First Year Exam: QA Flashcards
What is the daily QA output dose tolerance?
3%
In simple words, what is the purpose of Daily QA?
To check output and watch for trends in parameters
True or false
For photon energy tracking in daily QA, you are measuring absolute energy.
False
You are tracking relative change since baselines
How does a device like daily QA calculate output?
It has a central axis ion chamber which is corrected for temperature and pressure
A background collection is taken and subtracted out
You apply an array calibration which was first taken at time of output adjustment, when you were confident that delivered dose was exactly what you know
You measure relative to your baseline output (which you would’ve taken when everything was 1.000)
True or false
For electron energy tracking in daily QA, you are measuring absolute energy.
True
How setup is used by DQA3 to calculate electron energy?
5 ion chambers around the board, each with different buildup materials, thicknesses, and different depth dose
What is the accuracy of electron energy calculations?
4-6 MeV: within 2% accuracy
6-20 MeV: within 1% accuracy
What mechanical tests do you do for morning QA? What are their TG-142 recommended tolerances (assuming SRS/SBRT machine type)?
Laser localization: 1 mm
ODI: 2 mm at ISO
Collimator size indicator: 1 mm
What safety checks do you do on Linac for Daily QA?
Audio and Visual Door interlock Beam off interlock Beam on light indicators Collision interlocks
What imaging tests do you do daily? What are the tolerances?
Positioning/repositioning (IGRT) accuracy: 1 mm
Isocal (for SRS machine only really): 0.035, 0.05, etc
MV and kV alignment: 1 mm
Collision interlocks: (pass or fail)
What does Winston Lutz tell you about your machine?
Size of isocenter
What is the Winston Lutz test actually testing?
Coincidence of the radiation and mechanical isocenter
And for us specifically, also imaging isocenter since we use CBCT to setup the cube. But if you used lasers and no fine tuning with imaging, then you wouldn’t be using imaging isocenter
What are the tolerances for the WL test?
0.5 mm mean deviation
1 mm max deviation
How does doselab pro calculate result for winston lutz?
- First figures out the edge of the radiation field by finding the 50% threshold
- Then determines the center of each field
- Then locate sthe center of the ball using image thresholding as well
- Then the deltas are determined for x and y direction separately
- Total delta is then just distance formula
What does MPC check?
Geometric accuracy using just 6X
output for all energies
What is the tolerance for the picket fence test (assume SRS machine)
Technically, according to TG-142, it’s just visual inspection
But we use 0.5 mm max deviation since we quantitatively measure
What MLC tests are TG-142 recommended for monthly QA? What are the tolerances?
Pattern test, atleast 2 patterns: 2 mm
Travel speed: < 0.5 cm/s deviation from max
Picket Fence: 1 mm
What MLC tests are TG-142 recommended for annual QA?
Transmission
Picket Fence
Spoke shot
Coincidence of light and x-ray field
Segmental IMRT (step and shoot) test or moving window (IMRT) test for four cardinal gantry angles
What images are taken in isocal?
4 MV images at different collimator rotations
MVCT images
kVCT images
What is the purpose of isocal calibration?
To fine-tune imager alignment so that imagers are exactly centered to radiation isocenter with respect to gantry angle
The calibration calculates the needed longitudinal and lateral corrections of both MV and kV images at each angle, to make this happen
The angle-by-angle corrections are saved to the system and applied to all imaging thereafter
What is the tolerance for a monthly TG-51 output constancy check?
2%
What is the tolerance for a monthly profile constancy check?
1%
What is the R50 monthly tolerance?
2mm
What is the pdd(10) monthly tolerance?
2%
What does beam profile constancy mean?
Making sure your current profile is consistent with your baseline profile
What is the annual TG-51 output calibration tolerance for x-ray and electrons?
+-1%
What are some mechanical TG-142 monthly tests?
Light vs Radiation Field Jaw positioning accuracy ODI vs Front Pointer Laser localization Gantry/Collimator angle accuracy Accessory functionality Crosshair centering (walkout) Couch position shifts
What is the tolerance for Light vs Radiation Field?
2 mm symmetric (or 1%) on a side
1 mm asymmetric (or 1%) on a side
What is the tolerance for Jaw Positioning accuracy?
2 mm symmetric
1 mm assymmetric
Tolerance for ODI vs front pointer
1 mm at isocenter
Tolerance for monthly ODI vs front pointer
1 mm at isocenter
Tolerance for monthly laser localization
< 1 mm
Tolerance for monthly gantry/collimator angle accuracy
1 degree
Tolerance for monthly accessory functionality
pass/fail
Tolerance for monthly crosshair centering (walkout)
1 mm
Tolerance for monthly couch positioning (SRS machine)
1 mm / 0.5 degrees
What safety tests are recommended for monthly according to TG-142?
Laser Guard functionality
What OSMS/Calypso tests are recommended monthly by TG-142?
Output constancy
Phase/Amplitude beam control (magnitude and localization pretty much)
In-room respiratory monitoring
Gating interlock
What are the TG-142 recommended imaging tests for kV and MV, and their tolerances?
Imaging and treatment coordinate coincidence - < 1 mm (this is pretty much what we do for daily QA) Scaling - < 1 mm Spatial resolution - baseline Contrast - baseline Uniformity - baseline Noise - baseline
What are the TG-142 recommended imaging tests for Cone-Beam, and their tolerances?
Geometric distortion - < 1 mm Spatial resolution - baseline Contrast - baseline HU constancy - baseline Uniformity - baseline Noise - baseline
What does array calibration, in general, spit out?
The relative sensitivity differences between detectors in an array. The actual calibration spits out individual correction factors for each detector
What does dose calibration, in general for arrays, spit out?
A single value that converts relative dose values to absolute dose values
True or False
Array calibrations are energy and accelerator specific?
False
Array calibrations are all relative, so the machine that it’s done on is not important
True or False
Dose calibrations are energy and accelerator specific?
True for both!
But for us, since all three machines are matched, we use the same dose calibrations for each machine.
What does background measurements actually measure?
Leakage current of each individual detector. This calculates the background rate
What is the spacing and density of diodes in an arccheck?
Spacing: 10 mm
Density: Sub mm due to helical arrangement (diagonal short cut and entrance and exit overlapping)