity Pleak
What three requirements are very specific to FFF beams that aren’t as applicable to flattened beams?
- You need to be extra sure that Pion < 1.05 (since the higher dose per pulse makes that more challenge to accomplish)
- You need to be more thorough with calculating your Prp since it’s now more important
- You must use lead foil for flattened beams
According to the addendum uncertainty budget, what is the approximate combined uncertainty you can expect from the entire TG-51 process?
~ 1%

What is the minimum size ion chamber you’re allowed to use for reference dosimetry?
0.05 cc
What does the addendum say about ion chambers and Z dependence?
Don’t use ion chambers with high-Z electrodes
What does the addendum say about the use of plane parallel chambers for photon-reference dosimetry? What about electron?
Photons: DO NOT USE PLANE PARALLEL CHAMBERS, only cylindrical
Electrons: The addendum doesn’t talk about electrons
Does the addendum lean more trowards using or not using lead for flattened photon beams?
It actually leans towards not using lead, and instead just adding the 0.2% uncertainty to the budget
Which kQ values should you use in your clinic? The ones from the addendum, from original TG-51, or it doesn’t really make a big difference?
The addendum
Cause they’re more accurately derived (MC), and align better with experimental results
What is the maximum voltage you’re allowed to use for reference dosimetry?
300 V
How often should you determine an uncertainty budget?
Everytime your procedure changes or you use different equipment
What point do you perform your reference dosimetry output calculation for photons and electrons?
For photons: Always 10 cm depth
For electrons: Always dref
Why for electrons do we measure reference dosimetry at dref and not R50 or dmax?
R50 is in a range of rapid fall-off, so setup uncertainty is much greater
dmax is in a region with significant scatter electron contribution at higher energy fields
dref is the perfect balance where there is less scatter electron contribution, and it’s still in the relatively flat region so setup uncertainty doesn’t ruin measurements too much
Why in the addendum is it PREFERRED to use the correction formula to go from %dd(10) to %dd(10)X even for higher energy MV beams, instead of just using lead?
Placement of lead can introduce uncertainty in its own way, and it adds extra complexities. If you use the correction formula, yes you may be less accurate, but the accuracy is actually lower than what was previously expected. The benefit from using the lead foil is small.
TG-51 sites a maximum uncertainty in using the formula for higher energy photon beams of 0.4%
The addendum sites a max uncertainty in using the formula for higher energy photon beams of about 0.2%. It’s not really worth it to use the foil. It will improve uncertainty, but barely.
What is the range of photon energies that you may use TG-51 for?
Co-60 - 50 MV
What is the full equation for photon DQW using all of the correction factors?
DQW = kQPionPelecPTPPpolPrpPleakMraw
What is the actual meaning of %dd(10)X?
It’s the photon component of the percentage depth dose at 10 cm depth for a field size of 10x10cm2. This quantity either assumes no electron contamination (as is approximately the case for < 10 MV beams), or it has already factored out the electron component (as is the case of applying the equations)
At what energy electron beams does the cone size need to increase from 10x10 to 20x20, and why is this?
For electron energies > 20 MeV
This is because charged particular equilibrium is lost if you continue to use 10x10cm2 past 20 MeV
What setup is electron clinical reference dosimetry performed with?
Field size >= 10x10 cm2
An SSD value somewhere between 90 and 110 cm (usually 100 cm)
How is TG-51 a simplification from TG-21? (Two main benefits)
- We don’t need all the large stopping-power ratio and mass-energy absorption coefficient tables anymore
- We don’t need to calculate any theoretical dosimetry factors anymore
TG-51 was the 3rd generation of reference dosimetry protocols. TG-21 was the 2nd. What was the major error with the very first generation?
The very first protocol was overly simplified, and was found to lead to errors in beam calibrations of up to 5%
TG-21 fixed these issues, but at the expense of making things overly complex, especially for chamber-dependent factors.
What are the TG-51 requirements on phantoms that are appropriate to use?
For annual you MUST use water phantom of dimensions atleast 30x30x30 cm3
For routine and more frequent QA checks, you can use plastic materials (such as solid water) as long as you have a transfer factor (fudge factor) that was established relative to your water measurement
But reference dosimetry measurements in plastics and water-equivalent plastics are not allowed at all. It’s fine for QA, but you shouldn’t be adjusting anything off of it.
What are the similarities and differences between %dd(10) and %dd(10)Pb
Similarities: They are both directly measured and do not require any equation to get to them. Additionally, they both contain an electron component
Differences: One has lead placed, the other is open field. %dd(10)Pb has a predictable electron component, %dd(10) is not as predictable
What is the major difference between %dd(10) and %dd(10)Pb vs %dd(10)X?
To find %dd(10)X you first start at one of the other two, then apply a correction. Also %dd(10)X does not include the electron component of the photon beam, only the photon component
What are the units of DQW? What is the approximate magnitude?
Gy/C
_____ x 107 typically
Conceptually speaking, what is kQ? Why is it actually necessary?
It’s a quality conversion factor that accounts for the change in the absorbed-dose to water calibration factor between the beam quality of interest, Q, and the beam quality for which the calibration factor was first measured (Co-60)
Remember: the correction factor you get for ADCL/NIST only applies for a Co-60 unit. It doesn’t actually apply to your machine. So you need a value to convert the correction factor to your machine. That’s what kQ does
Keep in mind then, if you’re doing reference dosimetry with TG-51 for a Co-60 unit, kQ = 1.000
What is kQ for an electron beam broken down into? Why is this necessary?
kQ = kR50PQgr
It’s broken into two components, a gradient corrected and non-gradient corrected component. This is necessary because the gradient correction for electron beams is so variable, that it actually changes machine-to-machine. So you need one specifically for your Linac of interest.The kR50 does not change as much can be found in look up tables in TG-51
What are the two components of kR50? Why are they necessary?
kR50 = k’R50kecal
kecal is used to convert the ADCL calibration factor from photon to some arbitrary electron beam
k’R50 is needed to convert from an arbitrary electron calibration factor, to one that matches your machine’s beam quality
Is ND,W use dto measure absorbed dose at the chamber’s point of measurement in the absence of or with the chamber in place?
Obviously in the absence of a chamber. That’s the whole point of reference dosimetry
Otherwise why would you need a correction factor for something that you’re measuring directly anyway?
In general, if a factor is labeled ‘k’ with anything following it, what is it doing?
Converting the absorbed-dose calibration factor from one quality to another
In general, what are any of the ‘P’ correction factors responsible for?
Correcting for equipment or environment discrepencies
True or False
There are two independent equations to convert from %dd(10)Pb to %dd(10)X. One is for foil at 50 cm +- 5 cm, the other is for foil at 30 cm +- 1 cm.
True
How do you convert from a measured I50 value, to your wanted R50 value?
Use the conversion equations in TG-51 (equations 16 and 17)
What three options are there to measure R50?
- Measure using diode detector directly (but then your diode detector needs to be well established)
- Measure I50 directly, then use conversion equation in TG-51
- Measure the %I curve, then convert all of it to %dd curve (TG-25 method, but the general way to do this is an equation that tells you stopping power ratios as a function of R50 and depth from a Burns et al paper)
What photon beam would in general be hotter? One tweaked to give 1 cGy/MU at reference condition in SSD setup, or one in SAD setup?
Trick question
They both give 1 cGy/MU at dmax
For the SAD setup, the D10 will be higher, but then it just gets divided by TMR which is higher anyway (because the chamber is closer to the source). For a SSD setup, D10 is lower but it just gets divided by PDD which is lower anyway
In reality, you can perform clinical reference dosimetry calculations at any depth that you want. Why is this?
Because the only difference would be applying your PDD or TMR to get you to the actual point that you want to make 1 cGy/MU
Give the equation for dose to water from an electron beam
(use Mcorrected instead of Mraw multiplied by all the correction factors)
(also expand the kQ factor)
DQw = MPQgrk’R50kecalN60CoD,w
What process is used to avoid obtaining a absorbed dose in water calibration factor from ADCL for plane-parallel chambers? Give a general overview of how it works
Cross Calibration
Your end result that you want to find is (kecalN60CoD,w)pp = (Dw)cyl / (Mk’R50)PP
Where the numerator term is just your dose measured using a cylindrical chamber (with all corrections applied), and your denomenator is as it appears.
True or False
TG-51 and addendum can be applied to Tomo, GK and CyberKnife?
False
They are not capable of the setups specified in the report (100 SSD or SAD and 10 x 10 cm2 fields)
True or False
Plane parallel chambers may be used for photon reference dosimetry
True
But just keep in mind, they won’t be anywhere as stable as a farmer chamber or a most cylindrical chambers. Just expect their performance won’t be as good. But yes, if you absolutely wanted to and were willing to account for the added uncerataint, you can use certain models of plane parallel chambers
Why is having high-Z electrode such an issue for photon beams?
They’e mainly an issue with FFF beams, where they are not well specified by %dd(10)X
The addendum gives kQ equation for %dd(10)X between 63 and 86. What happens if you’re below 63%?
Lineraly interpolate from a kQ given %dd(10)X = 63%, to that of a Co-60 unit, kQ = 1.000 at %dd(10)X = 58%
When must you perform a full TG-51?
Annual and commissioning
Why are small voluem chambers not recommended for TG-51?
Anomalous recombination behavior
Large polarity effect
Complex effects from high-Z electrodes
Just overall instability
What does a fudge factor do?
Converts from dose in plastic to dose in water
Where do the dosimetric uncertainties for each item in the addendum uncertainty budget come from?
TG-142
This assumes your machine is within the tolerances of TG-142
Why do we usually only do a temperature and pressure correction, and not a hmidity correction?
Because humidity in most situations in the US and Canada only leads to an error of 0.15% at the absolute maximum. It’s small enough to ignore
Labs usually give a range of relative humidity for which the calibration coefficient is valid, and this range is very large (20-80% but ideally 40-60%)
What uncertainty k value do ADCL calibrations report? What about the recommended uncertainty budget?
Uncertainty budget recommends k = 1
ADCL reports k = 2
You would need to convert from k = 2 to k = 1 to get the uncertianty of the calibration coefficient
What does the stability (or time needed to get to equilibrium) (or chamber settling) of an ion chamber depend on?
Irradiation History
How long its been since it was last used
Whether polarity is being changed back and forth or not
If kQ values are calculated using montecarlo and taking into account the exact build of a chamber and the flatness and symmetry of the field, and would therefore already account for volume averaging, then why do we need a Prp factor at all?
Isn’t this redundant?
What uncertainty occurs from using both the implicit volume averaging correction in kQ and the Prp?
That’s because it doesn’t actuallya ccount for the correct flatness and symmetry. It assumes a relatively much more uniform profile than what is typically found in most LINACs
You have an error of at most 0.05% if you use both the kQ and Prp. Very insignificant. That’s why you always also measure the Prp
What is the chamber settling tolerance recommendation given by the addendum for a suitable ion chamber?
Less than a 0.5% change in chamber reading per monitor unit from a beam-on for a warmed up machine, to stabilization of the ionization chamber
So unstable to stable shoudl have a change in reading less than 0.5%
What is the restricted mass stopping power?
Quantity used to calculate dose to a medium with the restriction that only interactions below a certain cutoff energy are used in calculating dose
Reasoning: electrons of higher energy will travel away from site and contribute dose non-locally
What affect does not accounting for electron contamination have on your %dd at 10 cm?
Electron contamination will increase dose at dmax. As such your %dd at 10 cm will decrease if you don’t account out the electron contamination.
This effect is very small for < 10 MV because the electron contamination doesn’t penetrate that deep, but it reaches up to a 0.2% error in kQ for >= 10 MV beams
Ideally where should the lead be placed? What about if it can’t be placed in the ideal situation?
Ideally: 50 cm from phantom surface
Also acceptable: 30 cm from phantom surface
How thick is the lead foil?
1 mm
For reference dosimetry, what TMR/%DD value should you use to scale dose at 10 to dmax? What source should you get that data from, the commissioning data or the data you take on the day of measurement?
You must use the COMMISSIONING DATA! That’s because it’s what the TPS uses as well. If your TPS thinks you’re delivery 1 cGy/MU at dmax SAD or SSD setup, then you need to be consistent.
Do not use whatever %dd(10)X you measured on the day of. Even if it’s annual. You have to use whatever is in TPS
For electron and photon beam dosimetry, where are you trying to get 1 cGy/MU?
At dmax for both
What is the single item in the uncertainty budget that contributes the largest uncertainty?
The calibration factor you get from ADCL
That’s about a 0.75% uncertainty typically
What 5 values compose kQ?
kQ = [PwallPcelPflPgr(L/p)wair]Q / [PwallPcelPflPgr(L/p)wair]60Co
So notice, kQ accounts for properties of the physical chamber makeup that vary with respect to beam quality. All of them are pretty obvious in their naming, the only one that might not be is Pcel which is the central electrode correction factor
Which of the following is the most important characteristic that must match when choosing an appropriate chamber to resemble your chamber?
Central electrode material
Wall thickness
Wall material
Radius of cavity
Central electrode material
Wall thickness
Wall material
Radius of cavity
The only just the wall material is the same, you will be within 0.5%
Which correction factors are likely to be larger in FFF beams than in flattned beams?
Prp and Pion
How is the Pion value further decomposed in the TG-51 addendum?
Break it down into a Cgen and a Cinit value
Cgen is related to the dose per pulse in pulsed linac beams
Cinit should be equal to that measured by ADCL in Co-60, and should be < 0.002
Pion = 1 + Cinit + CgenDpp
The whole point of this breakdown is mainly for chamber commissioning purposes, but also suggests that Pion increases linearly with dose per pulse
Simply put, what is the end result of Bragg-Gray cavity theorem?
To relate dose measured in air to dose measured in water
What is the effective point of measurement for a single electron defined as?
The location at which the electron enters the chamber or detector
What is the addendum rule of thumb for uncertainty of output per 1 cm error in field size?
1%/cm uncertainty due to field size uncertainty